Use of a Combined SpO2/PtcCO2 Sensor in the Delivery Room

被引:4
|
作者
Rubortone, Serena Antonia [1 ]
De Carolis, Maria Pia [1 ]
Lacerenza, Serafina [1 ]
Bersani, Iliana [1 ]
Occhipinti, Federica [1 ]
Romagnoli, Costantino [1 ]
机构
[1] Univ Cattolica Sacro Cuore, Div Neonatol, Dept Paediat, I-00168 Rome, Italy
来源
SENSORS | 2012年 / 12卷 / 08期
关键词
oxygen saturation; partial pressure of carbon dioxide; pulse oximeter; delivery room; TOSCA sensor; neonate; CARBON-DIOXIDE TENSION; NEONATAL RESUSCITATION; OXYGEN-SATURATION; PULSE OXIMETRY; GUIDELINES; BLOOD;
D O I
10.3390/s120810980
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Arterial oxygen saturation (SaO(2)) and partial arterial pressure of carbon dioxide (PaCO2) are important respiratory parameters in critically ill neonates. A sensor combining a pulse oximeter with the Stow-Severinghaus electrode, required for the measurement of peripheral oxygen saturation (SpO(2)) and transcutaneous partial pressure of carbon dioxide (PtcCO(2)), respectively, has been recently used in neonatal clinical practice (TOSCA(500 (R))Radiometer). We evaluated TOSCA usability and reliability in the delivery room (DR), throughout three different periods, on term, late-preterm, and preterm neonates. During the first period (period A), 30 healthy term neonates were simultaneously monitored with both TOSCA and a MASIMO pulse oximeter. During the second period (period B), 10 healthy late-preterm neonates were monitored with both TOSCA and a transcutaneous device measuring PtcCO(2) (TINA (R) TCM3, Radiometer). During the third period (period C), 15 preterm neonates were monitored with TOSCA and MASIMO after birth, during stabilization, and during transport to the neonatal intensive care unit (NICU). Blood gas analyses were performed to compare transcutaneous and blood gas values. TOSCA resulted easily and safely usable in the DR, allowing reliable noninvasive SaO(2) estimation. Since PtcCO(2) measurements with TOSCA required at least 10 min to be stable and reliable, this parameter was not useful during the early resuscitation immediately after birth. Moreover, PtcCO(2) levels were less precise if compared to the conventional transcutaneous monitoring. However, PtcCO(2) measurement by TOSCA was useful as trend-monitoring after stabilization and during transport to NICU.
引用
收藏
页码:10980 / 10989
页数:10
相关论文
共 50 条
  • [1] Evaluation of a new combined SpO2/PtcCO2 sensor in anaesthetized paediatric patients
    Dullenkopf, A
    Bernardo, SD
    Berger, F
    Fasnacht, M
    Gerber, AC
    Weiss, M
    PAEDIATRIC ANAESTHESIA, 2003, 13 (09): : 777 - 784
  • [2] An evaluation of a new combined Spo2/PtcCO2 sensor in very low birth weight infants
    Lacerenza, Serafina
    De Carolis, Maria Pia
    Fusco, Francesca Paola
    La Torre, Giuseppe
    Chiaradia, Giacomina
    Romagnoli, Costantino
    ANESTHESIA AND ANALGESIA, 2008, 107 (01): : 125 - 129
  • [3] An evaluation of a new combined Spo2/PTCCO2 (vol 107, pg 125, 2008)
    Lacerenza, S.
    De Carolis, M. P.
    Fusco, F. P.
    La Torre, G.
    Chiaradia, G.
    Romagnoli, C.
    ANESTHESIA AND ANALGESIA, 2008, 107 (04): : 1389 - 1389
  • [4] Longtime performance and reliability of two different PtcCO2 and SpO2 sensors in neonates
    Bernet, Vera
    Doell, Carsten
    Cannizzaro, Vincenzo
    Ersch, Joerg
    Frey, Bernhard
    Weiss, Markus
    PEDIATRIC ANESTHESIA, 2008, 18 (09) : 872 - 877
  • [5] Longtime performance and reliability of two different PtcCO2 and SpO2 sensors in neonates
    Bernet, V
    Doll, C.
    Cannizarro, V
    Frey, B.
    Weiss, M.
    SWISS MEDICAL WEEKLY, 2007, 137 : 27S - 27S
  • [6] Skin-to-skin care in the delivery room: Impact of SpO2 monitoring
    Tourneux, P.
    Dubruque, E.
    Baumert, A.
    Carpentier, E.
    Caron-Lesenechal, E.
    Barcat, L.
    Elalouf, C.
    Blanchard, M. -C.
    Boniface, A.
    Poupart, C.
    Gondry, J.
    Fontaine, C.
    ARCHIVES DE PEDIATRIE, 2015, 22 (02): : 166 - 170
  • [7] FIO2 AND SpO2 TARGETS IN PRETERM DELIVERY ROOM RESUSCITATION - WHAT DO THEY MEAN?
    Vento, M.
    Saugstad, O.
    Wright, I.
    Finer, N.
    Rich, W.
    Rook, D.
    Kapadia, V.
    Rabi, J.
    Tarnow-Mordi, W.
    Oei, J. L.
    EUROPEAN JOURNAL OF PEDIATRICS, 2016, 175 (11) : 1565 - 1565
  • [8] Observational study of perioperative PtcCO2 and SPO2 in non-ventilated patients receiving epidural infusion or patient-controlled analgesia using a single earlobe monitor (TOSCA)
    Kopka, A.
    Wallace, E.
    Reilly, G.
    Binning, A.
    BRITISH JOURNAL OF ANAESTHESIA, 2007, 99 (04) : 567 - 571
  • [9] The utility of PtcCO2 in the diagnosis of sleep disordered breathing
    Pan, C. J. -H.
    Cossa, G.
    Brillante, R.
    JOURNAL OF SLEEP RESEARCH, 2018, 27
  • [10] Effect of lateral displacement of the sensor on SpO2 value
    Kimoto, Natsuko
    Kobayashi, Naoki
    Yarita, Masaru
    Tamura, Toshiyo
    Minato, Kotaro
    Transactions of Japanese Society for Medical and Biological Engineering, 2012, 50 (06): : 651 - 657