A total of 717 faeces samples were tested prospectively using the EntericBio Panel II (R) detection system (Serosep, Limerick, Ireland), in parallel with routine laboratory testing, which combines the EntericBio (R) system with retrospective culture of each specimen where a target is detected. Discrepancy analysis was conducted using molecular methods. The EntericBio Panel II (R) assay produced 585 negative and 132 positive results, namely, Campylobacter spp. (n = 66); SLT 1 and/or SLT 2 (n = 64); Salmonella spp. (n = 5); and Shigella spp. (n = 0). Three samples were positive for more than 1 target. Of these results, 4 Campylobacter spp. detections and 4 SLT 1/SLT 2 detections remained unconfirmed, and the system failed to detect 2 Campylobacter spp. targets detected by routine laboratory detection. The sensitivity, specificity, positive predictive value, negative predictive value, and efficiency were calculated to be 98.4%, 98.7%, 93.9%, 99.7%, and 98.6%, respectively. (C) 2013 Elsevier Inc. All rights reserved.