摘要
Background: Since blood culture bottles are validated by the manufacturer for blood only, an additional validation for the use with fluids of tissue preparations is necessary. Methods: Two 10-ml samples of cornea culture medium, histidine-tryptophan-ketoglutarate (HTK) solution, or Ringer solution at the end of femur head thermodisinfection were given into blood culture bottles (BD BACTEC (TM) Plus Aerobic/F, Anaerobic/F for cornea culture medium and BD BACTEC (TM) Standard Aerobic/Anaerobic for HTK and Ringer solution) and subsequently spiked with 10-100 colony forming units (CFU) of bacteria or fungi (aerobic bacteria: Staphylococcus aureus, Bacillus subtilis, Pseudomonas aeruginosa; anaerobic bacteria: Clostridium sporogenes; fungi: Candida albicans, Aspergillus brasiliensis) according to the European Pharmacopoeia Chapter 2.6.1. Results: All tested bacteria and fungi could be detected in all solutions. All positive and negative controls were tested correctly. Compared to the positive controls, the microbial growth was delayed in the antibiotic-containing cornea culture medium, and negative in two cases of B. subtilis spiking. Conclusion: The use of BACTEC (TM) blood culture bottles seems to be a suitable method for microbiological testing of HTK solution, Ringer solution, and, with limitations, also for testing of the antibiotic-containing cornea culture medium.
引用
收藏
页码:387 / 390
页数:4
相关论文
共 50 条
  • [1] Validation of Microbiological Testing of Tissue Preparations with Different Incubation Temperatures
    Herrlinger, Frithjof
    Schulz, Tino
    Pruss, Axel
    Schulz, Eva
    TRANSFUSION MEDICINE AND HEMOTHERAPY, 2021, 48 (01) : 23 - 30
  • [2] MICROBIOLOGICAL COMPARISON OF A NEW INFRARED BLOOD CULTURE SYSTEM (BACTEC-NR660) AND A RADIOMETRIC SYSTEM (BACTEC-460)
    CORKILL, JE
    RIMMER, K
    MEDICAL LABORATORY SCIENCES, 1987, 44 (02): : 150 - 159
  • [3] Validity of reducing blood culture incubation time for the BD BACTEC FX blood culture system considering microbiological and clinical aspects
    Hashimoto, Takehiro
    Yahiro, Takaaki
    Khan, Sakirul
    Kimitsuki, Kazunori
    Nishizono, Akira
    Hiramatsu, Kazufumi
    JOURNAL OF INFECTION AND CHEMOTHERAPY, 2025, 31 (01)
  • [4] Comparison of the clinical and microbiological characteristics of Campylobacter and Helicobacter bacteremia: The importance of time to blood culture positivity using the BACTEC blood culture systems
    Yamamoto K.
    Hayakawa K.
    Nagashima M.
    Shimada K.
    Kutsuna S.
    Takeshita N.
    Kato Y.
    Kanagawa S.
    Yamada K.
    Mezaki K.
    Kirikae T.
    Ohmagari N.
    BMC Research Notes, 10 (1)
  • [5] Validation of microbiological testing in cardiovascular tissue preservation
    Schroeter, J.
    Skenderi, Z.
    Schulz, E.
    Pruss, A.
    TRANSFUSION MEDICINE AND HEMOTHERAPY, 2017, 44 : 84 - 84
  • [6] The BD BACTEC FX blood culture system with the gentlemacs dissociator is suitable for sterility testing of heart valve and vascular allografts—A validation study
    Ramadan Jashari
    Sarah Vanzeebroeck
    Pieter Petit
    Hector Rodriguez-Villalobos
    Sharon Zahra
    Naziha Ben Said
    Vanessa Bouzet
    Stefano Mastrobuoni
    Cell and Tissue Banking, 2021, 22 : 453 - 466
  • [7] Isolation of Mycoplasma hominis using the BACTEC 9000 series blood culture system
    Kitson, KA
    Wright, KC
    JOURNAL OF CLINICAL PATHOLOGY, 1996, 49 (08) : 686 - 687
  • [8] CRYPTOCOCCEMIA AND THE BACTEC-460 BLOOD CULTURE SYSTEM
    LOVE, GL
    BOYD, GD
    LEBLANC, EJ
    ANNALS OF INTERNAL MEDICINE, 1987, 106 (04) : 633 - 634
  • [9] Time-to-positivity of aerobic blood cultures by using BacTec 9120 blood culture system
    Fokas, St.
    Fokas, Sp.
    Altouvas, G.
    Tsironi, M.
    Kaptanis, S.
    Kalkani, M.
    Dionysopoulou, M.
    INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 2007, 29 : S244 - S245
  • [10] The BD BACTEC FX blood culture system with the gentlemacs dissociator is suitable for sterility testing of heart valve and vascular allografts-A validation study
    Jashari, Ramadan
    Vanzeebroeck, Sarah
    Petit, Pieter
    Rodriguez-Villalobos, Hector
    Zahra, Sharon
    Ben Said, Naziha
    Bouzet, Vanessa
    Mastrobuoni, Stefano
    CELL AND TISSUE BANKING, 2021, 22 (03) : 453 - 466