Calibration Method for the Monochromator Based on Continuous Spectrum Light Source

被引:0
|
作者
Zhang Lei [1 ]
Dai Jing-min [1 ]
机构
[1] Harbin Inst Technol, Instrument Sci & Technol, Harbin 150001, Peoples R China
关键词
Grating monochromator; Continuous spectrum light source; Transmissivity function; Linear fit; RESPONSIVITY;
D O I
10.3964/j.issn.1000-0593(2015)08-2348-04
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
For the inevitable errors of the monochromator in the process of producing and installing, it must be calibrated before being used. For the traditional calibration method using mercury lamp, the number of wavelengths limits the calibration precision of the monochromator. Based on the principle of grating monochromator, calibration method for the monochromator based on continuous spectrum light source is proposed. Calibration system is designed and built, which is composed of broadband light source using tungsten halogen lamp, power source, light filter, the monochromator, spectroscopy to detect the intensity of the monochromator output, and computer. Divide the intensity detected on the output of monochromator by the intensity detected directly the tungsten halogen lamp. The result is the transmissivity function of the monochromator eliminating the influence of the halogen tungsten lamp radiation and the spectrometer response function. On a certain dial number, the transmissivity function is measured, normalized and the peak wavelength is calculated. 22 groups of wavelength data and dial number are acquired. 8, 14, 18, 22 points are chosen separately to calibrate the relationship between the wavelength and the dial number of monochromator by linear fitting and quadratic fitting. Experimental results show the calibration result in 22 points is better than others, (8 points similar to the calibration method using mercury lamp). The result of quadratic fitting with 22 points as the finalresult is better than linear fitting. So the calibration method for monochromator using the continuous spectrum light source with more calibration points is superior to the traditional calibration method using mercury lamp.
引用
收藏
页码:2348 / 2351
页数:4
相关论文
共 13 条
  • [1] Deng S. H., 2002, J INFRARED MILLIM W, V21, P133
  • [2] Fabio Frassetto, 2013, OPT EXPRESS, V21, P12996
  • [3] Absolute spectral responsivity measurements of solar cells by a hybrid optical technique
    Hamadani, Behrang H.
    Roller, John
    Dougherty, Brian
    Persaud, Fiona
    Yoon, Howard W.
    [J]. APPLIED OPTICS, 2013, 52 (21) : 5184 - 5193
  • [4] High-spectral-resolution characterization of broadband high-efficiency reflection gratings
    He, Kai
    Wang, Jianpeng
    Hou, Yongqiang
    Li, Xu
    Guan, Heyuan
    Kong, Fanyu
    Liu, Shijie
    Jin, Yunxia
    Yi, Kui
    [J]. APPLIED OPTICS, 2013, 52 (04) : 653 - 658
  • [5] Jin ZENG, 2012, ACTA OPT SINICA, V32(, P1
  • [6] Practical Implementation of the Mise en Pratique for the Definition of the Kelvin Above the Silver Point
    Machin, G.
    Bloembergen, P.
    Anhalt, K.
    Hartmann, J.
    Sadli, M.
    Saunders, P.
    Woolliams, E.
    Yamada, Y.
    Yoon, H.
    [J]. INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2010, 31 (8-9) : 1779 - 1788
  • [7] Measurement of normalized spectral responsivity of digital imaging devices by using a LED-based tunable uniform source
    Mahmoud, Khaled
    Park, Seongchong
    Park, Seung-Nam
    Lee, Dong-Hoon
    [J]. APPLIED OPTICS, 2013, 52 (06) : 1263 - 1271
  • [8] Monochromator-Based Absolute Calibration of a Standard Radiation Thermometer
    Mantilla, J. M.
    Hernanz, M. L.
    Campos, J.
    Martin, M. J.
    Pons, A.
    del Campo, D.
    [J]. INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2014, 35 (3-4) : 493 - 503
  • [10] Double-configuration grating monochromator for extreme-ultraviolet ultrafast pulses
    Poletto, L.
    Miotti, P.
    Frassetto, F.
    Spezzani, C.
    Grazioli, C.
    Coreno, M.
    Ressel, B.
    Gauthier, D.
    Ivanov, R.
    Ciavardini, A.
    de Simone, M.
    Stagira, S.
    De Ninno, G.
    [J]. APPLIED OPTICS, 2014, 53 (26) : 5879 - 5888