High-level disinfection of gastrointestinal endoscopes: Are current guidelines adequate?

被引:0
|
作者
Kovacs, BJ
Chen, YK
Kettering, JD
Aprecio, RM
Roy, I
机构
[1] Loma Linda Univ, Med Ctr, Div Gastroenterol, Dept Microbiol, Loma Linda, CA 92350 USA
[2] Loma Linda Univ, Med Ctr, Dept Hosp Epidemiol, Loma Linda, CA 92350 USA
来源
AMERICAN JOURNAL OF GASTROENTEROLOGY | 1999年 / 94卷 / 06期
关键词
D O I
暂无
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
OBJECTIVE: For a germicide to obtain a high level disinfection (HLD) claim, EDA requires demonstration of a 6-log reduction of mycobacterial inoculum under worst case conditions. The purpose of this study was to assess the adequacy of current guidelines for high level disinfection of GI endoscopes using alkaline glutaraldehyde in simulated-use testing. METHODS: Various gastrointestinal endoscopes were contaminated with Mycobacterium chelonae in 46 experiments. Quantitative cultures were obtained from each endoscope channel separately after each step: inoculation, standardized manual cleaning, immersion in 2% glutaraldehyde (Cidex) for 10, 20, or 45 min at room temperature, 70% isopropanol rinse, and drying. RESULTS: Manual cleaning alone achieved a 4-log reduction. After 10 min of glutaraldehyde exposure, but before alcohol rinse, two of 10 experiments failed to achieve a 6-log reduction. However, after alcohol rinse, all 10 experiments achieved HLD. After 20 min of glutaraldehyde exposure, but before alcohol rinse, one of 18 experiments failed to achieve a 6-log reduction. After alcohol rinse, all 18 experiments achieved HLD. After 45 min of glutaraldehyde exposure, but before alcohol rinse, one of 18 experiments failed to achieve a 6-log reduction. After alcohol rinse, all 18 experiments achieved HLD. Thus, if the entire reprocessing protocol including manual cleaning, glutaraldehyde exposure, alcohol rinse, and drying was taken into account, the required 6-log reduction of mycobacteria was achieved with a minimum of 10 min of glutaraldehyde exposure at room temperature. CONCLUSIONS: Current guidelines for high level disinfection using glutaraldehyde are appropriate. Alcohol rinse is a valuable adjunctive step for drying and for its bactericidal effects. (Am J Gastroenterol 1999;94:1546-1550. (C) 1999 by Am. Coll. of Gastroenterology).
引用
收藏
页码:1546 / 1550
页数:5
相关论文
共 50 条
  • [1] High-level disinfection of flexible GI endoscopes: Are current ASGE recommendations adequate?
    Kovacs, BJ
    Chen, YK
    Kettering, JD
    Roy, I
    Aprecio, RM
    Walter, MH
    Wong, W
    GASTROINTESTINAL ENDOSCOPY, 1997, 45 (04) : 103 - 103
  • [2] High-level disinfection or ''sterilization'' of endoscopes?
    Muscarella, LF
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 1996, 17 (03): : 183 - 187
  • [3] Recent Update on Microbiological Monitoring of Gastrointestinal Endoscopes after High-Level Disinfection
    Shin, Suk Pyo
    Kim, Won Hee
    CLINICAL ENDOSCOPY, 2015, 48 (05) : 369 - 373
  • [4] Mycobacteria and glutaraldehyde: Is high-level disinfection of endoscopes possible?
    Urayama, S
    Kozarek, RA
    Sumida, S
    Raltz, S
    Merriam, L
    Pethigal, P
    GASTROINTESTINAL ENDOSCOPY, 1996, 43 (05) : 451 - 456
  • [5] Bacteriologic testing of endoscopes after high-level disinfection
    Rejchrt, S
    Cermák, P
    Pavlatová, L
    Mícková, E
    Bures, J
    GASTROINTESTINAL ENDOSCOPY, 2004, 60 (01) : 76 - 78
  • [6] High-level disinfection (HLD) of GI endoscopes: Is there a problem?
    Foliente, RL
    Aprecio, RM
    Bains, HJ
    Kettering, JD
    Chen, YK
    GASTROINTESTINAL ENDOSCOPY, 1999, 49 (04) : AB141 - AB141
  • [7] High-level disinfection of flexible gastrointestinal endoscopes using a combination of glutaraldehyde and peracetic acid.
    Foliente, RL
    Aprecio, RM
    Kovacs, BJ
    Kettering, JD
    Roy, I
    Chen, YK
    GASTROINTESTINAL ENDOSCOPY, 1998, 47 (04) : AB49 - AB49
  • [8] Disinfection of endoscopes: Review of new chemical sterilants used for high-level disinfection
    Rutala, WA
    Weber, DJ
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 1999, 20 (01): : 69 - 76
  • [9] Guidelines for standardizing cleansing and disinfection of gastrointestinal endoscopes
    Iwakiri, Ryuichi
    Tanaka, Kiyohito
    Gotoda, Takuji
    Oka, Shiro
    Ohtsuka, Takao
    Sakata, Yasuhisa
    Chiba, Toshimi
    Higuchi, Kazuhide
    Masuyama, Hironori
    Nozaki, Ryoichi
    Matsuda, Koji
    Shimono, Nobuyuki
    Fujimoto, Kazuma
    Tajiri, Hisao
    DIGESTIVE ENDOSCOPY, 2019, 31 (05) : 477 - 497
  • [10] Use of flexible endoscopes for NOTES: sterilization or high-level disinfection?
    Georg O. Spaun
    Trudie A. Goers
    Richard A. Pierce
    Maria A. Cassera
    Sandy Scovil
    Lee L. Swanstrom
    Surgical Endoscopy, 2010, 24 : 1581 - 1588