Quantification of typical antipsychotics in human plasma by ultra-high performance liquid chromatography tandem mass spectrometry for therapeutic drug monitoring

被引:32
|
作者
Gradinaru, Julieta [1 ]
Vullioud, Astrid [1 ]
Eap, Chin B. [1 ,2 ]
Ansermot, Nicolas [1 ]
机构
[1] Univ Lausanne Hosp, Hosp Cery, Ctr Psychiat Neurosci,Dept Psychiat, Unit Pharmacogenet & Clin Psychopharmacol, CH-1008 Prilly, Switzerland
[2] Univ Lausanne, Univ Geneva, Sch Pharmaceut Sci, Geneva, Switzerland
关键词
Antipsychotics; Protein precipitation; UHPLC-MS/MS; Method validation; Therapeutic drug monitoring; SOLID-PHASE EXTRACTION; ELECTROSPRAY-IONIZATION; GAS-CHROMATOGRAPHY; HUMAN SERUM; QUANTITATIVE-DETERMINATION; CHEMICAL-IONIZATION; PSYCHOTROPIC-DRUGS; HALOPERIDOL; METABOLITES; ZUCLOPENTHIXOL;
D O I
10.1016/j.jpba.2013.07.041
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A selective and sensitive method was developed for the simultaneous quantification of seven typical antipsychotic drugs (cis-chlorprothixene, flupentixol, haloperidol, levomepromazine, pipamperone, promazine and zuclopenthixol) in human plasma. Ultra-high performance liquid chromatography (UHPLC) was used for complete separation of the compounds in less than 4.5 min on an Acquity UPLC BEH C18 column (2.1 mm x 50 mm; 1.7 mu m) with a gradient elution of ammonium formate buffer pH 4.0 and acetonitrile at a flow rate of 400 mu l/min. Detection was performed on a tandem quadrupole mass spectrometer (MS/MS) equipped with an electrospray ionization interface. A simple protein precipitation procedure with acetonitrile was used for sample preparation. Thanks to the use of stable isotope-labeled internal standards for all analytes, internal standard-normalized matrix effects were in the range of 92-108%. The method was fully validated to cover large concentration ranges of 0.2-90 ng/ml for haloperidol, 0.5-90 ng/ml for flupentixol, 1-450 ng/ml for levomepromazine, promazine and zuclopenthixol and 2-900 ng/ml for cis-chlorprothixene and pipamperone. Trueness (89.1-114.8%), repeatability (1.8-9.9%), intermediate precision (1.9-16.3%) and accuracy profiles (<30%) were in accordance with the latest international recommendations. The method was successfully used in our laboratory for routine quantification of more than 500 patient plasma samples for therapeutic drug monitoring. To the best of our knowledge, this is the first UHPLC-MS/MS method for the quantification of the studied drugs with a sample preparation based on protein precipitation. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:36 / 44
页数:9
相关论文
共 50 条
  • [1] Fast quantification of ten psychotropic drugs and metabolites in human plasma by ultra-high performance liquid chromatography tandem mass spectrometry for therapeutic drug monitoring
    Ansermot, Nicolas
    Brawand-Amey, Marlyse
    Kottelat, Astrid
    Eap, Chin B.
    JOURNAL OF CHROMATOGRAPHY A, 2013, 1292 : 160 - 172
  • [2] Quantification of α-ketoglutarate cyanohydrin in swine plasma by ultra-high performance liquid chromatography tandem mass spectrometry
    Mitchell, Brendan L.
    Rockwood, Gary A.
    Logue, Brian A.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2013, 934 : 60 - 65
  • [3] Rapid and simple quantification of belimumab in human plasma using ultra-high performance liquid chromatography with tandem mass spectrometry
    Yoshijima, Chisato
    Suzuki, Yosuke
    Tanaka, Ryota
    Ono, Hiroyuki
    Oda, Ayako
    Ozaki, Takashi
    Shibata, Hirotaka
    Itoh, Hiroki
    Ohno, Keiko
    CLINICAL BIOCHEMISTRY, 2024, 124
  • [4] Development and validation of an ultra-high performance liquid chromatography-tandem mass spectrometry method to quantify antiretroviral drug concentrations in human plasma for therapeutic monitoring
    West III, Raymond E.
    Oberly, Patrick J.
    Riddler, Sharon A.
    Nolin, Thomas D.
    Devanathan, Aaron S.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2024, 240
  • [5] Buprenorphine and norbuprenorphine quantification in human plasma by simple protein precipitation and ultra-high performance liquid chromatography tandem mass spectrometry
    Luethi, Guillaume
    Blangy, Valeria
    Eap, Chin B.
    Ansermot, Nicolas
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2013, 77 : 1 - 8
  • [6] Quantification of levetiracetam in human plasma by liquid chromatography-tandem mass spectrometry: Application to therapeutic drug monitoring
    Matar, Kamal M.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2008, 48 (03) : 822 - 828
  • [7] Determination of unbound piperaquine in human plasma by ultra-high performance liquid chromatography tandem mass spectrometry
    Huang, Liusheng
    Sok, Vong
    Aslam-Mir, Usman
    Marzan, Florence
    Whalen, Meghan
    Rosenthal, Philip J.
    Aweeka, Francesca
    JOURNAL OF CHROMATOGRAPHY OPEN, 2022, 2
  • [8] Simultaneous quantification of five antiretrovirals in human tissues using ultra-high performance liquid chromatography-tandem mass spectrometry methods for therapeutic drug monitoring at the sites of action
    West III, Raymond E.
    Oberly, Patrick J.
    Saylor, Amanda J.
    Riddler, Sharon A.
    Nolin, Thomas D.
    Devanathan, Aaron S.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2024, 1241
  • [9] Ultra-high throughput dual channel liquid chromatography with tandem mass spectrometry for quantification of four immunosuppressants in whole blood for therapeutic drug monitoring
    Zijp, Tanja R.
    van Hateren, Kai
    Kuiper, Hiltjo
    Jongedijk, Erwin M.
    Touw, Daan J.
    JOURNAL OF CHROMATOGRAPHY A, 2023, 1702
  • [10] Quantification of plasma phospholipids by ultra performance liquid chromatography tandem mass spectrometry
    Rabagny, Yannick
    Herrmann, Wolfgang
    Geisel, Juergen
    Kirsch, Susanne H.
    Obeid, Rima
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 401 (03) : 891 - 899