Low-uncertainty absolute radiometric calibration of a CCD

被引:19
|
作者
Ferrero, A [1 ]
Campos, J [1 ]
Pons, A [1 ]
机构
[1] CSIC, Inst Fis Aplicada, Madrid 28006, Spain
关键词
D O I
10.1088/0026-1394/43/2/S04
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This work presents an experimental method for the low-uncertainty calibration of the spectral radiant exposure responsivity of a CCD detector. It comprises a description of the experimental set-up as well as an analysis of the various sources of uncertainty. The overall result shows that the calibration procedure-related uncertainty is only 0.18% as long as the temperature is kept constant and the field of view is kept lower than 13 degrees.
引用
收藏
页码:S17 / S21
页数:5
相关论文
共 50 条
  • [1] Spatial characterization of cameras for low-uncertainty radiometric measurements
    Ferrero, A.
    Lopez, M.
    Campos, J.
    Sperling, A.
    METROLOGIA, 2014, 51 (03) : 316 - 325
  • [2] Absolute Radiometric Calibration and Uncertainty Validation for a CubeSat Thermal Infrared Radiometric Imager in Low Earth Orbit
    Osterman, David P.
    Amparan, Alfonso B.
    Kampe, Thomas U.
    Kerrigan, Patrick W.
    Mcpherson, Brandi N.
    Piqueira, David M.
    Reavis, Gretchen W.
    Veto, Michael S.
    Warden, Robert M.
    Willett, Kevin C.
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2024, 62
  • [3] Low-uncertainty calibration of accelerometers below 1.0 Hz: Practical considerations
    Peterson, Edward L.
    Rusen, William M.
    Morrison, John C.
    Test Engineering and Management, 2004, 66 (02): : 10 - 12
  • [4] Experimental Demonstration of Low-Uncertainty Calibration Methods for Bragg Grating Interrogators
    Luis de Miguel, Jose
    Galindo-Santos, Juan
    Pulido de Torres, Concepcion
    Salgado, Pedro
    Velasco, Aitor V.
    Corredera, Pedro
    SENSORS, 2018, 18 (06)
  • [5] Uncertainty analysis of the absolute radiometric calibration of full waveform airborne LiDAR
    Lu H.
    Pang Y.
    Li Z.
    Wang D.
    Chen B.
    Ma Z.
    Yaogan Xuebao/Journal of Remote Sensing, 2020, 24 (11): : 1353 - 1362
  • [6] Uncertainty budget for detector-based absolute radiometric calibration with GLAMR
    Wang, Zhipeng
    Barsi, Julia
    Thome, Kurtis
    Wenny, Brian N.
    Mcandrew, Brendan
    Efremova, Boryana
    Mccorkel, Joel
    APPLIED OPTICS, 2024, 63 (12) : 3015 - 3028
  • [7] VALIDATION FOR THE ABSOLUTE RADIOMETRIC CALIBRATION OF THE HJ-1B CCD SENSORS OF CHINA
    Jiang, Hongbo
    Qin, Qiming
    Li, Jun
    Zhao, Shaohua
    Dong, Heng
    Yuan, Weilin
    Cui, Rongbo
    2010 IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, 2010, : 2876 - 2879
  • [8] A low-uncertainty measurement of the Boltzmann constant
    de Podesta, Michael
    Underwood, Robin
    Sutton, Gavin
    Morantz, Paul
    Harris, Peter
    Mark, Darren F.
    Stuart, Finlay M.
    Vargha, Gergely
    Machin, Graham
    METROLOGIA, 2013, 50 (04) : 354 - 376
  • [9] Comment on 'A low-uncertainty measurement of the Boltzmann constant'
    Macnaughton, Donald B.
    METROLOGIA, 2016, 53 (01) : 108 - 115
  • [10] Hybrid data fusion strategy for the low-uncertainty 3D calibration of cylinder standards
    Liu, Pei
    Jusko, Otto
    Tutsch, Rainer
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2017, 28 (06)