Evaluation of the leukocyte differential on a new automated flow cytometry hematology analyzer

被引:9
|
作者
Kim, J. -E. [1 ]
Kim, B. -R. [1 ]
Woo, K. -S. [1 ]
Han, J. -Y. [1 ]
机构
[1] Dong A Univ, Dept Lab Med, Coll Med, Pusan 602715, South Korea
关键词
Blood cell analyzer; leukocyte differential; blast; cell lineage; BLOOD;
D O I
10.1111/j.1751-553X.2012.01432.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The Hematoflow (Beckman Coulter, USA) is a new automated hematology analyzer, which provides a 16-part white blood cell count (WBC) differential. Methods: We evaluated the differential WBC count performance of the Hematoflow. 101 blood samples from patients were selected for comparison analysis. Results: The methodological comparison of the WBC differential parameters of neutrophils, lymphocytes, monocytes and eosinophils showed good correlations among 4 different analyses. More than 1% of blast cells were counted in 30 of 101 samples. A good correlation for blast cell counts obtained by Hematoflow was found with the reference manual method (r=0.9637, P 0.0001). For blast B, Hematoflow shows good correlation with reference method results but did not identify blast T. Conclusions: These results demonstrate that the Hematoflow has a comparable performance with the Sysmex XE-2100 and indicate that B cell lineage ALL can be identified by the use of the Hematoflow in an initial evaluation of acute leukemia.
引用
收藏
页码:547 / 550
页数:4
相关论文
共 50 条
  • [1] Evaluation and comparison of automated hematology analyzer, flow cytometry, and digital morphology analyzer for monocyte counting
    Buoro, S.
    Moioli, V.
    Seghezzi, M.
    Previtali, G.
    Alessio, M. G.
    Lopez, R. Simon
    Ortolani, C.
    Ottomano, C.
    Lippi, G.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (05) : 577 - 585
  • [2] Performance evaluation of leukocyte differential on the hematology analyzer Celltac G compared with two hematology analyzers, reference flow cytometry method, and two manual methods
    Yamade, Kenji
    Yamaguchi, Toshihiro
    Nagai, Yutaka
    Kamisako, Toshinori
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2021, 35 (08)
  • [3] Use of an automated hematology analyzer and flow cytometry to assess bone marrow cellularity and differential cell count
    Kim, N
    Kim, J
    Lim, J
    Kim, Y
    Han, K
    Kang, CS
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2004, 34 (03): : 307 - 313
  • [4] EVALUATION OF A NEW 5 DIFFERENTIAL COMPACT AUTOMATED HEMATOLOGY ANALYZER: ACCURACY & ERGONOMICS
    Czyzewska, Emilia
    Wisniewska, Agnieszka
    Blocka-Gumowska, Monika
    Hetmanczuk, Jakub
    Bobilewicz, Dagna
    Duroux, Christophe
    Pastore, Manuela
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 29 - 29
  • [5] LEUKOCYTE DIFFERENTIATION BY FLOW CYTOMETRY: COMPARISON OF TWO FLOW CYTOMETRIC PROTOCOLS, LEUKOFLOW AND CYTODIFF™, WITH MICROSCOPY AND HEMATOLOGY ANALYZER.
    van de Geijn, G. J. M.
    van Dijk, M.
    Meijer, R.
    Blaauw, Y.
    Bakker, R.
    Zwarteveld, Y.
    van der Heide, E.
    Janssen, J. W. J.
    Beunis, M. H.
    Njo, T. L.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 59 - 60
  • [6] Performance evaluation the new automated Atellica Hema 580 hematology analyzer
    Schapkaitz, E.
    Baiden, A.
    Raburabu, S.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 (01) : 63 - 71
  • [7] Evaluation of performance parameters of an automated hematology analyzer
    Pares, Jesica
    Borda, Natalia
    Damian Santiago, Sergio
    Benito, Cristina
    Aranda, Claudio
    ACTA BIOQUIMICA CLINICA LATINOAMERICANA, 2015, 49 (04): : 399 - 407
  • [8] CLINICAL EVALUATION OF LEUKOCYTE DIFFERENTIAL COUNT IN PERIPHERAL BLOOD BY FLOW CYTOMETRY
    Guo, Ping
    Wang, Jianbiao
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 42 - 42
  • [9] IMPROVED AUTOMATED LEUKOCYTE COUNTING AND DIFFERENTIAL IN NEWBORNS ACHIEVED BY THE HEMATOLOGY ANALYZER CELL-DYN 3500(TM)
    DORNER, K
    SCHULZE, S
    REINHARDT, M
    SEEGER, H
    VANHOVE, L
    CLINICAL AND LABORATORY HAEMATOLOGY, 1995, 17 (01): : 23 - 30
  • [10] Preliminary evaluation of a new flow cytometry method for the routine hematology workflow
    Seghezzi, Michela
    Moioli, Valentina
    Previtali, Giulia
    Manenti, Barbara
    Lopez, Ramon-Simon
    Kono, Mari
    Tirloni, Ezio
    Alessio, Maria Grazia
    Buoro, Sabrina
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2019, 57 (10) : 1608 - 1622