Magnetic Resonance-Based Motion Correction for Positron Emission Tomography Imaging
被引:76
|作者:
Ouyang, Jinsong
[1
]
Li, Quanzheng
[1
]
El Fakhri, Georges
[1
]
机构:
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Ctr Adv Radiol Sci,Dept Radiol, Boston, MA 02114 USA
关键词:
RESPIRATORY GATED PET;
ATTENUATION CORRECTION;
HEAD MOVEMENT;
RECONSTRUCTION;
MRI;
REGISTRATION;
IMPACT;
MODEL;
TRANSFORMATION;
COMPENSATION;
D O I:
10.1053/j.semnuclmed.2012.08.007
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Positron emission tomography (PET) image quality is limited by patient motion. Emission data are blurred owing to cardiac and/or respiratory motion. Although spatial resolution is 4 mm for standard clinical whole-body PET scanners, the effective resolution can be as low as 1 cm owing to motion. Additionally, the deformation of attenuation medium causes image artifacts. Previously, gating has been used to "freeze" the motion, but led to significantly increased noise level. Simultaneous PET/magnetic resonance (MR) modality offers a new way to perform PET motion correction. MR can be used to measure 3-dimensional motion fields, which can then be incorporated into the iterative PET reconstruction to obtain motion-corrected PET images. In this report, we present MR imaging techniques to acquire dynamic images, a nonrigid image registration algorithm to extract motion fields from acquired MR images, and a PET reconstruction algorithm with motion correction. We also present results from both phantom and in vivo animal PET/MR studies. We demonstrate that MR-based PET motion correction using simultaneous PET/MR improves image quality and lesion detectability compared with gating and no motion correction. Semin Nucl Med 43:60-67 (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:60 / 67
页数:8
相关论文