Cascaded systems analysis of charge sharing in cadmium telluride photon-counting x-ray detectors

被引:23
|
作者
Tanguay, Jesse [1 ]
Cunningham, Ian A. [2 ,3 ,4 ]
机构
[1] Univ British Columbia Okanagan, Dept Phys, Kelowna, BC V1Y 1V7, Canada
[2] Univ Western Ontario, Imaging Res Labs, Robarts Res Inst, London, ON N6A 5B7, Canada
[3] Univ Western Ontario, Schulich Sch Med & Dent, Dept Med Biophys, London, ON N6A 5C1, Canada
[4] Univ Western Ontario, Schulich Sch Med & Dent, Biomed Engn, London, ON N6A 5C1, Canada
关键词
cadmium telluride; cascaded systems analysis; photoncounting; spatial resolution; spectroscopic x-ray detectors; x-ray detectors; TASK-BASED DETECTABILITY; MEDICAL IMAGING-SYSTEMS; CONE-BEAM CT; COMPUTED-TOMOGRAPHY; QUANTUM EFFICIENCY; NOISE TRANSFER; SPECTRAL MAMMOGRAPHY; READOUT CHIP; PULSE PILEUP; RESOLVED CT;
D O I
10.1002/mp.12853
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeSingle-photon-counting (SPC) and spectroscopic x-ray detectors are under development in academic and industry laboratories for medical imaging applications. The spatial resolution of SPC and spectroscopic x-ray detectors is an important design criterion. The purpose of this article was to extend the cascaded systems approach to include a description of the spatial resolution of SPC and spectroscopic x-ray imaging detectors. MethodsA cascaded systems approach was used to model reabsorption of characteristic x rays, Coulomb repulsion, and diffusion in SPC and spectroscopic x-ray detectors. In addition to reabsorption, diffusion, and Coulomb repulsion, the model accounted for x-ray conversion to electron-hole (e-h) pairs, integration of e-h pairs in detector elements, electronic noise, and energy thresholding. The probability density function (PDF) describing the number of e-h pairs was propagated through each stage of the model and was used to derive new theoretical expressions for the large-area gain and modulation transfer function (MTF) of CdTe SPC x-ray detectors, and the energy bin sensitivity functions and MTFs of CdTe spectroscopic detectors. Theoretical predictions were compared with the results of MATLAB-based Monte Carlo (MC) simulations and published data. Comparisons were also made with the MTF of energy-integrating systems. ResultsUnder general radiographic conditions, reabsorption, diffusion, and Coulomb repulsion together artificially inflate count rates by 20% to 50%. For thicker converters (e.g. 1000 m) and larger detector elements (e.g. 500 m pixel pitch) these processes result in modest inflation (i.e. approximate to 10%) in apparent count rates. Our theoretical and MC analyses predict that SPC MTFs will be degraded relative to those of energy-integrating systems for fluoroscopic, general radiographic, and CT imaging conditions. In most cases, this degradation is modest (i.e., approximate to 10% at the Nyquist frequency). However, for thicker converters, the SPC MTF can be degraded by up to 25% at the Nyquist frequency relative to EI systems. Additionally, unlike EI systems, the MTF of spectroscopic systems is strongly dependent on photon energy, which results in energy-bin-dependent spatial resolution in spectroscopic systems. ConclusionsThe PDF-transfer approach to modeling signal transfer through SPC and spectroscopic x-ray imaging systems provides a framework for understanding system performance. Application of this approach demonstrated that charge sharing artificially inflates the SPC image signal and degrades the MTF of SPC and spectroscopic systems relative to energy-integrating systems. These results further motivate the need for anticharge-sharing circuits to mitigate the effects of charge sharing on SPC and spectroscopic x-ray image quality.
引用
收藏
页码:1926 / 1941
页数:16
相关论文
共 50 条
  • [1] Cascaded systems analyses of photon-counting x-ray detectors
    Tanguay, Jesse
    Yun, Seungman
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [2] Evaluation of Energy Loss and Charge Sharing in Cadmium Telluride Detectors for Photon-Counting Computed Tomography
    Xu, Cheng
    Danielsson, Mats
    Bornefalk, Hans
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (03) : 614 - 625
  • [3] The detective quantum efficiency of cadmium telluride photon-counting x-ray detectors in breast imaging applications
    Day, James A.
    Tanguay, Jesse
    MEDICAL PHYSICS, 2022, 49 (03) : 1481 - 1494
  • [4] Photon-counting x-ray detectors for CT
    Danielsson, Mats
    Persson, Mats
    Sjolin, Martin
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03):
  • [5] The detective quantum efficiency of photon-counting x-ray detectors using cascaded-systems analyses
    Tanguay, Jesse
    Yun, Seungman
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL PHYSICS, 2013, 40 (04)
  • [6] Extension of cascaded systems analysis to single-photon-counting x-ray detectors
    Tanguay, Jesse
    Yun, Seungman
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [7] Monte-Carlo Simulation of the Detective Quantum Efficiency of Cadmium Telluride X-Ray Detectors for Photon-Counting Mammography
    Day, J.
    Tanguay, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [8] Monte Carlo simulation of charge sharing effects in silicon and GaAs photon-counting X-ray imaging detectors
    Nilsson, HE
    Fröjdh, C
    Dubaric, E
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (04) : 1636 - 1640
  • [9] Monte-Carlo study of contrast-enhanced spectral mammography with cadmium telluride photon-counting x-ray detectors
    Day, James A.
    Tanguay, Jesse
    MEDICAL PHYSICS, 2024, 51 (04) : 2479 - 2498
  • [10] Detective quantum efficiency of photon-counting x-ray detectors
    Tanguay, Jesse
    Yun, Seungman
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL PHYSICS, 2015, 42 (01) : 491 - 509