Scout View X-Ray Attenuation Versus Weight-Based Selection of Reduced Peak Tube Voltage in Cardiac CT Angiography

被引:20
|
作者
Ghafourian, Kambiz [2 ]
Younes, Desiree
Simprini, Lauren A. [2 ]
Weigold, Wm Guy [2 ]
Weissman, Gaby [2 ]
Taylor, Allen J. [1 ]
机构
[1] Georgetown Univ Hosp, Dept Med, Div Cardiol, Washington, DC 20007 USA
[2] Washington Hosp Ctr, Div Cardiol, Washington, DC 20010 USA
关键词
cardiac computed tomographic angiography; radiation exposure; AMERICAN-HEART-ASSOCIATION; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; IMAGE QUALITY; CARDIOVASCULAR RADIOLOGY; CLINICAL CARDIOLOGY; INTERVENTION; COMMITTEE; COUNCIL;
D O I
10.1016/j.jcmg.2011.12.022
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The study evaluated the relationship between cardiac computed tomography (CT) scout view x-ray attenuation and CT image noise compared with weight or body mass index (BMI). BACKGROUND Decreasing peak tube voltage from 120 to 100 kVp on the basis of body size reduces radiation exposure. Methods to better predict CT image noise may lead to more effective selection of reduced tube voltage in cardiac CT. METHODS Image quality was graded subjectively (1 [excellent] to 4 [nondiagnostic]) and objectively (SD of the aortic attenuation value) in cardiac CT angiograms (N = 106) acquired at either 100 or 120 kVp. X-ray attenuation characteristics on the scout view (120 kVp, 30 mA) were measured within a 3-cm region of interest across the chest in the frontal x-ray. Receiver-operating characteristic curve analysis was performed comparing scout view attenuation versus weight and BMI in predicting CT image noise and quality. RESULTS CT image noise correlated with both BMI (r = 0.40; p < 0.001) and the scout view attenuation value (r = 0.52; p < 0.001). In linear regression models with controlling for BMI (or weight) and tube potential, scout view attenuation was the best predictor of the CT image noise (p < 0.001), and increased model fit statistic from 0.23 to 0.41 (p model <0.001). At 120 kVp, scout view attenuation predicted CT image noise <30 Hounsfield units (HU) more accurately than BMI (area under the curve: 0.89 vs. 0.77). For CT images acquired at 120 kVp, those with a scout view attenuation <-120 HU had significantly lower noise and higher signal-to-noise ratios, with similar mean aortic attenuation values. A majority (89.3%) of "low-noise" CT images at 120 kVp had scout view attenuation values of <-120 HU. CONCLUSIONS Scout view attenuation predicts cardiac CT image noise better than weight or BMI and could enable broader application of reduced x-ray tube voltage as a radiation sparing technique. (J Am Coll Cardiol Img 2012; 5: 589-95) (C) 2012 by the American College of Cardiology Foundation
引用
收藏
页码:589 / 595
页数:7
相关论文
共 13 条
  • [1] SCOUT FILM ATTENUATION AS A GUIDE FOR CARDIAC CT RADIATION SPARING USING REDUCED TUBE VOLTAGE
    Ghafourian, Kambiz
    Younes, Desiree
    Simprini, Lauren A.
    Weigold, Wm Guy
    Weissman, Gaby
    Taylor, Allen J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (14) : E873 - E873
  • [2] Patient-specific tube-voltage selection at coronary CT angiography based on the combination of X-ray attenuation on scout views and body mass index: how can appropriate radiation dose be achieved?
    Noto, Daichi
    Funama, Yoshinori
    Utsunomiya, Daisuke
    Oda, Seitaro
    Yuki, Hideaki
    Yamashita, Yasuyuki
    ACTA RADIOLOGICA, 2015, 56 (10) : 1171 - 1179
  • [3] Feasibility of Weight-based Tube Voltage and Iodine Delivery Rate for Coronary Artery CT Angiography
    Wang, Ying
    Zhang, Yan
    Di, Aihui
    Wang, Qizheng
    Chen, Yongye
    Yuan, Huishu
    Lang, Ning
    CURRENT MEDICAL IMAGING, 2024, 20
  • [4] Automated attenuation-based selection of tube voltage and tube current for coronary CT angiography: Reduction of radiation exposure versus a BMI-based strategy with an expert investigator
    Layritz, Christian
    Muschiol, Gerd
    Flohr, Thomas
    Bietau, Christian
    Marwan, Mohamed
    Schuhbaeck, Annika
    Schmid, Jasmin
    Ropers, Dieter
    Achenbach, Stephan
    Pflederer, Tobias
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2013, 7 (05) : 303 - 310
  • [5] Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison
    Krazinski, Aleksander W.
    Meinel, Felix G.
    Schoepf, U. Joseph
    Silverman, Justin R.
    Canstein, Christian
    De Cecco, Carlo N.
    Geyer, Lucas L.
    EUROPEAN RADIOLOGY, 2014, 24 (11) : 2677 - 2684
  • [6] Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison
    Aleksander W. Krazinski
    Felix G. Meinel
    U. Joseph Schoepf
    Justin R. Silverman
    Christian Canstein
    Carlo N. De Cecco
    Lucas L. Geyer
    European Radiology, 2014, 24 : 2677 - 2684
  • [7] Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol
    Albrecht, Moritz H.
    Nance, John W.
    Schoepf, U. Joseph
    Jacobs, Brian E.
    Bayer, Richard R., II
    Litwin, Sheldon E.
    Reynolds, Michael A.
    Otani, Katharina
    Mangold, Stefanie
    Varga-Szemes, Akos
    De Santis, Domenico
    Eid, Marwen
    Apfaltrer, Georg
    Tesche, Christian
    Goeller, Markus
    Vogl, Thomas J.
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2018, 28 (05) : 2134 - 2142
  • [8] Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol
    Moritz H. Albrecht
    John W. Nance
    U. Joseph Schoepf
    Brian E. Jacobs
    Richard R. Bayer
    Sheldon E. Litwin
    Michael A. Reynolds
    Katharina Otani
    Stefanie Mangold
    Akos Varga-Szemes
    Domenico De Santis
    Marwen Eid
    Georg Apfaltrer
    Christian Tesche
    Markus Goeller
    Thomas J. Vogl
    Carlo N. De Cecco
    European Radiology, 2018, 28 : 2134 - 2142
  • [9] Reducing Iodine Contrast Volume in CT Angiography of the Abdominal Aorta Using Integrated Tube Potential Selection and Weight-Based Method Without Compromising Image Quality
    Vasconcelos, Rogerio
    Vrtiska, Terri J.
    Foley, Thomas A.
    Macedo, Thanila A.
    Cardona, Juan C. Montoya
    Williamson, Eric E.
    McCollough, Cynthia H.
    Fletcher, Joel G.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2017, 208 (03) : 552 - 563
  • [10] Impact of X-ray tube settings and metallic leads on neurological PET imaging when using CT-based attenuation correction
    Ay, Mohammad Reza
    Zaidi, Habib
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 411 - 414