A High-Voltage Bipolar Transconductance Amplifier for Electrotactile Stimulation

被引:13
|
作者
Schaning, Matthew A.
Kaczmarek, Kurt A. [1 ,2 ]
机构
[1] Univ Wisconsin, Dept Orthoped & Rehabil Med, Madison, WI 53706 USA
[2] Univ Wisconsin, Dept Biomed Engn, Madison, WI 53706 USA
关键词
Current control; electrocutaneous; electrotactile; transconductance amplifier (TA);
D O I
10.1109/TBME.2008.926675
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes a high-performance transconductance amplifier specifically designed for electrotactile (electrocutaneous) stimulation. It enables voltages up to +/- 600 V to be produced at the output that will allow the psychophysiological performance associated with stimulation of the fingertip using various stimulation waveforms to be studied more thoroughly. The design has a transconductance of up to 20 mA/V, an 8.8-M Omega output resistance, and can provide output currents up to +/- 20 mA. A complete schematic diagram is presented along with a discussion of theory of operation and safety issues as well as performance and derating plots from the implemented design.