Flow Patterns at Stented Coronary Bifurcations Computational Fluid Dynamics Analysis

被引:58
|
作者
Katritsis, Demosthenes G. [1 ]
Theodorakakos, Andreas [2 ]
Pantos, Ioannis [1 ,3 ]
Gavaises, Manolis [4 ]
Karcanias, Nicos [4 ]
Efstathopoulos, Efstathios P. [3 ]
机构
[1] Athens Euroclin, Dept Cardiol, Athens 11521, Greece
[2] Fluid Res, Athens, Greece
[3] Univ Athens, Dept Radiol, GR-10679 Athens, Greece
[4] City Univ London, Sch Engn & Math Sci, London EC1V 0HB, England
关键词
angioplasty; stents; hemodynamics; bifurcation; coronary circulation; WALL SHEAR-STRESS; PERCUTANEOUS CORONARY; ELUTING STENTS; PULSATILE FLOW; LESIONS; ARTERY; IMPLANTATION; ATHEROSCLEROSIS; HEMODYNAMICS; PREDICTORS;
D O I
10.1161/CIRCINTERVENTIONS.112.968347
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-The ideal bifurcation stenting technique is not established, and data on the hemodynamic characteristics at stented bifurcations are limited. Methods and Results-We used computational fluid dynamics analysis to assess hemodynamic parameters known affect the risk of restenosis and thrombosis at coronary bifurcations after the use of various single-and double-stenting techniques. We assessed the distributions and surface integrals of the time averaged wall shear stress (TAWSS), oscillatory shear index (OSI), and relative residence time (t(r)). Single main branch stenting without side branch balloon angioplasty or stenting provided the most favorable hemodynamic results (integrated values of TAWSS=4.13.10(-4) N, OSI=7.52.10(-6) m(2), t(r)=5.57.10(-4) m(2)/Pa) with bifurcational area subjected to OSI values >0.25, >0.35, and >0.45 calculated as 0.36 mm(2), 0.04 mm(2), and 0 mm(2), respectively. Extended bifurcation areas subjected to these OSI values were seen after T-stenting: 0.61 mm(2), 0.18 mm(2), and 0.02 mm(2), respectively. Among the considered double-stenting techniques, crush stenting (integrated values of TAWSS=1.18.10(-4) N, OSI=7.75.10(-6) m(2), t(r)=6.16.10(-4) m(2)/Pa) gave the most favorable results compared with T-stenting (TAWSS=0.78.10(-4) N, OSI=10.40.10(-6) m(2), t(r)=6.87.10(-4) m(2)/Pa) or the culotte technique (TAWSS=1.30.10(-4) N, OSI=9.87.10(-6) m(2), t(r)=8.78.10(-4) m(2)/Pa). Conclusions-In the studied models of computer simulations, stenting of the main branch with our without balloon angioplasty of the side branch offers hemodynamic advantages over double stenting. When double stenting is considered, the crush technique with the use of a thin-strut stent may result in improved immediate hemodynamics compared with culotte or T-stenting. (Circ Cardiovasc Interv. 2012; 5:530-539.)
引用
收藏
页码:530 / 539
页数:10
相关论文
共 50 条
  • [1] Computational fluid dynamics of stented coronary bifurcations studied with a hybrid discretization method
    Chiastra, Claudio
    Morlacchi, Stefano
    Pereira, Simon
    Dubini, Gabriele
    Migliavacca, Francesco
    EUROPEAN JOURNAL OF MECHANICS B-FLUIDS, 2012, 35 : 76 - 84
  • [2] Computational Fluid Dynamics Analysis of Blood Flow through Stented Arteries
    Mohammad, Shafiullah
    Majumdar, Pradip
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2013, VOL 3A, 2014,
  • [3] Blood flow in stented coronary artery :: numerical fluid dynamics analysis
    Bénard, N
    Perrault, R
    Coisne, D
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 3800 - 3803
  • [4] Local blood flow patterns in stented coronary bifurcations: an experimental and numerical study
    Raben, Jaime S.
    Morlacchi, Stefano
    Burzotta, Francesco
    Migliavacca, Francesco
    Vlachos, Pavlos P.
    JOURNAL OF APPLIED BIOMATERIALS & FUNCTIONAL MATERIALS, 2015, 13 (02) : E116 - E126
  • [5] Local Flow Patterns After Implantation of Bioresorbable Vascular Scaffold in Coronary Bifurcations - Novel Findings by Computational Fluid Dynamics -
    Li, Yingguang
    Li, Zehang
    Holck, Emil N.
    Xu, Bo
    Karanasos, Antonios
    Fei, Zhenyu
    Chang, Yunxiao
    Chu, Miao
    Dijkstra, Jouke
    Christiansen, Evald H.
    Reiber, Johan H. C.
    Holm, Niels R.
    Tu, Shengxian
    CIRCULATION JOURNAL, 2018, 82 (06) : 1575 - +
  • [6] NON-NEWTONIAN EFFECTS ON BLOOD FLOW DYNAMICS IN STENTED CORONARY ARTERY BIFURCATIONS
    Zarandi, Marjan Molavi
    Mongrain, Rosaire
    Bertrand, Olivier F.
    Bonakdar, Ali
    PROCEEDINGS OF THE ASME FLUIDS ENGINEERING DIVISION SUMMER CONFERENCE - 2010 - VOL 1, PTS A-C, 2010, : 1889 - 1893
  • [7] EXPERIMENTAL INVESTIGATION OF THE LOCAL BLOOD FLOW PATTERN IN STENTED CORONARY BIFURCATIONS
    Morlacchi, Stefano
    Schmieg, Jaime
    Cooper, Dan
    Burzotta, Francesco
    Migliavacca, Francesco
    Vlachos, Pavlos P.
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE 2011, PTS A AND B, 2011, : 165 - 166
  • [8] Analysis of wall shear stress in stented coronary artery using 3D computational fluid dynamics modeling
    Dehlaghi, Vahab
    Shadpoor, Mohammad Tafazoli
    Najarian, Siamak
    JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2008, 197 (1-3) : 174 - 181
  • [9] Computational Fluid Dynamics Analysis of Nasal Flow
    Moesges, R.
    Buechner, B.
    Kleiner, M.
    Freitas, R.
    Hoerschler, I.
    Schroeder, W.
    B-ENT, 2010, 6 (03) : 161 - 165
  • [10] Computational Fluid Dynamics Analysis of Coronary Stent Malapposition
    Hudrea, Iuliana-Claudia
    Totorean, Alin-Florin
    Gaita, Dan
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1, 2019, 68 (01): : 761 - 765