Motor feedback physiological control for a continuous flow ventricular assist device

被引:0
|
作者
Waters, T
Allaire, P [1 ]
Tao, G
Adams, M
Bearnson, G
Wei, N
Hilton, E
Baloh, M
Olsen, D
Khanwilkar, P
机构
[1] Univ Virginia, Dept Elect Engn, Charlottesville, VA 22901 USA
[2] Univ Virginia, Dept Mech Aerosp & Nucl Engn, Charlottesville, VA 22901 USA
[3] Univ Virginia, Dept Biomed Engn, Charlottesville, VA 22901 USA
[4] Univ Utah, Artificial Heart Lab, Salt Lake City, UT 84112 USA
[5] Medquest Prod Inc, Salt Lake City, UT USA
关键词
ventricular assist device; pump performance; automatic feedback controller;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The response of a continuous flow magnetic bearing supported ventricular assist device, the CFVAD3 (CF3) to human physiologic pressure and flow needs is varied by adjustment of the motor speed. This paper discusses a model of the automatic feedback controller designed to develop the required pump performance. The major human circulatory, mechanical, and electrical systems were evaluated using experimental data from the CF3 and linearized models developed. An open-loop model of the human circulatory system was constructed with a human heart and a VAD included. A feedback loop was then closed to maintain a desired reference differential pressure across the system. A proportional-integral (PI) controller was developed to adjust the motor speed and maintain the system reference differential pressure when changes occur in the natural heart. The effects of natural heart pulsatility on the control system show that the reference blood differential pressure is maintained without requiring CF3 motor pulsatility.
引用
收藏
页码:480 / 486
页数:7
相关论文
共 50 条
  • [1] A multi-objective physiological control system for continuous-flow left ventricular assist device with non-invasive physiological feedback
    Liu, Hongtao
    Liu, Shuqin
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2025, 100
  • [2] Blood flow in a continuous flow ventricular assist device
    Allaire, PE
    Wood, HG
    Awad, RS
    Olsen, DB
    ARTIFICIAL ORGANS, 1999, 23 (08) : 769 - 773
  • [3] Improving arterial pulsatility by feedback control of a continuous flow left ventricular assist device via in silico modeling
    Bozkurt, Selim
    van de Vosse, Frans N.
    Rutten, Marcel C. M.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2014, 37 (10): : 773 - 785
  • [4] Physiological control of rotary left ventricular assist device
    Yi, Wu
    PROCEEDINGS OF THE 26TH CHINESE CONTROL CONFERENCE, VOL 6, 2007, : 469 - 474
  • [5] Continuous-flow ventricular assist device - Reply
    Miller, Leslie W.
    Pagani, Francis D.
    NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22): : 2305 - 2306
  • [6] VELOCITY CONTROL OF MOTOR DRIVEN VENTRICULAR ASSIST DEVICE
    GREATHOUSE, S
    LEWIS, J
    OSTROFF, A
    GESELOWITZ, DB
    ROSENBERG, G
    PIERCE, WS
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (08) : 532 - 532
  • [7] Impact of perioperative right ventricular assist device on ventricular arrhythmias in patients with continuous flow left ventricular assist device
    Dhawan, R.
    Selim, A.
    Jhand, A.
    Lundgren, S.
    Zolty, R.
    Khan, F.
    EUROPEAN HEART JOURNAL, 2020, 41 : 1096 - 1096
  • [8] SYNCHRONIZED PULSATILE FLOW WITH A CONTINUOUS FLOW LEFT VENTRICULAR ASSIST DEVICE
    Rutten, M. C. M.
    Tuzun, E.
    de Vosse, F. N. van
    de Mol, B. A.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2010, 33 (07): : 449 - 449
  • [9] Design of a continuous flow centrifugal pediatric ventricular assist device
    Throckmorton, AL
    Wood, HG
    Day, SW
    Song, X
    Click, PC
    Allaire, PE
    Olsen, DB
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2003, 26 (11): : 1015 - 1031
  • [10] Continuous-Flow Left Ventricular Assist Device REPLY
    Slaughter, Mark S.
    Rogers, Joseph G.
    NEW ENGLAND JOURNAL OF MEDICINE, 2010, 362 (12): : 1149 - 1149