Indirect potentiometric determination of chlorpromazine with an oxidative column in a flow injection system

被引:0
|
作者
Kojlo, A
机构
关键词
solid-phase reactor; flow injection; pharmaceutical analysis;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-injection method is proposed for the determination of chlorpromazine and other N-substituted phenothiazines. The procedure is based on the oxidation of analyte with lead dioxide entrapped into polymeric material in a packed-bed reactor. The oxidation of the drug yields soluble Pb2+, which is monitored by means of the lead ion selective electrode in the wall-jet configuration. The calibration graph is linear over the range of 0.01 - 2 mu g ml(-1) of chlorpromazine with relative standard deviation of 1.4% and sample throughput 20 h(-1). The developed method was applied to the determination of chlorpromazine in pharmaceutical preparations.
引用
收藏
页码:2353 / 2363
页数:11
相关论文
共 50 条