Correction factors for water-proofing sleeves in kilovoltage x-ray beams

被引:20
|
作者
Ma, CM [1 ]
Seuntjens, JP [1 ]
机构
[1] NATL RES COUNCIL CANADA,INST NATL MEASUREMENT STAND,IONIZING RADIAT STAND,OTTAWA,ON K1A 0R6,CANADA
关键词
radiotherapy; kilovoltage x-ray dosimetry; perturbation factor; waterproofing sleeve; EGS4 Monte Carlo;
D O I
10.1118/1.598039
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper investigates the effect of the waterproofing sleeve on the calibration of kilovoltage photon beams (50-300 kV). The sleeve effect correction factor, p(s), has been calculated using the Monte Carlo method as the ratios of the air kerma in an air cavity of a cylindrical chamber without the waterproofing sleeve to that with a sleeve. Three sleeve materials have been studied, PMMA, nylon and polystyrene. The calculations were carried out using the EGS4 (Electron Gamma Shower version 4) code system with the application of a correlated-sampling variance-reduction technique. The results show that the sleeve correction factor for a I-mm thick PMMA waterproofing sleeve surrounding a Farmer-type cylindrical chamber at 2 cm depth in water is about 0.991 at 50 kV (1.04 mm Al HVL) and 1.000 at 300 kV (21.5 mm Al HVL). For I-mm thick nylon and polystyrene sleeves, p(s) varies from 0.992 to 1.000 and from 0.981 to 1.000, respectively, for the same beam quality range. The p(s) factor varies with sleeve thickness, beam quality and phantom depth. No significant dependence of the p(s) factor on field size and source-surface distance has been found. Measurements for PMMA, nylon and polystyrene sleeves of various thicknesses have also been carried out and show excellent agreement with the Monte Carlo calculations. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1507 / 1513
页数:7
相关论文
共 50 条
  • [1] Correction factors for water-proofing sleeves in kilovoltage x-ray beams
    Ionizing Radiation Standards, Inst. for Natl. Msrmt. Standards, National Research Council Canada, Ottawa, Ont. K1A 0R6, Canada
    不详
    MED. PHYS., 9 (1507-1513):
  • [2] Monte Carlo calculated correction factors for water-proofing sleeves in kilovoltage X-ray beams
    Ma, CM
    Seuntjens, JP
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 449 - 451
  • [3] Determination of Backscatter Factors in Diagnostic Kilovoltage X-Ray Beams
    Tanabe, R.
    Araki, F.
    Ohno, T.
    MEDICAL PHYSICS, 2020, 47 (06) : E525 - E525
  • [4] A dosimetric evaluation of water equivalent phantoms for kilovoltage x-ray beams
    Hill, R
    Holloway, L
    Baldock, C
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21): : N331 - N344
  • [5] Energy Response of Thermoluminescent Microcube Dosimeters in Water for Kilovoltage X-Ray Beams
    Di Maso, L.
    Lawless, M.
    Culberson, W.
    DeWerd, L.
    MEDICAL PHYSICS, 2016, 43 (06) : 3624 - 3624
  • [6] The dosimetry of eye shields for kilovoltage X-ray beams
    Wang, D.
    Sobolewski, M.
    Hill, R.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (04) : 491 - 495
  • [7] The dosimetry of eye shields for kilovoltage X-ray beams
    D. Wang
    M. Sobolewski
    R. Hill
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 491 - 495
  • [8] Stereotactic breast irradiation with kilovoltage x-ray beams
    Garnica-Garza, H. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02): : 983 - 995
  • [9] Study of dosimetry consistency for kilovoltage x-ray beams
    Ma, CM
    Li, XA
    Seuntjens, JP
    MEDICAL PHYSICS, 1998, 25 (12) : 2376 - 2384
  • [10] Determination of backscatter factors based on the quality index for diagnostic kilovoltage x-ray beams
    Tanabe, Ryuki
    Araki, Fujio
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 : 48 - 53