Errors in measurement of glomerular filtration rate using the slope-intercept technique and their identification

被引:0
|
作者
Oliveira, Ana [1 ,2 ]
Conway, Lauren [1 ,2 ]
Heraghty, Neil [1 ,2 ]
Peters, A. Michael [1 ,2 ,3 ]
机构
[1] Kings Coll Hosp London, Mol Imaging Ctr, Harley St Clin, London, England
[2] Kings Coll Hosp London, Dept Nucl Med, London, England
[3] Kings Coll Hosp London, Dept Nucl Med, Denmark Hill, London SE5 9RS, England
关键词
glomerular filtration rate; kidney; quality control;
D O I
10.1097/MNM.0000000000001788
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundGFR measured from plasma sampling may be expressed as slope-intercept GFR (SI-GFR) and scaled to body surface area (mGFR/BSA) or as GFR per unit extracellular fluid volume (mGFR/ECV), which is based only on half-time. Measurement errors comprise 3 categories. Pre-injection error arises from error in administered marker and is suspected when mGFR/BSA and mGFR/ECV disagree. Injection errors include 'tissued' injections. Post-injection errors include inaccurate sample timing, inaccurate pipetting, sample haemolysis and sampling through long IV lines through which marker was administered. The aim of the study was to evaluate the impact of errors on mGFR.MethodsWe compared mGFR/BSA with mGFR/ECV in 898 patients undergoing routine investigation. To investigate post-injection error, we took two further patient datasets with r values (correlation coefficient of the 3-sample fit) of 1.0 and introduced errors, in isolation, into each of the 3 recorded sample values, as follows: pipetting (volume) errors of -20%, -10%, -5%, 5%, 10% and 20%, and timing errors of -15 min, -10 min, -5 min, 5 min, 10 min and 15 min.ResultsThe correlation between mGFR/BSA and mGFR/ECV was close and independent of r. Post-injection error depended on the time of the sample in which it occurred. r correlated poorly with error magnitude for both volume and timing errors. When a 'rogue' sample is suspected its error needed to be substantial for it to be identified by single sample estimates applied to the other samples.ConclusionSI-GFR is resistant to post-injection timing and volume errors but not to pre-injection error.
引用
收藏
页码:51 / 60
页数:10
相关论文
共 50 条
  • [1] Using the slope-only technique and estimated glomerular filtration rate for checking the reliability of slope-intercept measurement of glomerular filtration rate
    Bird, Nicholas J.
    Peters, Christina
    Michell, A. Robert
    Peters, A. Michael
    NUCLEAR MEDICINE COMMUNICATIONS, 2008, 29 (12) : 1086 - 1092
  • [2] Correction of the slope-intercept method for the measurement of glomerular filtration rate
    Blake, Glen M.
    Barnfield, Mark C.
    Burniston, Maria T.
    Fleming, John S.
    Cosgriff, Philip S.
    Siddique, Musib
    NUCLEAR MEDICINE COMMUNICATIONS, 2014, 35 (12) : 1277 - 1283
  • [3] The effect of measurement errors on glomerular filtration rate: a comparison of the slope-intercept and slope-only methods
    Holness, J. L.
    Warwick, J. M.
    Fleming, J. S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S358 - S358
  • [4] Commentary to 'Correction of the slope-intercept method for the measurement of glomerular filtration rate'
    Jodal, Lars
    Brochner-Mortensen, Jens
    NUCLEAR MEDICINE COMMUNICATIONS, 2015, 36 (04) : 407 - 408
  • [5] Robustness of slope-intercept versus slope-only for measurement of glomerular filtration rate
    Peters, AM
    NUCLEAR MEDICINE COMMUNICATIONS, 2001, 22 (01) : 115 - 117
  • [6] Calculating the precision of slope-intercept glomerular filtration rate measurements
    Memmott, Matthew J.
    Lawson, Richard S.
    NUCLEAR MEDICINE COMMUNICATIONS, 2022, 43 (08) : 959 - 966
  • [7] Robustness of slope-intercept versus slope-only for measurement of glomerular filtration rate - Reply
    De Sadeleer, C
    Piepsz, A
    Ham, HR
    NUCLEAR MEDICINE COMMUNICATIONS, 2001, 22 (01) : 117 - 117
  • [8] The effect of measurement errors on glomerular filtration rate: a comparison of the slope-intercept, single-sample and slope-only methods
    Holness, J.
    Fleming, J.
    Warwick, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S19 - S19
  • [9] Correction of slope-intercept glomerular filtration rate measurement without scaling for body size
    McMeekin, Helena
    Burniston, Maria T.
    Barnfield, Mark
    Peters, A. Michael
    NUCLEAR MEDICINE COMMUNICATIONS, 2021, 42 (01) : 68 - 72
  • [10] Propagation of measurement errors in glomerular filtration rate determination: a comparison of slope-intercept, single-sample and slope-only methods
    Holness, Jennifer L.
    Fleming, John S.
    Chirehwa, Maxwell T.
    Warwick, James M.
    NUCLEAR MEDICINE COMMUNICATIONS, 2019, 40 (04) : 333 - 342