Using Bland-Altman Analysis to Identify Appropriate Clonogenic Assay Colony Counting Techniques

被引:0
|
作者
Buryska, Seth [1 ]
Arji, Sanjana [1 ]
Wuertz, Beverly [1 ]
Ondrey, Frank [1 ]
机构
[1] Univ Minnesota, Dept Otolaryngol Head & Neck Surg, 2001 Sixth St SE,LRB 270, Minneapolis, MN 55455 USA
关键词
Bland-Altman; clonogenic; colony; counting; algorithm; cancer;
D O I
10.1177/15330338231214250
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Objective: Determine the interchangeability of various methods utilized for counting colonies in clonogenic assays. Methods: Clonogenic assays of 2 head and neck cancer cell lines were counted through 4 different counting modalities: Manual counting pen, via microscope, 1 publicly available automated algorithm, and a semiautomated algorithm presented by the authors. Each method counted individual wells (N = 24). Pen and microscopic counts were performed by 2 observers. Parameters included both low-growth (<150 colonies/well) and high-growth (>150 colonies/well) cell lines. Correlational and Bland-Altman analyses were performed using SPSS software. Results: Interobserver manual pen count correlation R-2 value in both growth conditions was 0.902; controlling for only low-growth conditions decreased R-2 to 0.660. Correlation of microscopic versus pen counts R-2 values for observers 1 and 2 were 0.955 and 0.775, respectively. Comparing techniques, Bland-Altman revealed potential bias with respect to the magnitude of measurement (P < .001) for both observers. Correlation of microscopic counts for both interobserver (R-2 = 0.902) and intraobserver (R-2 = 0.916) were analyzed. Bland-Altman revealed no bias (P = .489). Automated versus microscopic counts revealed no bias between methodologies (P = .787) and a lower correlation coefficient (R-2 = 0.384). Semiautomated versus microscopic counts revealed no bias with respect to magnitude of measurement for either observer (P = .327, .229); Pearson correlation was 0.985 (R-2 = 0.970) and 0.965 (R-2 = 0.931) for observer 1 and 2. Semiautomated versus manual pen colony counts revealed a significant bias with respect to magnitude of measurement (P < .001). Conclusion: Counting with a manual pen demonstrated significant bias when compared to microscopic and semiautomated colony counts; 2 methods were deemed to be interchangeable. Thus, training algorithms based on manual counts may introduce this bias as well. Algorithms trained to select colonies based on size (pixels(2)) and shape (circularity) should be prioritized. Solely relying on Bland-Altman or correlational analyses when determining method interchangeability should be avoided; they rather should be used in conjunction.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Using Bland-Altman Analysis to Identify Appropriate Clonogenic Assay Colony Counting Techniques
    Buryska, Seth
    Arji, Sanjana
    Wuertz, Beverly
    Ondrey, Frank
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2023, 22
  • [2] Using the variance function to generalize Bland-Altman analysis
    Sadler, William A.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2019, 56 (02) : 198 - 203
  • [3] Bland-Altman analysis for pupillometer comparison
    Brown, Sandra
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (10): : 1802 - 1803
  • [4] Bland-Altman analysis for pupillometer comparison REPLY
    Scheffel, Magdalena
    Kuehne, Christoph
    Kohnen, Thomas
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (10): : 1803 - 1804
  • [5] Bland-Altman analysis: A paradigm to understand correlation and agreement
    Dogan, Nurettin Ozgur
    TURKISH JOURNAL OF EMERGENCY MEDICINE, 2018, 18 (04): : 139 - 141
  • [6] BA.plot: An R function for Bland-Altman analysis
    Karun, Kalesh M.
    Puranik, Amitha
    CLINICAL EPIDEMIOLOGY AND GLOBAL HEALTH, 2021, 12
  • [7] Comparison of two methods of measurements: the Bland-Altman analysis.
    Grenier, B
    Dubreuil, M
    Journois, D
    ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2000, 19 (02): : 128 - 135
  • [8] Using the Bland-Altman method to measure agreement with repeated measures
    Myles, P. S.
    Cui, J.
    BRITISH JOURNAL OF ANAESTHESIA, 2007, 99 (03) : 309 - 311
  • [9] Statistical Analysis Using Bland-Altman Agreement for Technical Methods of Measuring Blood Glucose
    Dias, Cifha
    Kamath, Surekha
    Vidyasagar, Sudha
    2017 INTERNATIONAL CONFERENCE ON ELECTRICAL, ELECTRONICS, COMMUNICATION, COMPUTER, AND OPTIMIZATION TECHNIQUES (ICEECCOT), 2017, : 693 - 698
  • [10] Assessment of Operator Reliability in Ultrasound of the Median and Ulnar Nerve Using Bland-Altman Analysis
    Philipps, Joerg
    Mork, Hannah
    Katz, Maria
    Knaup, Mark
    Beyer, Kira
    Spies, Kristina
    Corbach, Johannes
    Schellinger, Peter D.
    DIAGNOSTICS, 2021, 11 (11)