Continuous cardiac monitoring in epilepsy: an implantable loop manual activation algorithm for improving ECG signal acquisition accuracy

被引:0
|
作者
Davtyan, Karapet [1 ]
Serdyuk, Svetlana [1 ]
Topchyan, Arpi [1 ]
Simonyan, Georgiy [1 ]
Kharlap, Maria [1 ]
Burd, Sergey [2 ]
机构
[1] Natl Med Res Ctr Therapy & Prevent Med, Petroverigskiy Lane 10-3, Moscow 101990, Russia
[2] Pirogov Russian Res Med Univ, Ostrovitianov str 1, Moscow 117997, Russia
关键词
Cardiac arrhythmias; Continuous monitoring; Seizure; False-positive arrhythmias; ARRHYTHMIAS;
D O I
10.1186/s12872-024-03721-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The muscle artifacts, caused by prominent muscle contractions, mimicking cardiac arrhythmias, might compromise the ECG signal quality and the implantable loop recorder memory capacity in patients with epilepsy. We developed an epileptic seizures clinical pattern-based implantable loop recorder manual activation algorithm, presenting its real-world efficacy here.Methods One hundred ninety-three patients (18-60 years) with drug-resistant focal epilepsy were consecutively enrolled and underwent a subcutaneous loop recorder implantation. Patients with focal onset-aware seizures and patients with focal impaired awareness seizures /bilateral tonic-clonic seizures without aura were recommended to use the activator once - just after the episode. Patients with focal impaired awareness seizures/bilateral tonic-clonic seizures with aura, the caregivers of patients experiencing status epilepticus, were advised to use the activator twice - during the aura and after the episode/ regaining consciousness.Results Six thousand four hundred ninety-four ECG traces (4826 - auto-triggered events, 1668 - person-activated events) were recorded and analyzed. The rate of true positive events in the person-activated group was statistically higher than in the autoactivation group (72.5% vs.19.4%, p < 0.0001). Person-activated false-positive events were observed in 30.5% of patients with focal impaired awareness seizures and 27.7% in patients with bilateral tonic-clonic seizures. The highest rate of false-positive events (61.5%) was detected in patients undergoing epileptic status, and the lowest rate (3.8%) - was in patients with focal onset aware seizures. The rate of false-positive events was significantly higher in patients with impaired awareness seizures without aura both in focal impaired awareness (45.5% vs. 19.3%, p < 0.0001) and bilateral tonic-clonic seizure groups (38.8% vs. 5.9%, p < 0.0001).Conclusions Arrhythmias with varying clinical outcomes are expected in epilepsy patients and have been monitored continuously. The specified loop recorder external activation algorithm can improve the clinically relevant cardiac arrhythmia detection accuracy in epilepsy patients and the value of future studies.
引用
收藏
页数:5
相关论文
共 9 条
  • [1] Continuous cardiac monitoring in epilepsy: an implantable loop manual activation algorithm for improving ECG signal acquisition accuracy
    Karapet Davtyan
    Svetlana Serdyuk
    Arpi Topchyan
    Georgiy Simonyan
    Maria Kharlap
    Sergey Burd
    BMC Cardiovascular Disorders, 24
  • [2] 1 IN 8 PATIENTS WITH SYNCOPE ARE MISDIAGNOSED AS EPILEPSY ON LONG TERM CARDIAC RHYTHM MONITORING BY AN IMPLANTABLE ECG LOOP RECORDER
    Petkar, S.
    Hamid, T.
    Rose, S.
    Clifford, A.
    Homa, S.
    Garratt, C.
    Clarke, B.
    Cooper, P.
    Fitzpatrick, A.
    JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2009, 80 (11):
  • [3] Continuous ECG Monitoring Trial for Outpatient - Patient Receptiveness and Signal Accuracy
    Wong, David Liang Tai
    Sathappan, Selva Muthu Kumaran
    Yu, Jufeng
    Heng, Chun Huat
    Kojodjojo, Pipin
    Feng, Mengling
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 1144 - 1148
  • [4] 1 IN 8 PATIENTS WITH SYNCOPE ARE MISDIAGNOSED AS EPILEPSY ON LONG-TERM CARDIAC RHYTHM MONITORING BY AN IMPLANTABLE ECG LOOP RECORDER (ILR)
    Hamid, T.
    Rose, S.
    Clifford, A.
    Homa, S.
    Garratt, C.
    Clarke, B.
    Cooper, P.
    Fitzpatrick, A.
    Petkar, S.
    EPILEPSIA, 2009, 50 : 35 - 35
  • [5] Which Brugada patient deserves continuous ECG monitoring through implantable loop recorder? An evidence update
    Calvelli, Pierangelo
    Cerrato, Natascia
    Giustetto, Carla
    Saglietto, Andrea
    Anselmino, Matteo
    Curcio, Antonio
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2025, 26 (02) : 64 - 71
  • [6] Atrial fibrillation and treatment changes in cryptogenic stroke patients with an implantable loop recorder for continuous cardiac rhythm monitoring
    Ungar, A.
    Rieger, G.
    West, T.
    Purerfellner, H.
    Bucx, J.
    Topper, R. F.
    Kuster, S.
    Nagele, H.
    De Melis, M.
    Duru, F.
    EUROPEAN HEART JOURNAL, 2013, 34 : 62 - 62
  • [7] Incidence, duration, pattern, and burden of de novo atrial arrhythmias detected by continuous ECG monitoring using an implantable loop recorder following ablation of the cavotricuspid isthmus
    Musat, Dan L.
    Milstein, Nicolle S.
    Pimienta, Jacqueline
    Bhatt, Advay
    Preminger, Mark W.
    Sichrovsky, Tina C.
    Flynn, Laura
    Pistilli, Carissa
    Shaw, Richard E.
    Mittal, Suneet
    CARDIOVASCULAR DIGITAL HEALTH JOURNAL, 2020, 1 (03): : 114 - 122
  • [8] Value of long term cardiac rhythm monitoring by an implantable ECG loop recorder (ILR) in patients presenting to the neurologist with transient loss of consciousness (T-LOC)
    Hamid, T.
    Rose, S.
    Clifford, A.
    Homa, S.
    Garratt, C.
    Clarke, B.
    Cooper, P.
    Fitzpatrick, A.
    Petkar, S.
    McHugh, C.
    JOURNAL OF THE NEUROLOGICAL SCIENCES, 2009, 285 : S112 - S112