Very Accurate Flexible pH Microsensor Based on Nanoporous Titanium Nitride Material for In Vivo Application

被引:1
|
作者
Nguyen, Thuy [1 ]
Lissorgues, Gaelle [1 ]
Takhedmit, Hakim [1 ]
Rousseau, Lionel [1 ]
Journet, Bernard [2 ]
Ridoux, Audrey [3 ]
Lecarpentier, Edouard [3 ]
机构
[1] Univ G Eiffel, CNRS UMR 9007, ESYCOM, F-77454 Marne La Vallee, France
[2] Univ Paris Saclay, CNRS, ENS Paris Saclay, LUMIN UMR 9024, F-91190 Gif Sur Yvette, France
[3] Univ Paris Est Creteil, Ctr Hosp Intercommunal Creteil, INSERM, U955 IMRB, F-94000 Creteil, France
关键词
Flexible sensors; microfabrication; microsensors; pH sensors; TiN material; SENSORS; LABOR;
D O I
10.1109/JSEN.2023.3321075
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Recently, the need for a real-time and less invasive solution to replace in-laboratory blood analysis by analyzing in situ biological fluids (sweat, saliva, tear, urine, and interstitial fluid) has grown alongside the development of microfabrication techniques. pH determination is crucial for health monitoring (e.g., fetal acidosis supervision, sport activity, tumor detection, and wound monitoring). Our work focuses on developing a potentiometric microsensor, targeting pH measurement on the fetal scalp during labor, employing nanoporous titanium nitride (TiN) as an alternative sensing material. The pH probe contains a working electrode (WE) of 400 x 800 mu m in size, integrated on a flexible polyimide (PI) substrate, creating an ultrathin sensor with thickness below 20 mu m. Microfabrication process and a complete characterization on performance of our flexible pH microsensor are presented, demonstrating a sensitivity of 53.2 mV/pH. Hysteresis was found at 3-mV maximum after six cycles, demonstrated a good reproducibility, and led to a precision of 0.05 in pH. A comparison of the performance between porous and flat TiN (sensitivity of 53.7 mV/pH, hysteresis maximum of 5 mV, and precision of 0.09 in pH) is presented, along with an evaluation of the dependence in the electrode size versus their performances. A preliminary in vivo test was conducted on mice with the probe implanted under the skin. We recorded a stable open-circuit voltage (OCV) signal under anesthesia conditions. A compact measurement system was designed with a customized liquid crystal display (LCD) screen for signal visualization, an EZO ORP circuit for data acquisition, and an integrated secure digital (SD card) for data storage.
引用
收藏
页码:27019 / 27030
页数:12
相关论文
共 50 条
  • [1] A novel pH electrode based on titanium nitride
    Lu, XL
    Yuan, HY
    Xiao, D
    CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE, 2003, 24 (08): : 1400 - 1402
  • [2] Titanium nitride based plasmonic nanoparticles for photovoltaic application
    Akhtary, Nowshin
    Zubair, Ahmed
    OPTICS CONTINUUM, 2023, 2 (07): : 1701 - 1715
  • [3] Flexible Advanced Asymmetric Supercapacitors Based on Titanium Nitride-based Nanowire Electrodes
    Liu, Yi
    Xiao, Rujian
    Qiu, Yongfu
    Fang, Yiqi
    Zhang, Peng
    ELECTROCHIMICA ACTA, 2016, 213 : 393 - 399
  • [4] Porous Titanium Nitride electrodes on miniaturised pH sensor for foetal health monitoring application
    Nguyen, T.
    Journet, B.
    Takhedmit, H.
    Lissorgues, G.
    2022 SYMPOSIUM ON DESIGN, TEST, INTEGRATION AND PACKAGING OF MEMS/MOEMS, DTIP, 2022,
  • [5] Superhard Composite Material Based on Cubic Boron Nitride Plated With Titanium Compounds
    Rudenok, L. P.
    Suvorov, S. A.
    Rumyantsev, V., I
    Bulatov, O. N.
    REFRACTORIES AND INDUSTRIAL CERAMICS, 2022, 62 (06) : 663 - 667
  • [6] Superhard Composite Material Based on Cubic Boron Nitride Plated With Titanium Compounds
    L. P. Rudenok
    S. A. Suvorov
    V. I. Rumyantsev
    O. N. Bulatov
    Refractories and Industrial Ceramics, 2022, 62 : 663 - 667
  • [7] Titanium Nitride as a Plasmonic Material from Near-Ultraviolet to Very-Long-Wavelength Infrared Range
    Judek, Jaroslaw
    Wrobel, Piotr
    Michalowski, Pawel Piotr
    Ozga, Monika
    Witkowski, Bartlomiej
    Seweryn, Aleksandra
    Struzik, Michal
    Jastrzebski, Cezariusz
    Zberecki, Krzysztof
    MATERIALS, 2021, 14 (22)
  • [8] In Vivo Monitoring of pH in Subacute PD Mouse Brains with a Ratiometric Electrochemical Microsensor Based on Poly(melamine) Films
    Zhang, Ziyi
    Li, Mengyin
    Zuo, Yimei
    Chen, Shu
    Zhuo, Yi
    Lu, Ming
    Shi, Guoyue
    Gu, Hui
    ACS Sensors, 2022, 7 (01): : 235 - 244
  • [9] In Vivo Monitoring of pH in Subacute PD Mouse Brains with a Ratiometric Electrochemical Microsensor Based on Poly(melamine) Films
    Zhang, Ziyi
    Li, Mengyin
    Zuo, Yimei
    Chen, Shu
    Zhuo, Yi
    Lu, Ming
    Shi, Guoyue
    Gu, Hui
    ACS SENSORS, 2021, : 235 - 244
  • [10] Electrochemical flexible supercapacitor based on manganese dioxide-titanium nitride nanotube hybrid
    Xie, Yibing
    Fang, Xiuqin
    ELECTROCHIMICA ACTA, 2014, 120 : 273 - 283