共 50 条
Development of hard X-ray spectrometer with full digital data acquisition for runaway electron studies at HL-2M
被引:4
|作者:
Cheng, Shikui
[1
]
Zhang, Jie
[1
]
Zhang, Yipo
[1
]
He, Li
[2
]
Xu, Hongbing
[1
]
Zhu, Yuxuan
[1
]
Zhan, Xuwen
[1
]
Shi, Zhongbing
[1
]
Zhong, Wulv
[1
]
机构:
[1] Southwestern Inst Phys, Chengdu 610041, Sichuan, Peoples R China
[2] Chengdu Univ Technol, Coll Nucl Technol & Automat Engn, Chengdu 610059, Sichuan, Peoples R China
基金:
国家重点研发计划;
中国国家自然科学基金;
关键词:
Nuclear instruments and methods for hot plasma diagnostics;
Gamma detectors (scin- tillators;
CZT;
HPGe;
HgI etc);
X-ray detectors;
DISRUPTIONS;
BEHAVIOR;
D O I:
10.1088/1748-0221/18/02/T02006
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
The development of hard X-ray (HXR) spectrometers with full digital data acquisition for runaway electron studies on the HL-2M tokamak is presented. Consisting of large-sized NaI:Tl detectors and high-performance multichannel analyzers, this system can realize the measurement of HXR with energies ranging from 0.5 to 10 MeV, and spectrum temporal resolution of 1 ms. An average energy resolution of 56 keV at 662 keV has been achieved at counts rates of about 100 kcps. The spectra data is acquired by a digital-based data acquisition system with fast digitization and software signal processing technology. The energy calibrations are performed by using several radioisotope gamma-ray sources, and the detection efficiencies are simulated by MCNP code. First experimental results from HL-2M Ohmic discharges are presented in this paper.
引用
收藏
页数:12
相关论文