Fully automated F-wave corridor extraction and analysis algorithm for F-wave analyses and MUNE studies

被引:1
|
作者
Artug, N. Tugrul [1 ]
机构
[1] Istanbul Univ Cerrahpasa, Vocat Sch Tech Sci, Dept Elect, Istanbul, Turkiye
来源
SCIENTIFIC REPORTS | 2023年 / 13卷 / 01期
关键词
MULTIPLE POINT STIMULATION; BASE-LINE CORRECTION; THENAR MOTOR UNITS; NUMBER; PHYSIOLOGY; FEATURES;
D O I
10.1038/s41598-023-41183-5
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
F-waves are used in motor unit number estimation (MUNE) studies, which require rapid dedicated software to perform calculations. The aim of this study is to define a mathematical method for a fully automated F-wave extraction algorithm to perform F-wave and MUNE studies while performing baseline corrections without distorting traces. Ten recordings from each class, such as healthy controls, polio patients and ALS patients, were included. Submaximal stimuli were applied to the median and ulnar nerves to record 300 traces from the abductor pollicis brevis and abductor digiti minimi muscles. The autocorrelation function and the signal of sum of all traces were used to find the location for the maximum amplitude of the F-waves. F-waves were revealed by using a cutting window. Linear line estimation was preferred for baseline corrections because it did not cause any distortion in the traces. The algorithm automatically revealed F-waves from all 30 recordings in accordance with the locations marked by a neurophysiologist. The execution of the algorithm was less than 2 (usually<1) minutes when 300 traces were analyzed. Mean sMUP amplitudes and MUNE values are important for differentiating healthy controls from patients. Moreover, F-wave parameters belonging to polio patients on whom there was a relatively low number of studies conducted were also evaluated.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Fully automated F-wave corridor extraction and analysis algorithm for F-wave analyses and MUNE studies
    N. Tuğrul Artuğ
    Scientific Reports, 13 (1)
  • [3] Josephson Current in f-Wave Superconductor/f-Wave Superconductor Junctions
    Liao Yan-Hua
    Dong Zheng-Chao
    Liang Zhi-Peng
    Yin Zai-Feng
    COMMUNICATIONS IN THEORETICAL PHYSICS, 2009, 52 (04) : 715 - 720
  • [4] Anodal block in F-wave studies
    Kirshblum, S
    Cai, PT
    Johnston, MV
    Shah, V
    O'Connor, K
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1998, 79 (09): : 1059 - 1061
  • [5] Reproducibility in nerve conduction studies and F-wave analysis
    Pinheiro, Denise Spinola
    Manzano, Gilberto Mastrocola
    Maciel Nobrega, Joao Antonio
    CLINICAL NEUROPHYSIOLOGY, 2008, 119 (09) : 2070 - 2073
  • [6] F-WAVE DETERMINATION IN POLYNEUROPATHIES
    KIMURA, J
    SHIVAPOUR, E
    YAMADA, T
    ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1981, 52 (03): : S100 - S100
  • [7] VARIABILITY OF F-WAVE SHAPE
    KOMORI, T
    HIROSE, K
    TSUBAKI, T
    ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1987, 67 (03): : P72 - P72
  • [8] ARE F-WAVE RESPONSES RANDOM
    PETAJAN, JH
    MUSCLE & NERVE, 1983, 6 (07) : 536 - 536
  • [9] CLINICAL AND PATHOLOGICAL STUDIES OF ATRIAL F-WAVE
    SUZUKI, Y
    OHKUBO, T
    JAPANESE CIRCULATION JOURNAL, 1969, 33 (08): : 875 - &
  • [10] Comparison of automated and manual F-wave latency measurements
    Fisher, MA
    CLINICAL NEUROPHYSIOLOGY, 2005, 116 (02) : 264 - 269