Accuracy of a Patient-Specific Total Ankle Arthroplasty Instrumentation

被引:2
|
作者
April, Pierre-Marc [1 ]
Locke, Eric [1 ]
Champagne, Philippe-Hugo [2 ]
Angers, Magalie [2 ]
Martinez-Gomez, Adriana Patricia [1 ]
Seidel, Angela [1 ]
Jibri, Zaid [1 ]
Lalonde, Karl-Andre [1 ]
Meulenkamp, Brad [1 ]
机构
[1] Civ Hosp, Dept Orthopaed, Ottawa, ON, Canada
[2] CHUS, Dept Orthopaed, Sherbrooke, PQ, Canada
关键词
foot and ankle surgery; total ankle arthroplasty; patient specific instrumentation; alignment outcomes; END-STAGE ANKLE; LONG-TERM OUTCOMES; REPLACEMENT; ARTHRODESIS; INTERMEDIATE; SURVIVORSHIP; MISALIGNMENT; ALIGNMENT; SERIES;
D O I
10.1177/10711007231194049
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Total ankle arthroplasty (TAA) is a preferred surgical option for end-stage ankle osteoarthritis; however, it is a demanding procedure with a higher historical rate of revision compared with ankle fusion. Patient-specific instrumentation (PSI) has been introduced to optimize prosthesis alignment and theoretically overall improve TAA outcomes. The goal of this study is to report on the experience and surgical outcomes of one implant with specific evaluation of the accuracy and reproducibility of the system with respect to prosthesis alignment and prediction of implant size. Methods: A retrospective, multicentered study involving 4 foot and ankle fellowship-trained orthopaedic surgeon's patients undergoing TAA between January 1, 2015, and December 31, 2018, using the PROPHECY PSI system. Results: 80 TAA procedures were performed. On average the postoperative tibial component alignment was 89.9 (range, 86.1-96.5) degrees in the coronal plane, with a mean sagittal alignment of 88.1 (range, 81.3-96.7) degrees. The mean deviation from neutral sagittal alignment improved from 4.9 +/- 3.9 degrees preoperatively to 2.7 +/- 1.7 degrees postoperatively, whereas the mean coronal alignment improved from 3.3 +/- 2.5 degrees to 1.3 +/- 1.1 degrees. The PSI software correctly determined the tibial implant size in 70 patients (89%). Prediction of talar implant sizing was less accurate than the tibial component, with 56 patients (71%) using the predicted sized implant. The overall implant survival at a mean follow-up of 45 months (range, 27-76) was 97.5%. Conclusion: We found that this PSI system accurately and reliably assisted in implant total ankle prosthesis positioning within a clinically acceptable margin and without significant outliers. Prediction of implant size was not as accurate as component orientation. Level of Evidence: Level III, retrospective study.
引用
收藏
页码:1150 / 1157
页数:8
相关论文
共 50 条
  • [1] Accuracy and Reproducibility Using Patient-Specific Instrumentation in Total Ankle Arthroplasty
    Daigre, Justin
    Berlet, Gregory
    Van Dyke, Bryan
    Peterson, Kyle S.
    Santrock, Robert
    FOOT & ANKLE INTERNATIONAL, 2017, 38 (04) : 412 - 418
  • [2] Patient-specific instrumentation in total ankle arthroplasty
    Mazzotti, Antonio
    Arceri, Alberto
    Zielli, Simone
    Bonelli, Simone
    Viglione, Valentina
    Faldini, Cesare
    WORLD JOURNAL OF ORTHOPEDICS, 2022, 13 (03): : 230 - 237
  • [3] Accuracy of patient-specific instrumentation in total ankle arthroplasty: A comparative study
    Saito, Guilherme H.
    Sanders, Austin E.
    O'Malley, Martin J.
    Deland, Jonathan T.
    Ellis, Scott J.
    Demetracopoulos, Constantine A.
    FOOT AND ANKLE SURGERY, 2019, 25 (03) : 383 - 389
  • [4] Accuracy of patient-specific instrumentation for implant positioning in custom-made total ankle arthroplasty
    Mazzotti, Antonio
    Zielli, Simone Ottavio
    Arceri, Alberto
    Artioli, Elena
    Langone, Laura
    Sgubbi, Federico
    Geraci, Giuseppe
    Faldini, Cesare
    JOURNAL OF EXPERIMENTAL ORTHOPAEDICS, 2024, 11 (03)
  • [5] Patient-specific instrumentation (PSI) in total ankle arthroplasty: a systematic review
    Wang, Qiuyuan
    Zhang, Nianfei
    Guo, Wanshou
    Wang, Weiguo
    Zhang, Qidong
    INTERNATIONAL ORTHOPAEDICS, 2021, 45 (09) : 2445 - 2452
  • [6] Patient-specific instrumentation (PSI) in total ankle arthroplasty: a systematic review
    Qiuyuan Wang
    Nianfei Zhang
    Wanshou Guo
    Weiguo Wang
    Qidong Zhang
    International Orthopaedics, 2021, 45 : 2445 - 2452
  • [7] Accuracy of Patient-Specific Instrumentation Compared With Conventional Instrumentation in Total Knee Arthroplasty
    Jiang, Jin
    Kang, Xin
    Lin, Qiao
    Teng, Yuanjun
    An, Liping
    Ma, Jingling
    Wang, Jing
    Xia, Yayi
    ORTHOPEDICS, 2015, 38 (04) : E305 - E313
  • [8] Effectiveness and Accuracy of a Patient-Specific Instrumentation System for Total Hip Arthroplasty
    Zhang, Teng
    Jia, Zhao
    Han, Wei
    Wang, Junqiang
    Li, Jinqi
    Gong, Maoqi
    Jiang, Xieyuan
    ORTHOPAEDIC SURGERY, 2023, 15 (03) : 878 - 887
  • [9] Patient-Specific Instrumentation Does Not Improve Accuracy in Total Knee Arthroplasty
    Shen, Chong
    Tang, Zhi-Hong
    Hu, Jun-Zu
    Zou, Guo-Yao
    Xiao, Rong-Chi
    Yan, Dong-Xue
    ORTHOPEDICS, 2015, 38 (03) : E178 - E188
  • [10] Accuracy of patient-specific instrumentation in anatomic and reverse total shoulder arthroplasty
    Dallalana, Richard James
    McMahon, Ryan A.
    East, Ben
    Geraghty, Liam
    INTERNATIONAL JOURNAL OF SHOULDER SURGERY, 2016, 10 (02): : 59 - 66