A Modified Technique for Interventional MRI-Guided Deep Brain Stimulation Using the ClearPoint System

被引:1
|
作者
Harland, Tessa A. [1 ]
Brougham, Jared [2 ]
Gupta, Shruti [3 ]
Strahan, Jessica [3 ]
Hefner, Matthew [4 ]
Wilden, Jessica [2 ,4 ]
机构
[1] Albany Med Ctr, Dept Neurosurg, Albany, NY USA
[2] LSU Sch Med, Dept Neurosurg, 1501 Kings Hwy, Shreveport, LA 71103 USA
[3] ClearPoint Neuro Inc, Solana Beach, CA USA
[4] Willis Knighton Hlth Syst, Dept Neurosurg, Shreveport, LA USA
关键词
Asleep; Deep brain stimulation; Interventional MRI; MRI-guided; ClearPoint; PARKINSONS-DISEASE; CLINICAL-OUTCOMES; NUCLEUS; AWAKE;
D O I
10.1227/ons.0000000000000698
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND:Deep brain stimulation (DBS) is increasingly performed using imaging-based techniques, which are rapidly evolving to minimize invasiveness and use limited resources efficiently.OBJECTIVE:To describe a modified surgical technique for interventional MRI-guided stereotactic procedures using ClearPoint.METHODS:A total of 160 consecutive DBS cases for Parkinson disease (48%), essential tremor (44%), dual essential tremor/Parkinson disease (3%), and dystonia (5%) were studied. Technique modifications included (1) preplanning the target/trajectory, (2) adoption of a smaller incision and new stereotactic frame, (3) changing the sequence of surgical steps, aiming to final target PRIOR to drilling, (4) reducing burr hole size, (5) minimizing dural opening, and (6) anchoring leads with a bio cement/plate combination in lieu of commercial devices. Demographic data, radial error, number of passes, and case time were prospectively collected while complications and clinical outcomes were retrospectively assessed.RESULTS:One hundred and forty seven DBS patients had 262 total leads placed. Average radial error was 0.41 mm bilaterally. Seventy nine leads had error 1 mm. Pin-to-stylet insertion time, pin-to-frame removal time, and pin-to-room departure time were 2 hours 1 minute, 2 hours 48 minutes, and 3 hours 16 minutes, respectively, and were significantly shorter than those obtained with the original ClearPoint technique for our first 150 DBS cases (P < .0001). Complications included infection with hardware removal (2.5%) and superficial hemorrhage without permanent neurological sequelae (1.9%).CONCLUSION:The modified ClearPoint technique maintains accuracy while reducing invasiveness and MRI time. This workflow can be expanded to other applications, including laser interstitial thermal therapy and gene delivery.
引用
收藏
页码:81 / 86
页数:6
相关论文
共 50 条
  • [1] Interventional MRI-Guided Deep Brain Stimulation Lead Implantation
    Lee, Philip S.
    Richardson, Robert Mark
    NEUROSURGERY CLINICS OF NORTH AMERICA, 2017, 28 (04) : 535 - +
  • [2] ClearPoint versus frame-based MRI-guided and MRI-verified deep brain stimulation
    Zrinzo, Ludvic
    Akram, Harith
    Hariz, Marwan
    JOURNAL OF NEUROSURGERY, 2020, 133 (05) : 1624 - 1625
  • [3] ClearPoint versus frame-based MRI-guided and MRI-verified deep brain stimulation Response
    Sharma, Vibhash D.
    Bezchlibnyk, Yarema B.
    Gross, Robert E.
    JOURNAL OF NEUROSURGERY, 2020, 133 (05) : 1625 - 1626
  • [4] Interventional MRI-guided deep brain stimulation in pediatric dystonia: first experience with the Clear Point system
    Starr, Philip A.
    Markun, Leslie C.
    Larson, Paul S.
    Volz, Monica M.
    Martin, Alastair J.
    Ostrem, Jill L.
    JOURNAL OF NEUROSURGERY-PEDIATRICS, 2014, 14 (04) : 400 - 408
  • [5] Clinical outcomes using ClearPoint interventional MRI for deep brain stimulation lead placement in Parkinson's disease
    Ostrem, Jill L.
    Ziman, Nathan
    Galifianakis, Nicholas B.
    Starr, Philip A.
    San Luciano, Marta
    Katz, Maya
    Racine, Caroline A.
    Martin, Alastair J.
    Markun, Leslie C.
    Larson, Paul S.
    JOURNAL OF NEUROSURGERY, 2016, 124 (04) : 908 - 916
  • [6] Design of a Robotic System for MRI-Guided Deep Brain Stimulation Electrode Placement
    Cole, Gregory
    Pilitsis, Julie
    Fischer, Gregory S.
    ICRA: 2009 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-7, 2009, : 1034 - +
  • [7] Ultra High Field MRI-Guided Deep Brain Stimulation
    Forstmann, Birte U.
    Isaacs, Bethany R.
    Temel, Yasin
    TRENDS IN BIOTECHNOLOGY, 2017, 35 (10) : 904 - 907
  • [8] Interventional MRI-guided brain biopsies using inductively coupled surface coils
    Staubert, A
    Vester, M
    Tronnier, VM
    Renz, W
    Wirtz, CR
    Bonsanto, MM
    Kunze, S
    MAGNETIC RESONANCE IN MEDICINE, 2000, 43 (02) : 278 - 283
  • [9] Wide-bore 1.5 T MRI-guided deep brain stimulation surgery: initial experience and technique comparison
    Sillay, Karl A.
    Rusy, Deborah
    Buyan-Dent, Laura
    Ninman, Nancy L.
    Vigen, Karl K.
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2014, 127 : 79 - 85
  • [10] A Technique for MRI-Guided Transrectal Deep Pelvic Abscess Drainage
    Nour, Sherif Gamal
    Derakhshan, Jamal J.
    Akhtar, Nila J.
    Ayres, Martin A.
    Clampitt, Mark E.
    Stellato, Thomas A.
    Duerk, Jeffrey L.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 191 (04) : 1182 - 1185