Left atrial disc device thrombosis after atrial defect closure with the Gore® Cardioform septal occluder: A case series

被引:4
|
作者
Kramer, Peter [1 ,2 ,3 ,4 ,5 ,6 ]
Opgen-Rhein, Bernd [1 ,2 ,3 ,4 ,5 ]
Berger, Felix [1 ,2 ,3 ,4 ,5 ]
Nordmeyer, Johannes [1 ,2 ,3 ,4 ,5 ]
机构
[1] Deutsch Herzzentrum Charite, Dept Congenital Heart Dis Pediat Cardiol, Berlin, Germany
[2] Free Univ Berlin, Berlin, Germany
[3] Humboldt Univ, Berlin, Germany
[4] Charite, Berlin, Germany
[5] DZHK, German Ctr Cardiovasc Res, Berlin, Germany
[6] Deutsch Herzzentrumder Charite, Dept Congenital Heart Dis Pediat Cardiol, Augustenburger Pl 1, D-13353 Berlin, Germany
关键词
device thrombosis; Gore (R) Cardioform septal occluder; patent foramen ovale closure; secundum type atrial septal defect closure; PATENT FORAMEN OVALE;
D O I
10.1002/ccd.30747
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Device thromboses after patent foramen ovale (PFO) and atrial septal defect (ASD) closure are rare but serious complications. They have been reported in different devices of virtually all manufacturers. Here, we report three cases of left atrial device thrombosis after atrial defect closure with the Gore (R) Cardioform septal occluder (GSO) in our recent institutional experience. All patients were symptomatic with new-onset neurological impairments and evidence of cerebral thromboembolism. Device thromboses have occurred despite antiplatelet therapy in two patients and moreover were observed late, at approximately 2 years after implantation, in two patients. One device was surgically explanted while in two cases, thrombi completely resolved with initiated anticoagulation. All patients had a favorable neurological recovery. Our observations suggest that regular follow-up echocardiography beyond 6 months after device implantation might be advisable in patients with GSO devices to rule out late device thromboses. More long-term follow-up data considering safety and late complications of contemporary PFO and ASD devices is required to establish confident evidence-based recommendations on long-term follow-up management and antithrombotic therapy regimen after PFO and ASD closure.
引用
收藏
页码:359 / 363
页数:5
相关论文
共 50 条
  • [1] Rescue atrial septal defect closure with the new GORE® cardioform atrial septal defect occluder
    Pizzuto, Alessandra
    Cuman, Magdalena
    Cantinotti, Massimiliano
    Franchi, Eliana
    Corana, Giulia
    Viacava, Cecilia
    Assanta, Nadia
    Santoro, Giuseppe
    ANNALS OF PEDIATRIC CARDIOLOGY, 2022, 15 (02) : 192 - 194
  • [2] Device-Related Thrombosis After Transcatheter Closure of Atrial Septal Defect Using the Gore Cardioform ASD Occluder
    Moolla, Muhammad
    Linjawi, Hatem
    Taylor, Dylan
    Mathew, Anoop
    JOURNAL OF INVASIVE CARDIOLOGY, 2024, 36 (07):
  • [3] Reversible atrioventricular block after atrial septal defect closure with a Gore Cardioform Septal Occluder
    Figueras-Coll, Marc
    Sabate-Rotes, Anna
    Roguera-Sopena, Marc
    Betrian-Blasco, Pedro
    ARCHIVOS DE CARDIOLOGIA DE MEXICO, 2020, 90 (01): : 90 - 92
  • [4] Upsizing of GORE® Cardioform ASD Occluder for Atrial Septal Defect With Atrial Septal Aneurysm
    Hayama, Hiromasa
    Makino, Kenji
    Yazaki, Yoshiyuki
    Hara, Hidehiko
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (06)
  • [5] Perforating the GORE® CARDIOFORM septal occluder and atrial septal defect occluder to gain access to the left atrium
    Anderson, Jason H.
    Adamson, Teagan
    Migliati, Elton
    Herlihy, Jennipher
    Danieu, Pierrette
    Daly, John
    Allen, Carissa
    Hua, Khoa
    Inglessis, Ignacio
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 96 (06) : E660 - E665
  • [6] Device-related thrombosis on atrial septal defect occluder after simultaneous closure of left atrial appendage and atrial septal defect: a case report
    Lu HE
    Ya-Juan DU
    Ge-Sheng CHENG
    Yu-Shun ZHANG
    Journal of Geriatric Cardiology, 2019, 16 (06) : 490 - 494
  • [7] Device-related thrombosis on atrial septal defect occluder after simultaneous closure of left atrial appendage and atrial septal defect: a case report
    He, Lu
    Du, Ya-Juan
    Cheng, Ge-Sheng
    Zhang, Shun
    JOURNAL OF GERIATRIC CARDIOLOGY, 2019, 16 (06) : 490 - 494
  • [8] Device-related thrombosis on atrial septal defect occluder after simultaneous closure of left atrial appendage and atrial septal defect: a case report
    Lu HE
    Ya-Juan DU
    Ge-Sheng CHENG
    Yu-Shun ZHANG
    Journal of Geriatric Cardiology, 2019, 16 (06) : 490 - 494
  • [9] Recurrent Strokes After Gore Septal Occluder Device Closure of Atrial Septal Defect
    Alhadlaq, Adnan
    Hussain, Arif
    Friesen, Camille Hancock
    Dhillon, Santokh
    CANADIAN JOURNAL OF CARDIOLOGY, 2017, 33 (12) : 1736.e9 - 1736.e11
  • [10] Initial Clinical Experience With the GORE® CARDIOFORM ASD Occluder for Transcatheter Atrial Septal Defect Closure
    de Hemptinne, Quentin
    Horlick, Eric M.
    Osten, Mark D.
    Millan, Xavier
    Tadros, Victor-Xavier
    Pighi, Michele
    Barlatey, Francisco Gonzalez
    Alnasser, Sami M.
    Miro, Joaquim
    Asgar, Anita W.
    Ibrahim, Reda
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 90 (03) : 495 - 503