Interaction of surface topography and taper mismatch on head-stem modular junction contact mechanics during assembly in modern total hip replacement

被引:9
|
作者
Gustafson, Jonathan A. [1 ]
Mell, Steven P. [1 ]
Levine, Brett R. [1 ]
Pourzal, Robin [1 ]
Lundberg, Hannah J. [1 ]
机构
[1] Rush Univ, Dept Orthoped Surg, Med Ctr, 1611 W Harrison St,Suite 204, Chicago, IL 60612 USA
关键词
arthroplasty-hip; biomechanics; finite element analysis; modeling; FINITE-ELEMENT; NECK JUNCTION; FRETTING WEAR; CORROSION; DAMAGE; COMPONENTS; INTERFACE; FIXATION; STRENGTH; IMPLANTS;
D O I
10.1002/jor.25357
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Implant failure due to fretting corrosion at the head-stem modular junction is an increasing problem in modular total hip arthroplasty. The effect of varying microgroove topography on modular junction contact mechanics has not been well characterized. The aim of this study was to employ a novel, microgrooved finite element (FEA) model of the hip taper interface and assess the role of microgroove geometry and taper mismatch angle on the modular junction mechanics during assembly. A two-dimensional, axisymmetric FEA model was created using a modern 12/14 taper design of a CoCrMo femoral head taper and Ti6Al4V stem taper. Microgrooves were modeled at the contacting interface of the tapers and varied based on height and spacing measurements obtained from a repository of measured retrievals. Additionally, taper angular mismatch between the head and stem was varied to simulate proximal- and distal-locked engagement. Forty simulations were conducted to parametrically evaluate the effects of microgroove surface topography and angular mismatch on predicted contact area, contact pressure, and equivalent plastic strain. Multiple linear regression analysis was highly significant (p < 0.001; R-2 > 0.74) for all outcome variables. The regression analysis identified microgroove geometry on the head taper to have the greatest influence on modular junction contact mechanics. Additionally, there was a significant second order relationship between both peak contact pressure (p < 0.001) and plastic strain (p < 0.001) with taper mismatch angle. These modeling techniques will be used to identify the implant parameters that maximize taper interference strength via large in-silico parametric studies.
引用
收藏
页码:418 / 425
页数:8
相关论文
共 13 条
  • [1] Quantification of the Contact Area at the Head-Stem Taper Interface of Modular Hip Prostheses
    Witt, Florian
    Guehrs, Julian
    Morlock, Michael M.
    Bishop, Nicholas E.
    PLOS ONE, 2015, 10 (08):
  • [2] Geometric Variations of Modular Head-Stem Taper Junctions of Total Hip Replacements
    Wade, A.
    Beadling, A. R.
    Neville, A.
    De Villiers, D.
    Cullum, C. J.
    Collins, S.
    Bryant, M. G.
    MEDICAL ENGINEERING & PHYSICS, 2020, 83 : 34 - 47
  • [3] Contact Mechanics and Plastic Deformation at the Local Surface Topography Level After Assembly of Modular Head-Neck Junctions in Modern Total Hip Replacement Devices
    Lundberg, Hannah J.
    Ha, Nguyen Q.
    Hall, Deborah J.
    Urban, Robert M.
    Levine, Brett R.
    Pourzal, Robin
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 59 - 82
  • [4] A combined experimental and finite element approach to analyse the fretting mechanism of the head-stem taper junction in total hip replacement
    Bitter, Thom
    Khan, Imran
    Marriott, Tim
    Lovelady, Elaine
    Verdonschot, Nico
    Janssen, Dennis
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2017, 231 (09) : 862 - 870
  • [5] A computational approach to fretting wear prediction at the head-stem taper junction of total hip replacements
    English, Russell
    Ashkanfar, Ariyan
    Rothwell, Glynn
    WEAR, 2015, 338 : 210 - 220
  • [6] Method for Characterization of Material Loss from Modular Head-Stem Taper Surfaces of Hip Replacement Devices
    Racasan, Radu
    Bills, Paul
    Blunt, Liam
    Hart, Alister
    Skinner, John
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 132 - 146
  • [7] The wear of the taper junction of the stem trunnion with the femoral head of a total modular hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    JURNAL TRIBOLOGI, 2019, 23 : 61 - 75
  • [8] Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers
    Bechstedt, Maren
    Gustafson, Jonathan A.
    Mell, Steven P.
    Guehrs, Julian
    Morlock, Michael M.
    Levine, Brett R.
    Lundberg, Hannah J.
    JOURNAL OF BIOMECHANICS, 2020, 103
  • [9] An adaptive finite element simulation of fretting wear damage at the head neck taper junction of total hip replacement: The role of taper angle mismatch
    Fallahnezhad, Khosro
    Oskouei, Reza H.
    Badnava, Hojjat
    Taylor, Mark
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 75 : 58 - 67
  • [10] Micromotion at the head-stem taper junction of total hip prostheses is influenced by prosthesis design-, patient- and surgeon-related factors
    Falkenberg, Adrian
    Biller, Sara
    Morlock, Michael M.
    Huber, Gerd
    JOURNAL OF BIOMECHANICS, 2020, 98 (98)