Case Study of Intrathecal Baclofen Pump Motor Shutdown Secondary to the Effect of the Magnetic Field Created by a Personal Digital Tablet and Magnetic Cover

被引:0
|
作者
Fostier, Michele [1 ,4 ]
Gustin, Thierry [2 ]
Deltombe, Thierry [3 ]
机构
[1] CHU UCL Namur Site Godinne, Dept Anesthesiol, Yvoir, Belgium
[2] CHU UCL Namur Site Godinne, Dept Neurosurg, Yvoir, Belgium
[3] CHU UCL Namur Site Godinne, Dept Phys Med & Rehabil, Yvoir, Belgium
[4] CHU UCL Namur Site Godinne, 1 Ave Gaston Therasse, B-5530 Yvoir, Belgium
关键词
Intrathecal Baclofen Pump; Spasticity; Withdrawal Syndrome; Magnetic Field; WITHDRAWAL;
D O I
10.1097/PHM.0000000000002285
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
We present the case of a posttraumatic C6 AIS A tetraplegic patient with spasticity treated with an intrathecal baclofen pump, who noticed a transient increase in his spasticity each time he used a digital tablet (Ipad) protected by a magnetic shell placed on his abdomen. Telemetry confirmed transient motor shutdown responsible for withdrawal symptoms each time the tablet was used. Symptoms resolved after the removal of the protective shell. Effects of magnetic fields like magnetic resonance imaging are known to stall the pump rotor, which recover at the end of magnetic resonance imaging. Other sources of magnetic fields like laptops or new smartphones with magnet charging technology may also interfere with implanted devices. We therefore recommend patients to avoid close contact of magnetic devices with the intrathecal baclofen pump. More robust studies are warranted to assess the effect of the new magnetic technologies on the function of intrathecal pumps.
引用
收藏
页码:e160 / e161
页数:2
相关论文
共 13 条
  • [1] Study on Magnetic Field Effect of AC Motor
    Feng, Yue
    Zhao, Yang
    ADVANCED RESEARCH ON INDUSTRY, INFORMATION SYSTEMS AND MATERIAL ENGINEERING, PTS 1-7, 2011, 204-210 : 1705 - 1709
  • [2] A nested case-control study of personal magnetic field measures and miscarriage
    Lee, GM
    Neutra, RR
    Yost, M
    Hristova, L
    Hiatt, RA
    EPIDEMIOLOGY, 2001, 12 (04) : S18 - S18
  • [3] A nested case-control study of residential and personal magnetic field measures and miscarriages
    Lee, GM
    Neutra, RR
    Hristova, L
    Yost, M
    Hiatt, RA
    EPIDEMIOLOGY, 2002, 13 (01) : 21 - 31
  • [4] Study the Effect of Magnetic Field on Gaseous Flames using Digital Speckle Pattern Interferometry
    Kumar, Manoj
    Agarwal, Shilpi
    Kumar, Varun
    Khan, Gufran S.
    Shakher, Chandra
    LIGHT AND ITS INTERACTIONS WITH MATTER, 2014, 1620 : 273 - 281
  • [5] OSCILLOGRAPHIC STUDY OF THE MAGNETIC FIELD OF A DIRECT CURRENT ELECTRIC MOTOR BY MEANS OF THE HALL EFFECT IN FILMS
    ELPATEVSKAIA, OD
    PERCHUK, VA
    SOVIET PHYSICS-TECHNICAL PHYSICS, 1957, 2 (07): : 1478 - 1479
  • [6] Digital holographic study of the effect of magnetic field on the potentiostatic current oscillations of iron in sulfuric acid
    Yang, XG
    Chen, SH
    Li, L
    Wang, C
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2006, 586 (02) : 173 - 179
  • [7] Application of in situ digital holography to the study of the effect of a magnetic field on the anodic dissolution of iron in thichloroacetic acid
    Yang, XG
    Chen, SH
    Wang, C
    Li, L
    JOURNAL OF THE SERBIAN CHEMICAL SOCIETY, 2006, 71 (01) : 67 - 73
  • [8] A nested case-control study of residential and personal magnetic field measures and miscarriages (vol 13, pg 21, 2002)
    Lee, GM
    Neutra, RR
    Hristova, L
    Yost, M
    Hiatt, RA
    EPIDEMIOLOGY, 2003, 14 (02) : 255 - 255
  • [9] Effect of the Magnetic Field Generated by a New NeFeB Cover Screw on Bone Healing around Endosseous Implants: A Case Series Report from Dental Practice
    Gallusi, Gianni
    Strappa, Enrico M.
    Monterubbianesi, Riccardo
    Ferrante, Luigi
    Sampalmieri, Francesco
    Meme, Lucia
    APPLIED SCIENCES-BASEL, 2023, 13 (01):
  • [10] Stabilization of Blue Emitters with Thermally Activated Delayed Fluorescence by the Steric Effect: A Case Study by means of Magnetic Field Effects
    Zhao, Chongguang
    Zhao, Fenggui
    Wang, Kai
    Yu, Haomiao
    Huang, Tianyu
    Wang, Rui
    Zhang, Caixia
    Hu, Bin
    Duan, Lian
    PHYSICAL REVIEW APPLIED, 2020, 14 (03):