Accuracy of iodine quantification in dual energy CT: A phantom study across 3 different CT systems

被引:1
|
作者
Dehlinger, N. [1 ]
Bach, J. [1 ]
Willaume, T. [1 ]
Ohana, M. [1 ,2 ,3 ]
Dillenseger, J. P. [1 ,2 ,3 ,4 ]
机构
[1] Hop Univ Strasbourg, Pole Imagerie Med, Strasbourg, France
[2] Univ Strasbourg, CNRS, ICube UMR 7357, Strasbourg, France
[3] Univ Strasbourg, Fac Medecine Maieut & Sci Sante, Strasbourg, France
[4] Fac Medecine Maieut & Sci Sante, 4Rue Kirschleger, F-67000 Strasbourg, France
关键词
Computed tomography; Phantoms; Dual-energy computed tomography; SQUAMOUS-CELL CARCINOMA; COMPUTED-TOMOGRAPHY; PARAMETERS; HEAD;
D O I
10.1016/j.radi.2023.11.015
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: No study has rigorously compared the performances of iodine quantification on recent CT systems employing different emission-based technologies, depending on the manufacturers and models.Methods: A specific bespoke phantom was used for this study, with 12 known concentrations of iodinated contrast agent: 0.4, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 10.0, 15.0, 20.0, 30.0 and 50.0 mg/mL. Three different dual-energy scanners were tested: one system using dual-source acquisition (CT#1) and two systems using Fast kilovolt-peak switching technology +/- artificial intelligence (AI) reconstruction methods (CT#2 and #3) from two different manufacturers. For each system, helical scans were performed following recommended clinical protocols. Four acquisitions were performed per iodine concentration (mg/mL), and measurements were made on iodine-maps using ROIs. Mean measured values were compared to the known concentrations, and the absolute quantification error (AQE) and the relative percentage error (RPE) were used to compare the performances of each CT.Results: The accuracy of the obtained measurements varied depending on the studied model but not on the acquisition mode (dual-source vs kVp switch +/- AI). The quantification was more precise at high concentrations. RPE values were below 10 % with CT#2 (kVp switch) and below 25 % with CT#1 (dualsource), but were significantly higher with CT#3 (kVp switch thorn AI), exceeding 50 % at low concentrations (<3 mg/mL).Conclusions: With the help of a phantom, we identified variability in the results accuracy depending on the CT model, with sometimes significant deviation. Considering the performances of the different DECT technologies in iodine mapping, dual-source (CT#1) and kVp switch (CT#2) technologies appear more accurate than kVp switch technology combined with deep-learning-based reconstruction (CT#3) especially at low concentrations (<3 mg/mL). Implications for practice: As the primary and daily user of medical imaging devices, the radiographer role is to be attentive to the performance of imaging systems, particularly when performing quantitative acquisitions like iodine-quantification. In CT quantitative imaging (iodine map), it's essential for radiographers to consider their CT systems as measuring tools, and to be aware of their accuracies and limits.(c) 2023 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:226 / 230
页数:5
相关论文
共 50 条
  • [1] Multivendor Comparison of Quantification Accuracy of Iodine Concentration and Attenuation Measurements by Dual-Energy CT: A Phantom Study
    Chen, Yong
    Zhong, Jingyu
    Wang, Lan
    Shi, Xiaomeng
    Chang, Rui
    Fan, Jing
    Jiang, Jiang
    Xia, Yihan
    Yan, Fuhua
    Yao, Weiwu
    Zhang, Huan
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (05) : 827 - 839
  • [2] Evaluation of Intra-Manufacturer Dual-Energy CT Iodine Quantification Accuracy: Phantom Studies
    Zhao, A.
    Borhani, A.
    Catania, R.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [3] Iodine Quantification With Dual-Energy CT: Phantom Study and Preliminary Experience With Renal Masses
    Chandarana, Hersh
    Megibow, Alec J.
    Cohen, Benjamin A.
    Srinivasan, Ramya
    Kim, Danny
    Leidecker, Christianne
    Macari, Michael
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (06) : W693 - W700
  • [4] Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom Study
    Jacobsen, Megan C.
    Schellingerhout, Dawid
    Wood, Cayla A.
    Tamm, Eric P.
    Godoy, Myrna C.
    Sun, Jia
    Cody, Dianna D.
    RADIOLOGY, 2018, 287 (01) : 224 - 234
  • [5] Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT
    Pelgrim, Gert Jan
    van Hamersvelt, Robbert W.
    Willemink, Martin J.
    Schmidt, Bernhard T.
    Flohr, Thomas
    Schilham, Arnold
    Milles, Julien
    Oudkerk, Matthijs
    Leiner, Tim
    Vliegenthart, Rozemarijn
    EUROPEAN RADIOLOGY, 2017, 27 (09) : 3904 - 3912
  • [6] Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT
    Gert Jan Pelgrim
    Robbert W. van Hamersvelt
    Martin J. Willemink
    Bernhard T. Schmidt
    Thomas Flohr
    Arnold Schilham
    Julien Milles
    Matthijs Oudkerk
    Tim Leiner
    Rozemarijn Vliegenthart
    European Radiology, 2017, 27 : 3904 - 3912
  • [7] Dual-Energy CT Iodine Accuracy Across Vendors and Platforms
    Jacobsen, M.
    Wood, C.
    Cody, D.
    MEDICAL PHYSICS, 2016, 43 (06) : 3835 - 3836
  • [8] Multivendor comparison of quantification accuracy of effective atomic number by Dual-Energy CT: A phantom study
    Qin, Tian
    Wang, Mengting
    Fan, Yihan
    Wang, Jing
    Gao, Zhizhen
    Wang, Fan
    Li, Ruomei
    Li, Kui
    Ruan, Chengcheng
    Liang, Baohui
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 180
  • [9] Deep learning image reconstruction to improve accuracy of iodine quantification and image quality in dual-energy CT of the abdomen: a phantom and clinical study
    Fukutomi, Akiyo
    Sofue, Keitaro
    Ueshima, Eisuke
    Negi, Noriyuki
    Ueno, Yoshiko
    Tsujita, Yushi
    Yabe, Shinji
    Yamaguchi, Takeru
    Shimada, Ryuji
    Kusaka, Akiko
    Hori, Masatoshi
    Murakami, Takamichi
    EUROPEAN RADIOLOGY, 2023, 33 (02) : 1388 - 1399
  • [10] Deep learning image reconstruction to improve accuracy of iodine quantification and image quality in dual-energy CT of the abdomen: a phantom and clinical study
    Akiyo Fukutomi
    Keitaro Sofue
    Eisuke Ueshima
    Noriyuki Negi
    Yoshiko Ueno
    Yushi Tsujita
    Shinji Yabe
    Takeru Yamaguchi
    Ryuji Shimada
    Akiko Kusaka
    Masatoshi Hori
    Takamichi Murakami
    European Radiology, 2023, 33 : 1388 - 1399