Impedance spectroscopy of enlarged cochlear implant stimulation electrodes - FEM simulations considering the perilymph

被引:2
|
作者
Sehlmeyer, Merle [1 ]
Bhavsar, Mit B. [2 ]
Biebighaeuser, Julian [1 ]
Hitzemann, Moritz [1 ]
Maier, Hannes [2 ]
Lippmann, Martin [1 ]
Schaefer, Christoph [1 ]
Zimmermann, Stefan [1 ]
机构
[1] Leibniz Univ Hannover, Inst Elect Engn & Measurement Technol, Dept Sensors & Measurement Technol, Appelstr 9A, D-30167 Hannover, Germany
[2] Hannover Med Sch, Dept Otolaryngol, NIFE, Cluster Excellence Hearing4all, Stadtfelddamm 34, D-30625 Hannover, Germany
关键词
cochlear implant; electrical conductivity; impedance; impedance spectroscopy; perilymph;
D O I
10.1515/teme-2023-0091
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Cochlear implants are hearing prostheses for patients with severe to total hearing loss but intact auditory nerve. An external speech processor captures sound from the environment, which is subsequently converted into electrical signals and transmitted to an electrode array in the patient's inner ear. The metallic stimulation electrodes of the electrode array electrically stimulate the spiral ganglion cells of the auditory nerve. The functionality of cochlear implants strongly depends on the possible maximum current stimulating the spiral ganglion cells, which can be affected by, e.g., cell growth around the stimulation electrodes. This in turn decreases the stimulation efficiency leading to decreased hearing. Cell growth, implant position and other changes in the surrounding medium are reflected in a change of the impedance of the stimulation electrodes. The impedance measurement of the stimulation electrodes is already implemented in all common cochlear implant systems to check functionality of the stimulation electrodes after implantation, but the frequency spectrum is normally not analyzed. Although this method can detect cell growth on the stimulation electrodes, it faces limitations when other interfering effects, such as changes in the perilymph and implant position, influence the impedance. This work shows impedance spectroscopic measurements using enlarged cochlear implant models to electrically analyze the surrounding medium, the perilymph, to understand changes in electrode impedance and to later monitor the stimulation efficiency of cochlear implants and to identify possible reasons for decreased hearing ability by impedance spectroscopy. In addition, we use FEM simulations to numerically model the influence of the perilymph composition on the impedance measurement. As shown by a final validation, this model can serve as a basis for an extended simulation model including implant position and cell growth monitoring to predict hearing deterioration in cochlear implant patients. In this context, this work serves as a basis for the development of a holistic prediction model and considers in the first step exclusively the influence of the perilymph composition on the impedance between two stimulation electrodes.
引用
收藏
页码:809 / 821
页数:13
相关论文
共 27 条
  • [1] How Cochlear Implant Position and Perilymph Composition affect Stimulation Electrode Impedance
    Sehlmeyer, Merle
    Biebighaeuser, Julian
    Klinger, Britta
    Bhaysar, Mit B.
    Maier, Hannes
    Zimmermann, Stefan
    2024 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS, MEMEA 2024, 2024,
  • [2] A simple electrical circuit model for impedance spectroscopy with cochlear implant electrodes
    Sehlmeyer, Merle
    Bhavsar, Mit B.
    Zimmermann, Stefan
    Maier, Hannes
    HEARING RESEARCH, 2024, 453
  • [3] Comparison of the electrical stimulation performance of four cochlear implant electrodes
    Choi, CTM
    Lai, WD
    Chen, YB
    IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (05) : 1920 - 1923
  • [4] Electro-Mechanical Stimulation of the Cochlea by Vibrating Cochlear Implant Electrodes
    Mueller, Mathias
    Salcher, Rolf
    Majdani, Omid
    Lenarz, Thomas
    Maier, Hannes
    OTOLOGY & NEUROTOLOGY, 2015, 36 (10) : 1753 - 1758
  • [5] In vivo impedance spectroscopy of deep brain stimulation electrodes
    Lempka, Scott F.
    Miocinovic, Svjetlana
    Johnson, Matthew D.
    Vitek, Jerrold L.
    McIntyre, Cameron C.
    JOURNAL OF NEURAL ENGINEERING, 2009, 6 (04)
  • [6] Study on the effect of the number of electrodes programmed and the area of stimulation in cochlear implant patients
    Naito, A
    Kitano, Y
    Iida, M
    Ishida, K
    Sakai, M
    Koga, K
    UPDATES IN COCHLEAR IMPLANTATION, 2000, 57 : 412 - 416
  • [7] Impedance development after implantation of hybrid-L24 cochlear implant electrodes
    Konrad, Simon
    Buechner, Andreas
    Lenarz, Thomas
    Paasche, Gerrit
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2023, 62 (12) : 1137 - 1144
  • [8] Simulations of tonotopically mapped speech processors for cochlear implant electrodes varying in insertion depth
    Faulkner, A
    Rosen, S
    Stanton, D
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2003, 113 (02): : 1073 - 1080
  • [9] Simulations of tonotopically mapped speech processors for cochlear implant electrodes varying in insertion depth
    Faulkner, A. (andyf@phon.ucl.ac.uk), 1600, Acoustical Society of America (113):
  • [10] Optimizing the number of electrodes with high-rate stimulation of the clarion CII cochlear implant
    Frijns, JHM
    Klop, WMC
    Bonnet, RM
    Briaire, JJ
    ACTA OTO-LARYNGOLOGICA, 2003, 123 (02) : 138 - 142