Camera Wavelength Selection for Multi-wavelength Pulse Transit Time based Blood Pressure Monitoring

被引:6
|
作者
Huang, Yukai [1 ,2 ]
Huang, Dongmin [1 ]
Huang, Jia [4 ]
Lu, Hongzhou [4 ]
He, Min [2 ,3 ]
Wang, Wenjin [1 ]
机构
[1] Southern Univ Sci & Technol, Dept Biomed Engn, Shenzhen, Peoples R China
[2] Chinese Acad Sci, Hangzhou Inst Med HIM, Hangzhou 310022, Zhejiang, Peoples R China
[3] Zhejiang Canc Hosp, Key Lab Head Neck Canc Translat Res Zhejiang Prov, Hangzhou, Peoples R China
[4] Third Peoples Hosp Shenzhen, Intens Care Unit, Shenzhen, Peoples R China
基金
国家重点研发计划; 中国国家自然科学基金;
关键词
Blood pressure; camera photoplethysmography; multi-wavelength; pulse transmit time;
D O I
10.1109/EMBC40787.2023.10340068
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Multi-wavelength pulse transmit time (MV-PTT) is a potential tool for remote blood pressure (BP) monitoring. It uses two wavelengths, typically green (G) and near-infrared (NIR), that have different skin penetration depths to measure the PTT between artery and arterioles of a single site of the skin for BP estimation. However, the impact of wavelength selection for MV-PTT based BP calibration is unknown. In this paper, we explore the combination of different wavelengths of camera photoplethysmography for BP measurement using a modified narrow-band camera centered at G-550/R-660/NIR-850 nm, especially focused on the comparison between G-R (full visible) and G-NIR (hybrid). The experiment was conducted on 17 adult participants in a dark chamber with their BP significantly changed by the protocol of ice water stimulation. The experimental results show that the MV-PTT obtained by G-NIR has a higher correlation with BP, and the fitted model has lower MAE in both the systolic pressure (5.78 mmHg) and diastolic pressure (6.67 mmHg) than others. It is confirmed that a hybrid wavelength of visible (G) and NIR is still essential for accurate BP calibration due to their difference in skin penetration depth that allows proper sensing of different skin layers for this measurement.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] A Preliminary Study on Multi-Wavelength PPG Based Pulse Transit Time Detection for Cuffless Blood Pressure Measurement
    Liu, Jing
    Zhang, Yuan-Ting
    Ding, Xiao-Rong
    Dai, Wen-Xuan
    Zhao, Ni
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 615 - 618
  • [2] Feasibility of Remote Blood Pressure Estimation via Narrow-band Multi-wavelength Pulse Transit Time
    Slapnicar, Gasper
    Wang, Wenjin
    Lustrek, Mitja
    ACM TRANSACTIONS ON SENSOR NETWORKS, 2024, 20 (04)
  • [3] AI-based Multi-Wavelength PPG Device for Blood Pressure Monitoring
    Botrugno, Chiara
    Dheman, Kanika
    Bonazzi, Pietro
    Dell'Olio, Francesco
    Magno, Michele
    2024 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS, MEMEA 2024, 2024,
  • [4] Feasibility of Remote Pulse Transit Time Estimation Using Narrow-band Multi-wavelength Camera Photoplethysmography
    Slapnicar, Gasper
    Wang, Wenjin
    Lustrek, Mitja
    2022 IEEE-EMBS INTERNATIONAL CONFERENCE ON BIOMEDICAL AND HEALTH INFORMATICS (BHI) JOINTLY ORGANISED WITH THE IEEE-EMBS INTERNATIONAL CONFERENCE ON WEARABLE AND IMPLANTABLE BODY SENSOR NETWORKS (BSN'22), 2022,
  • [5] Multi-Wavelength Photoplethysmography Device for the Measurement of Pulse Transit Time in the Skin Microvasculature
    Sirkia, Jukka-Pekka
    Panula, Tuukka
    Kaisti, Matti
    2020 COMPUTING IN CARDIOLOGY, 2020,
  • [6] Assessing systemic vascular resistance using arteriolar pulse transit time based on multi-wavelength photoplethysmography
    Lu, Yiqian
    Yu, Zengjie
    Liu, Jikui
    An, Qi
    Chen, Cong
    Li, Ye
    Wang, Yishan
    PHYSIOLOGICAL MEASUREMENT, 2022, 43 (07)
  • [7] Multi-wavelength pulse plethysmography for real-time drug delivery monitoring
    Adhikari, Pratik
    Magana, Isidro B.
    O'Neal, D. Patrick
    OPTICAL DIAGNOSTICS AND SENSING XIV: TOWARD POINT-OF-CARE DIAGNOSTICS, 2014, 8951
  • [8] Multi-wavelength reflectance pulse oximetry
    Mendelson, Y
    Lewinsky, RM
    Wasserman, Y
    ANESTHESIA AND ANALGESIA, 2002, 94 (01): : S26 - S30
  • [9] Camera-Based Blood Pressure Monitoring Based on Multisite and Multiwavelength Pulse Transit Time Features
    Huang, Yukai
    Huang, Dongmin
    Huang, Jia
    Wang, Guowei
    Pan, Liping
    Lu, Hongzhou
    He, Min
    Wang, Wenjin
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2024, 73
  • [10] Effects of Cuff Inflation and Deflation on Pulse Transit Time Measured from ECG and Multi-wavelength PPG
    Liu, Jing
    Li, Yao
    Ding, Xiao-Rong
    Dai, Wen-Xuan
    Zhang, Yuan-Ting
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 5973 - 5976