Determination of boron in borated concrete by means of a simplified prompt gamma activation analysis at the HOTNES facility

被引:2
|
作者
Pietropaolo, A. [1 ,2 ]
Russo, L. [2 ]
Calamida, A. [2 ]
Campoy, A. I. Castro [2 ]
Cottone, D. [3 ,4 ]
Bedogni, R. [2 ]
机构
[1] ENEA, Dept Fus & Technol Nucl Safety & Secur, Frascati, Italy
[2] INFN, Lab Nazl Frascati, Via Enrico Fermi 54, I-00044 Frascati, Italy
[3] Tecnostrutture Srl, Via A Meucci 26, I-30020 Noventa Di Piave, Italy
[4] SVA Srl BARIBLOCK, Via A Meucci 26, I-30020 Noventa Di Piave, Italy
来源
EUROPEAN PHYSICAL JOURNAL PLUS | 2023年 / 138卷 / 12期
关键词
THERMAL-NEUTRON FACILITY; SHIELDING EFFICIENCY; SORGENTINA-RF;
D O I
10.1140/epjp/s13360-023-04752-1
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The prompt gamma activation analysis (PGAA) is an elemental analysis based on the gamma-ray emission following neutron radiative capture by nuclides. A simplified and compact PGAA for the determination of boron in borated concrete was setup at the HOTNES neutron facility, relying on common laboratory equipment. The thermal neutron field has fluence rate of about 20 cm-2 s-1 and is obtained from a moderated americium-boron neutron source. The gamma-ray detector is a common 3 '' x 3 '' NaI(Tl) scintillator. Samples of borated resin with the same geometry of the concrete samples were manufactured and used as standards. A specific "blank sample" correction was developed to isolate the boron contribution in the spectra obtained by irradiating the concrete samples. Boron quantities in the order of 1.5-2.0 g were measured in the concrete samples with uncertainty in the order of +/- 6%, in agreement with manufacturer's expectations. The Detection Limit of this simplified and compact boron analysis is in the order of 0.3 g in terms of boron mass, in line with values given in literature for PGAA-based boron analysis performed at research fission reactors.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Determination of boron in borated concrete by means of a simplified prompt gamma activation analysis at the HOTNES facility
    A. Pietropaolo
    L. Russo
    A. Calamida
    A. I. Castro Campoy
    D. Cottone
    R. Bedogni
    The European Physical Journal Plus, 138
  • [2] Prompt gamma activation analysis for boron determination in the tens of milligram range at the HOTNES facility
    Bedogni, R.
    Russo, L.
    Campoy, A. I. Castro
    Dashdondog, D.
    Caballero-Pacheco, M. A.
    Calamida, A.
    Pietropaolo, A.
    EUROPEAN PHYSICAL JOURNAL PLUS, 2024, 139 (07):
  • [3] Determination of boron in a black mouse by prompt gamma activation analysis
    Cho, H. J.
    Chun, K. J.
    Park, K. W.
    Chung, Y. S.
    Kim, H. R.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2007, 272 (02) : 403 - 407
  • [4] Determination of boron in a black mouse by prompt gamma activation analysis
    H. J. Cho
    K. J. Chun
    K. W. Park
    Y. S. Chung
    H. R. Kim
    Journal of Radioanalytical and Nuclear Chemistry, 2007, 272 : 403 - 407
  • [5] Prompt gamma ray activation analysis for determination of boron in aqueous solutions
    Kudejová, P
    Jolie, J
    Baechler, S
    Schenker, JL
    CAPTURE GAMMA-RAY SPECTROSCOPY AND RELATED TOPICS, 2003, : 730 - 733
  • [6] Determination of boron in water samples by chemical prompt gamma neutron activation analysis
    Ramanjaneyulu, P. S.
    Sayi, Y. S.
    Nathaniel, T. Newton
    Reddy, A. V. R.
    Ramakumar, K. L.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2007, 273 (02) : 411 - 414
  • [7] Determination of trace boron in steels by prompt gamma-ray activation analysis
    Kim, I. J.
    Cho, K. H.
    Paul, R. L.
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2009, 22 (04): : 302 - 306
  • [8] Prompt gamma-ray activation analysis for determination of boron in aqueous solutions
    Baechler, S
    Kudejova, P
    Jolie, J
    Schenker, JL
    Stritt, N
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 488 (1-2): : 410 - 418
  • [9] Determination of boron in water samples by chemical prompt gamma neutron activation analysis
    P. S. Ramanjaneyulu
    Y. S. Sayi
    T. Newton Nathaniel
    A. V. R. Reddy
    K. L. Ramakumar
    Journal of Radioanalytical and Nuclear Chemistry, 2007, 273 : 411 - 414
  • [10] Determination of boron over a large dynamic range by prompt-gamma activation analysis
    Harrison, R. K.
    Landsberger, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2009, 267 (03): : 513 - 518