In-Vitro Diagnostic Reagent Evaluation of Commercially Available Cardiac Troponin I Assay Kits Using H/D Exchange Mass Spectrometry for Antibody-Epitope Mapping

被引:6
|
作者
Song, Dewei [1 ]
Zhou, Min [2 ]
Sun, Haofeng [1 ,2 ]
Ma, Lingyun [1 ]
Liu, Jianyi [1 ]
Gao, Yanhong [3 ]
Zhang, Qi [1 ,4 ]
Xiao, Peng [1 ]
Sun, Keqi [5 ]
Shen, Min [6 ]
Wang, Xiaojian [7 ]
机构
[1] Natl Inst Metrol, Div Chem Metrol & Analyt Sci, Beijing 100029, Peoples R China
[2] Nanjing Univ Sci & Technol, Sch Chem & Engn, Jiangsu 210094, Peoples R China
[3] First Med Ctr PLA Gen Hosp, Lab Dept, Beijing 100039, Peoples R China
[4] Beijing Univ Chem Technol, Sch Chem Engn, Beijing 100029, Peoples R China
[5] Maccura Biotechnol Co Ltd, Reference Syst Dept, Chengdu 611713, Peoples R China
[6] Med Syst Biotechnol Co Ltd, Reference Lab, Ningbo 315104, Peoples R China
[7] Beijing Strong Biotechnol Inc, Reference Lab, Beijing 101400, Peoples R China
关键词
ACUTE MYOCARDIAL-INFARCTION; STANDARDIZATION; SERUM;
D O I
10.1021/acs.analchem.2c03946
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cardiac troponin I (cTnI) is the biomarker of choice and considered a gold standard for the diagnosis of acute myocardial infarction. However, the quantitative results of cTnI assay kits from different manufacturers are not comparable. Based on the H/D exchange mass spectrometry (HDX-MS) workflow, we developed an in-vitro diagnostic reagent antibody evaluation strategy to analyze the interactions of epitopes and antibody cocktails-(R195, F12, S13) and (D1, D2, pAb2). The HDX results indicate that the quantitative result bias of the different reagents originates from the ability of antibodies to recognize various cTnI complex forms, such as free cTnI, hydrolyzed cTnI, and cTnI combined with cTnT or TnC as binary or ternary complexes (cTnIC, cTnTIC), in blood based on different epitopes. The data obtained from the peptide HDX of interest after treatment with various antibody cocktails clearly indicated epitope specificity. The consistency of quantitative results can be improved by a thorough investigation into the epitopes recognized by the antibodies of various diagnostic kits, which will lead to the standardization of cTnI diagnosis.
引用
收藏
页码:2278 / 2284
页数:7
相关论文
共 1 条
  • [1] Epitope mapping of 7S cashew antigen in complex with antibody by solution-phase H/D exchange monitored by FT-ICR mass spectrometry
    Guan, Xiaoyan
    Noble, Kyle A.
    Tao, Yeqing
    Roux, Kenneth H.
    Sathe, Shridhar K.
    Young, Nicolas L.
    Marshall, Alan G.
    JOURNAL OF MASS SPECTROMETRY, 2015, 50 (06): : 812 - 819