Noninvasive cardiac radioablation for ventricular tachycardia: dosimetric comparison between linear accelerator- and robotic CyberKnife-based radiosurgery systems

被引:1
|
作者
Wang, Ching-Yu [1 ]
Ho, Li-Ting [2 ]
Lin, Lian-Yu [2 ]
Chan, Hsing-Min [1 ]
Chen, Hung-Yi [1 ]
Yu, Tung-Lin [3 ]
Huang, Yu-Sen [5 ,6 ]
Kuo, Sung-Hsin [1 ,4 ]
Lee, Wen-Jeng [5 ,6 ]
Chen, Jenny Ling-Yu [1 ,4 ,6 ]
机构
[1] Natl Taiwan Univ Hosp, Dept Oncol, Div Radiat Oncol, Taipei, Taiwan
[2] Natl Taiwan Univ, Coll Med & Hosp, Dept Internal Med, Div Cardiol, Taipei, Taiwan
[3] Fu Jen Catholic Univ Hosp, Dept Radiat Oncol, Taipei, Taiwan
[4] Natl Taiwan Univ Canc Ctr, Dept Radiat Oncol, 57 Ln 155 Sec 3, Keelung Rd, Taipei 106, Taiwan
[5] Natl Taiwan Univ Hosp, Dept Med Imaging, Chung Shan S Rd 7, Taipei 100, Taiwan
[6] Natl Taiwan Univ, Coll Med, Dept Radiol, Taipei, Taiwan
关键词
CyberKnife; Linear accelerator; Cardiac radioablation; Ventricular tachycardia; BODY RADIATION-THERAPY; COMMITTEE; HEART; SBRT;
D O I
10.1186/s13014-023-02370-w
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Few dosimetric comparisons have been published between linear accelerator (LA)-based systems and CyberKnife (CK)-based robotic radiosurgery systems for cardiac radio-ablation in ventricular tachycardia. This study aimed to compare the dosimetry of noninvasive cardiac radio-ablation deliverable on LA with that on CK.Methods Thirteen patients who underwent noninvasive cardiac radio-ablation by LA were included. The prescribed dose was 25 Gy in 1 fraction, and the average planning target volume was 49.8 +/- 31.0 cm(3) (range, 14.4-93.7 cm(3)). CK plans were generated for comparison.Results Both the CK and LA plans accomplished appropriate dose coverage and normal tissue sparing. Compared with the LA plans, the CK plans achieved significantly lower gradient indices (3.12 +/- 0.71 vs. 3.48 +/- 0.55, p = 0.031) and gradient measures (1.00 +/- 0.29 cm vs. 1.17 +/- 0.29 cm, p < 0.001). They had similar equivalent conformity indices (CK vs. LA: 0.84 +/- 0.08 vs. 0.87 +/- 0.07, p = 0.093) and maximum doses 2 cm from the planning target volume (PTV) in any direction (CK vs. LA: 50.8 +/- 9.9% vs. 53.1 +/- 5.3%, p = 0.423). The dosimetric advantages of CK were more prominent in patients with a PTV of <= 50 cm(3) or a spherical PTV. In patients with a PTV of > 50 cm(3) or a non-spherical PTV, the LA and CK plans were similar regarding dosimetric parameters. CK plans involved more beams (232.2 +/- 110.8 beams vs. 10.0 +/- 1.7 arcs) and longer treatment times (119.2 +/- 43.3 min vs. 22.4 +/- 1.6 min, p = 0.007).Conclusions Both CK and LA are ideal modalities for noninvasive cardiac radio-ablation. Upfront treatment should be considered based on clinical intent.
引用
收藏
页数:11
相关论文
共 23 条
  • [1] Noninvasive cardiac radioablation for ventricular tachycardia: dosimetric comparison between linear accelerator- and robotic CyberKnife-based radiosurgery systems
    Ching-Yu Wang
    Li-Ting Ho
    Lian-Yu Lin
    Hsing-Min Chan
    Hung-Yi Chen
    Tung-Lin Yu
    Yu-Sen Huang
    Sung-Hsin Kuo
    Wen-Jeng Lee
    Jenny Ling-Yu Chen
    Radiation Oncology, 18
  • [2] Cardiac radioablation for ventricular tachycardia: photons versus protons dosimetric comparison
    Rossi, Eleonora
    Mirandola, Alfredo
    Gomez, Luca Maria Colombo
    Mantovani, Laura
    Trombetta, Luca
    Bonzano, Elisabetta
    Valentini, Adele
    Basla, Nicoletta
    Vitolo, Viviana
    Barcellini, Amelia
    Pella, Andrea
    Rordorf, Roberto
    Di Liberto, Riccardo
    Orlandi, Ester
    Ciocca, Mario
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4721 - S4725
  • [3] Dosimetric comparison of linear accelerator-based stereotactic radiosurgery systems
    Sharma, S. D.
    Kumar, Sudhir
    Dagaonkar, S. S.
    Bisht, Geetika
    Dayanand, S.
    Devi, Reena
    Deshpande, S. S.
    Chaudhary, S.
    Bhatt, B. C.
    Kannan, S.
    JOURNAL OF MEDICAL PHYSICS, 2007, 32 (01) : 18 - 23
  • [4] Radiation Therapy Workflow and Dosimetric Analysis from a Phase 1/2 Trial of Noninvasive Cardiac Radioablation for Ventricular Tachycardia
    Knutson, Nets C.
    Samson, Pamela P.
    Hugo, Geoffrey D.
    Goddu, S. Murty
    Reynoso, Francisco J.
    Kavanaugh, James A.
    Mutic, Sasa
    Moore, Kaitlin
    Hilliard, Jessica
    Cuculich, Phillip S.
    Robinson, Clifford G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 104 (05): : 1114 - 1123
  • [5] Dosimetric comparison of fractionated radiosurgery plans using frameless Gamma Knife ICON and CyberKnife systems with linear accelerator-based radiosurgery plans for multiple large brain metastases
    Han, Eun Young
    Wang, He
    Luo, Dershan
    Li, Jing
    Wang, Xin
    JOURNAL OF NEUROSURGERY, 2020, 132 (05) : 1473 - 1479
  • [6] Dosimetric Comparison of Robotic and Linear Accelerator Multi-Leaf Collimator-Based Stereotactic Radiosurgery for Arteriovenous Malformation
    Kaliyaperumal, Venkatesan
    Abraham, Susan
    Veni, Maragatha
    Banerjee, Susovan
    Tamilselvan, S.
    Gupta, Deepak
    Dayanithi, K.
    Manigandan, D.
    Mishra, Saumyaranjan
    Bisht, Shyam Singh
    Kataria, Tejinder
    JOURNAL OF MEDICAL PHYSICS, 2021, 46 (01) : 16 - 25
  • [7] Dosimetric Comparison between Cyberknife and Linac Based Hyperarc Stereotactic Radiosurgery for Brain Metastases
    Singh, K. K.
    Cao, J.
    MEDICAL PHYSICS, 2024, 51 (10) : 7952 - 7952
  • [8] Dosimetric quality and delivery efficiency of robotic radiosurgery for brain metastases: Comparison with C-arm linear accelerator based plans
    Zhang, Shuming
    Yang, Ruijie
    Wang, Xin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (11): : 104 - 110
  • [9] Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife®) stereotactic system plans for acoustic schwannoma
    Dutta, Debnarayan
    Subramanian, S. Balaji
    Murli, V.
    Sudahar, H.
    Kurup, P. G. Gopalakrishna
    Potharaju, Mahadev
    JOURNAL OF NEURO-ONCOLOGY, 2012, 106 (03) : 637 - 642
  • [10] Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife®) stereotactic system plans for acoustic schwannoma
    Debnarayan Dutta
    S. Balaji Subramanian
    V. Murli
    H. Sudahar
    P. G. Gopalakrishna Kurup
    Mahadev Potharaju
    Journal of Neuro-Oncology, 2012, 106 : 637 - 642