摘要
Background: Saxitoxin (STX) is the most toxic marine toxin, which can pose several adverse effects on human health. High sensitivity, fast response, and low-cost detection of STX contamination are of significance to reducing the fishery and seafood industries' loss. Among the various types of biosensors, the electrochemical biosensors have been extensively studied in the detection of STX, but the electrode surface modification material is easy to fall off, resulting in unstable electrochemical signals and poor reproducibility. It is imperative to have a ratiometric electrochemical biosensor for STX.Results: In this study, we developed a novel aptamer-based electrochemical sensor (AECs) for the sensitive detection of STX based on a K3Fe(CN)6 regulated silver nanoparticles (Ag NPs) modified with aptamer. The AECs was constructed by immobilizing aptamer on Ag NPs surfaces. Under optimized conditions, the AECs showed a linear response towards STX in the range from 0.04 to 0.15 mu M with the regression equation of Y =-8.0 + 233.7 X (R2 = 0.9956). The limit of detection (LOD) was calculated to be 1 nM (based on 3 N/S), which is significantly lower than the regulatory limits for STX in seafood. Moreover, the AECs showed excellent sensitivity, reproducibility and stability, as well as the detection in samples with acceptable recovery ranged from 71.2 % to 93.8 %, demonstrating its broad application prospects in detection of STX in seafood samples.Significance: This work proposed an AECs to achieve sensitive detection of STX. A reaction system of K3Fe(CN)6 etched Ag NPs was introduced and used as the signal source to avoid the instability of the electrochemical signal, which can produce a ratiometric electrochemical signal output mode, improving the stability and sensitivity of electrochemical detection of STX.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Towards the development of a sensitive and selective electrochemical aptamer-based ampicillin sensor
    Yu, Zhi-gang
    Sutlief, Arin L.
    Lai, Rebecca Y.
    SENSORS AND ACTUATORS B-CHEMICAL, 2018, 258 : 722 - 729
  • [2] Development of an electrochemical aptamer-based sensor with a sensitive Fe3O4 nanopaticle-redox tag for reagentless protein detection
    Zhang, Song
    Zhou, Guoliang
    Xu, Xiaoli
    Cao, Lili
    Liang, Guohai
    Chen, Hui
    Liu, Baohong
    Kong, Jilie
    ELECTROCHEMISTRY COMMUNICATIONS, 2011, 13 (09) : 928 - 931
  • [3] Peptide aptamer-based biosensor for the detection of human chorionic gonadotropin by converting silver nanoparticles-based colorimetric assay into sensitive electrochemical analysis
    Xia, Ning
    Chen, Zhihua
    Liu, Yadong
    Ren, Huizhu
    Liu, Lin
    SENSORS AND ACTUATORS B-CHEMICAL, 2017, 243 : 784 - 791
  • [4] Highly sensitive and selective thiourea electrochemical sensor based on novel silver nanoparticles/chitosan nanocomposite
    Rashed, Md A.
    Ahmed, Jahir
    Faisal, M.
    Alsareii, S. A.
    Jalalah, Mohammed
    Harraz, Farid A.
    COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 2022, 644
  • [5] Aptamer-based fluorescent platform for ultrasensitive adenosine detection utilizing Fe3O4 magnetic nanoparticles and silver nanoparticles
    Song, Yueyue
    Xu, Guanhong
    Wei, Fangdi
    Cen, Yao
    Sohail, Muhammad
    Shi, Menglan
    Xu, Xiaoman
    Ma, Yunsu
    Ma, Yujie
    Hu, Qin
    MICROCHIMICA ACTA, 2018, 185 (02)
  • [6] Sensitive chemiluminescence determination of phentolamine mesylate and phenoxybenzamine hydrochloride based on K3Fe(CN)6-H2O2-fluorescein
    Xing, Lili
    Tang, Yuhai
    Wang, Zhongcheng
    Song, Hailin
    Shi, Xiyan
    JOURNAL OF LUMINESCENCE, 2013, 137 : 162 - 167
  • [7] Aptamer-based fluorescent platform for ultrasensitive adenosine detection utilizing Fe3O4 magnetic nanoparticles and silver nanoparticles
    Yueyue Song
    Guanhong Xu
    Fangdi Wei
    Yao Cen
    Muhammad Sohail
    Menglan Shi
    Xiaoman Xu
    Yunsu Ma
    Yujie Ma
    Qin Hu
    Microchimica Acta, 2018, 185
  • [8] Rate performance of 3D ordered macroporous carbon based electrodes in aqueous K3Fe(CN)6
    Ortega, Ana T.
    Kraay, Bruce
    Brown, Kenneth L.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2011, 241
  • [9] Novel polymer electrolyte based on oligo(ethylene glycol) 600, K2PdCl4, and K3Fe(CN)6
    Universita degli Studi di Padova, Padova, Italy
    J Mater Res, 12 (3393-3403):
  • [10] A novel polymer electrolyte based on oligo(ethylene glycol) 600, K2PdCl4, and K3Fe(CN)(6)
    DiNoto, V
    JOURNAL OF MATERIALS RESEARCH, 1997, 12 (12) : 3393 - 3403