Evaluation of intrafraction motion with an open immobilization mask for HyperArc treatment of multiple brain metastases

被引:0
|
作者
Tejedor-Aguilar, Natalia [1 ]
Lliso, Francoise [1 ]
Ruiz-Rodriguez, Juan C. [1 ]
Gimeno-Olmos, Jose [1 ]
Carmona, Vicente [1 ]
Bonaque, Jorge [1 ]
Bautista, Juan A. [1 ]
Perez-Calatayud, Jose [1 ,2 ]
机构
[1] Hosp Univ & Politecn La Fe, Radiat Oncol Dept, Av Fernando Abril Martorell 106, Valencia 46026, Spain
[2] Clin Benidorm, Radiat Oncol Dept, Av Alfonso Puchades 8, Alicante 03501, Spain
来源
JOURNAL OF RADIOSURGERY AND SBRT | 2023年 / 8卷 / 04期
关键词
Intrafraction motion; Encompass (TM); radiosurgery; HyperArc; STEREOTACTIC RADIOSURGERY;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: In the implementation of the use of Encompass (TM) partially open immobilization mask to perform SRS of multiple brain metastasis, the evaluation of patient's intrafraction motion (IM) is deemed convenient to verify that the margins applied to the GTV are able to ensure adequate dose coverage to each lesion. Methods: IM was determined by comparing the pre- and post-treatment CBCT images with respect to the simulation CT for a total of 23 fractions. The dosimetric impact on GTV coverage due to translational errors in patient positioning and rotational uncertainties of LINAC's performance was also evaluated. Results: The absolute magnitude of IM was less than 1 mm in all cases. The dosimetric difference on GTV coverage due to patient's IM was inferior to 5%. There was not found any significant correlation between the dosimetric impact of rotational uncertainties with the distance to the isocenter. Conclusions: The margins applied to the GTV are adequate when using Encompass (TM) immobilization device.
引用
收藏
页码:283 / 290
页数:8
相关论文
共 50 条
  • [1] Intrafraction motion during frameless radiosurgery using Varian HyperArc™ and BrainLab Elements™ immobilization systems
    Shah, Amish P.
    Meeks, Dylan T.
    Willoughby, Twyla R.
    Ramakrishna, Naren
    Warner, Christopher J.
    Swanick, Cameron W.
    Kelly, Patrick
    Meeks, Sanford L.
    JOURNAL OF RADIOSURGERY AND SBRT, 2020, 7 (02): : 149 - 156
  • [2] Evaluation of the potential treatment delivery benefits of Varian HyperArc for brain metastases
    Williamson, A.
    James, A.
    Chalmers, A.
    Carruthers, R.
    Nowicki, S.
    McLoone, P.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S278 - S279
  • [3] Intra-fractional patient motion when using the Qfix Encompass immobilization system during HyperArc treatment of patients with brain metastases
    Komiyama, Riho
    Ohira, Shingo
    Ueda, Hikari
    Kanayama, Naoyuki
    Masaoka, Akira
    Isono, Masaru
    Ueda, Yoshihiro
    Miyazaki, Masayoshi
    Teshima, Teruki
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (03): : 254 - 260
  • [4] EVALUATION OF INTRAFRACTION MOTION FOR FRAMELESS STEREOTACTIC RADIOSURGERY (SRS) TREATMENT WITH THE MODIFIED BRAINLAB MASK SYSTEM
    Gete, E.
    Shaffer, R.
    Lee, R.
    McKenzie, M.
    Moiseenko, V.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S432 - S432
  • [5] Rapid Hippocampal Sparing HyperArc VMAT Radiosurgery for Multiple Brain Metastases
    Palmiero, A.
    Critchfield, L.
    Hunter, T.
    St Clair, W.
    Randall, M.
    Pokhrel, D.
    MEDICAL PHYSICS, 2020, 47 (06) : E841 - E841
  • [6] Hyperarc Vmat and Vmat planning for stereotactic radiosurgery in multiple brain metastases
    Celen, Yonca Yahsi
    Dinc, Ozcan
    Arslan, Nazmiye Deniz
    Dag, Sedef
    Dogan, Ayse Kutluhan
    Gunenc, Senar
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2023, 16 (04)
  • [7] Dosimetric advantage of HyperArc over an institutional RapidArc planning technique for stereotactic treatment of multiple brain metastases
    Basharat, Fateen
    Dekker, Kurtis
    Olding, Tim
    Alexander, Kevin
    Kerr, Andrew
    MEDICAL PHYSICS, 2024, 51 (08) : 5786 - 5786
  • [8] Intra-fractional motion error during HyperArc stereotactic radiosurgery on patients with brain metastases: Comparison of open and full-face clamshell-style immobilization devices
    Ohira, Shingo
    Komiyama, Riho
    Kanayama, Naoyuki
    Ueda, Yoshihiro
    Inui, Shoki
    Miyazaki, Masayoshi
    Koizumi, Masahiko
    Konishi, Koji
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (04):
  • [9] HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: a new treatment planning approach
    Ohira, Shingo
    Ueda, Yoshihiro
    Akino, Yuichi
    Hashimoto, Misaki
    Masaoka, Akira
    Hirata, Takero
    Miyazaki, Masayoshi
    Koizumi, Masahiko
    Teshima, Teruki
    RADIATION ONCOLOGY, 2018, 13
  • [10] HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: a new treatment planning approach
    Shingo Ohira
    Yoshihiro Ueda
    Yuichi Akino
    Misaki Hashimoto
    Akira Masaoka
    Takero Hirata
    Masayoshi Miyazaki
    Masahiko Koizumi
    Teruki Teshima
    Radiation Oncology, 13