On the Importance of Consistent Insonation Conditions During Hydrophone Calibration and Use

被引:3
|
作者
Rajagopal, Srinath [1 ]
Robinson, Stephen P. P. [1 ]
Ablitt, Justin [1 ]
Miloro, Piero [1 ]
Wang, Lian [1 ]
Zeqiri, Bajram [1 ]
Hurrell, Andrew [2 ]
机构
[1] Natl Phys Lab, Teddington TW11 0LW, England
[2] Precis Acoust Ltd, Dorchester DT2 8QH, England
关键词
Sonar equipment; Calibration; Acoustics; Sensors; Transducers; Sensitivity; Reflection; exposimetry; hydrophone; ULTRASONIC HYDROPHONES; COMPLEX DECONVOLUTION; FREQUENCY-RANGE; BAND; MHZ; AMPLITUDE; FIELD; KHZ;
D O I
10.1109/TUFFC.2022.3205851
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Hydrophones are generally calibrated in acoustic fields with temporally localized (short pulse) or long duration (tone burst) signals. Free-field conditions are achieved by time gating any reflections from the hydrophone body, mounting structures, and surrounding water tank boundaries arriving at the active sensing element. Consequently, the sensitivity response of the hydrophone is a result of direct waves incident on its active element, free from any contaminating effects of reflections. However, when using tone bursts below 400 kHz to calibrate hydrophones, it may not be possible to isolate the direct wave from reflection artifacts. This means that the sensitivity responses derived at these frequencies using short pulse and tone burst signals might not be comparable as they can be characteristic of the acoustic field interaction with either/both the hydrophone active element alone or the hydrophone active element and body. Therefore, there is a need to consider an appropriate calibration method for a given hydrophone type, depending on whether the eventual application employs short pulse or tone burst acoustic fields. This article presents the findings from a short study comprising four needle-type hydrophones of active element diameters in the range of 1-4 mm. These hydrophones were calibrated from 30 kHz to 1.6 MHz using established calibration methodologies within the underwater acoustics (UWA) and ultrasound (US) areas employed at the National Physical Laboratory (NPL), Teddington, U.K. In UWA tone, burst acoustic fields are used, while in US, it is short pulses. The 2- and 4-mm-diameter needle hydrophones showed the largest variation at the overlapping frequencies, in which the maximum disagreement of UWA calibration was 30% relative to US calibration. For the 4-mm hydrophone, UWA calibration exhibited resonant sensitivity structure between 100 and 450 kHz, but which was absent in US calibration. This observed behavior was further investigated theoretically by using a validated acoustic wave solver to confirm the resonant sensitivity structure seen in the case of UWA calibration. The work contained within illustrates the need to ensure that the method of calibration is carefully considered in the context of the duration of the acoustic signals for which the hydrophone is intended.
引用
收藏
页码:120 / 127
页数:8
相关论文
共 50 条
  • [1] Hydrophone calibration in the presence of reflecting elements with the use of matched spatial filtering
    Isaev, A. E.
    ACOUSTICAL PHYSICS, 2008, 54 (03) : 363 - 370
  • [2] Hydrophone calibration in the presence of reflecting elements with the use of matched spatial filtering
    A. E. Isaev
    Acoustical Physics, 2008, 54 : 363 - 370
  • [3] Importance of Calibration for Effective use of Therapeutic Ultrasound
    Tuna, Hakan
    TURKIYE FIZIKSEL TIP VE REHABILITASYON DERGISI-TURKISH JOURNAL OF PHYSICAL MEDICINE AND REHABILITATION, 2011, 57 (02): : 94 - 100
  • [4] Development of a PVDF membrane hydrophone for use in air-coupled ultrasonic transducer calibration
    Galbraith, W
    Hayward, G
    Benny, G
    1996 IEEE ULTRASONICS SYMPOSIUM, PROCEEDINGS, VOLS 1 AND 2, 1996, : 917 - 920
  • [5] Development of a PVDF membrane hydrophone for use in air-coupled ultrasonic transducer calibration
    Univ of Strathclyde, Glasgow, United Kingdom
    IEEE Trans Ultrason Ferroelectr Freq Control, 6 (1549-1558):
  • [6] Development of a PVDF membrane hydrophone for use in air-coupled ultrasonic transducer calibration
    Galbraith, W
    Hayward, G
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (06) : 1549 - 1558
  • [7] On the use of monochromators during AXAF calibration
    Swartz, DA
    Elsner, RF
    Kolodziejczak, JJ
    O'Dell, SL
    Tennant, AF
    Sulkanen, ME
    Weisskopf, MC
    Edgar, RJ
    X-RAY OPTICS, INSTRUMENTS, AND MISSIONS, 1998, 3444 : 189 - 197
  • [8] Calibration and conditions of use of refractory thermocouples up to 2300 K
    Morice, R
    Devin, E
    TEMPMEKO 2001: 8TH INTERNATIONAL SYMPOSIUM ON TEMPERATURE AND THERMAL MEASUREMENT IN INDUSTRY AND SCIENCE, VOL 1 & 2, PROCEEDINGS, 2002, : 73 - 78
  • [9] The importance of consistent use of denominators across patient groups in assessing responses in clinical trials
    Davies, JK
    Cavenagh, JD
    BRITISH JOURNAL OF HAEMATOLOGY, 2006, 132 (06) : 794 - 795
  • [10] Compulsion and substance use disorder: potential importance of boundary conditions
    Gipson, Cassandra D.
    Beckmann, Joshua S.
    NEUROPSYCHOPHARMACOLOGY, 2023, 48 (03) : 432 - 433