Delayed perforation of an atrial pacemaker electrode: Lifelong risk for a rare but serious complication

被引:1
|
作者
Waechter, Christian [1 ,3 ]
Koenig, Alexander M. [2 ]
Chatzis, Georgios [1 ]
Mueller, Julian [1 ]
Schieffer, Bernhard [1 ]
Luesebrink, Ulrich [1 ]
机构
[1] Philipps Univ Marburg, Univ Hosp Marburg, Dept Cardiol, Marburg, Germany
[2] Philipps Univ Marburg, Univ Hosp Marburg, Clin Diagnost & Intervent Radiol, Marburg, Germany
[3] Univ Hosp Marburg, Dept Cardiol, Baldingerstr, D-35043 Marburg, Germany
来源
CLINICAL CASE REPORTS | 2023年 / 11卷 / 06期
关键词
atrial lead; cardiac perforation; CIED; pacemaker; pericardial effusion; tamponade; LEAD PERFORATION; CARDIAC PERFORATION;
D O I
10.1002/ccr3.7525
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Key Clinical MessageNonspecific symptoms such as pleuritic or pericardial chest pain in cardiovascular implantable electronic devices patients, even with unremarkable ECG or device parameters, should always raise suspicion of electrode perforation, regardless of how long ago the implantation was performed. We report the successful percutaneous management of a 77-year-old woman who had a dual-chamber pacemaker implanted more than 1 year ago and presented with pericarditis pain and compensated pericardial hemorrhagic tamponade. The symptoms were due to very late acute perforation of the atrial lead. This report is intended to raise awareness of procedure-related complications in the large group of cardiovascular implantable electronic device patients. Pleuritic or pericardial pain in these patients should raise suspicion of electrode perforation, as the risk of perforation is not restricted to the period immediately after implantation and a lifelong risk cannot apparently be excluded.
引用
收藏
页数:4
相关论文
共 47 条
  • [1] RARE PACEMAKER COMPLICATION THROUGH PERFORATION OF A VENTRICULAR SCREW-IN ELECTRODE LEADING TO A HEMATOTHORAX
    GERCKENS, U
    GRUBE, E
    KIRCHHOFF, PG
    LUDERITZ, B
    ZEITSCHRIFT FUR KARDIOLOGIE, 1986, 75 (05): : 303 - 305
  • [2] A rare but serious complication of continuous ambulatory peritoneal dialysis: delayed perforation of the colon by the Tenckhoff catheter
    Chu, P. Y.
    Siu, K. L.
    HONG KONG MEDICAL JOURNAL, 2016, 22 (03) : 286 - 288
  • [3] Esophageal perforation: A rare but serious complication of cervical mediastinoscopy
    Landstrom, Fredrik
    Stenberg, Erik
    Wickbom, Anders
    ACTA OTO-LARYNGOLOGICA CASE REPORTS, 2024, 9 (01): : 125 - 128
  • [4] Delayed lead perforation: a rare cause of pacemaker dysfunction
    dos Santos, Luis Ferreira
    Costa, Filipa
    Correia, Emanuel
    Rodrigues, Bruno
    Moreira, Davide
    Correia, Hugo
    Nunes, Luis
    Pipa, Joao
    Costa, Antonio
    de Carvalho, Lopes
    Cabral, Costa
    Ribeiro, Nuno
    de Olival, Ramos
    Santos, Oliveira
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2011, 30 (06) : 599 - 609
  • [5] Inflammatory Pericardial Pseudocyst Secondary to Atrial Myocardial Perforation: A Rare Complication following Transvenous Pacemaker Implantation
    Mowaswes, Mohammad
    Khoury, Tawfik
    Lahav, Ziv
    Sanduka, Ashraf
    Shapira-Daniels, Ayelet
    Shapira, Oz M.
    CASE REPORTS IN CARDIOLOGY, 2018, 2018
  • [6] PERFORATION BY PERMANENT PACEMAKER LEAD A RARE COMPLICATION: A CASE REPORT
    Maqbool, Muhammad Furrakh
    Tawb, Shehzad
    Noeman, Ahmed
    Sajid, Muhammad
    PAKISTAN HEART JOURNAL, 2016, 49 (04): : 200 - 202
  • [7] TWIDDLER'S SYNDROME: A RARE BUT A SERIOUS COMPLICATION OF PACEMAKER IMPLANTATION
    Mansur, Sarmed
    Kassab, Ihab
    Sarsam, Nadine
    Hansalia, Riple
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 71 (11) : 2586 - 2586
  • [8] DELAYED PULMONARY PERFORATION - A RARE COMPLICATION OF TUBE THORACOSTOMY
    RESNICK, DK
    CHEST, 1993, 103 (01) : 311 - 313
  • [9] Gallbladder perforation following transarterial chemoembolisation; a rare but serious complication
    Lim, Eu Jin
    Spanger, Manfred
    Lubel, John S.
    FRONTLINE GASTROENTEROLOGY, 2013, 4 (02) : 135 - 137
  • [10] Delayed perforation of right ventricle with cardiac tamponade: a complication of pacemaker implantation
    Foo, Ning-Ping
    Lin, Hung-Jung
    EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2005, 12 (02) : 89 - 91