Metal artifacts reduction in kV-CT images with polymetallic dentures and complex metals based on MV-CBCT images in radiotherapy

被引:3
|
作者
Ni, Xiaochen [1 ]
Shi, Zhonghua [2 ]
Song, Xinmao [1 ]
Tang, Tianci [1 ]
Li, Shengwei [1 ]
Hou, Zhenfeng [2 ]
Zhang, Wei [2 ]
Wang, Wei Fang [1 ]
Chen, Fu [1 ]
Li, Ji [1 ]
Yang, Gang [1 ]
Li, Ruichen [1 ]
Wang, Xiaoshen [1 ,3 ]
机构
[1] Fudan Univ, Eye & ENT Hosp, Dept Radiotherapy, 83 Fenyang Rd, Shanghai 200031, Peoples R China
[2] Shanghai United Imaging Healthcare Co Ltd, Shanghai 201800, Peoples R China
[3] Fudan Univ, Jiangyue Rd 2600, Shanghai, Peoples R China
关键词
RECONSTRUCTION; NECK; NMAR;
D O I
10.1038/s41598-023-35736-x
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This paper proposes a metal artifact reduction method of using MV-CBCT images to correct metal artifacts in kV-CT images, especially for the complex metal artifacts caused by multi-metal interaction of patients with head and neck tumors. The different tissue regions are segmented in the MV-CBCT images to obtain template images and the metal region is segmented in the kV-CT images. Forward projection is performed to get sinogram of the template images, kV-CT images and metal region images. Artifact images can be reconstructed through those sonograms. Corrected images is generated by subtracting the artifact images from the original kV-CT images. After the first correction, the template images are generated again and brought into the previous step for iteration to get better correction result. CT data set of 7 patients are used in this study, compared with linear interpolation metal artifact (LIMAR) and normalized metal artifact reduction method, mean relative error of CT value is reduced by 50.5% and 63.3%, noise is reduced by 56.2% and 58.9%. The Identifiability Score of the tooth, upper/lower jaw, tongue, lips, masseter muscle and cavity in the corrected images by the proposed method have significantly improved (P < 0.05) than original images. The artifacts correction method proposed in this paper can effectively remove the metal artifacts in the images and greatly improve the CT value accuracy, especially in the case of multi-metal and complex metal implantation.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Metal artifacts reduction in kV-CT images with polymetallic dentures and complex metals based on MV-CBCT images in radiotherapy
    Xiaochen Ni
    Zhonghua Shi
    Xinmao Song
    Tianci Tang
    Shengwei Li
    Zhenfeng Hou
    Wei Zhang
    Wei Fang Wang
    Fu Chen
    Ji Li
    Gang Yang
    Ruichen Li
    Xiaoshen Wang
    Scientific Reports, 13
  • [2] Generation of hybrid sinograms for the recovery of kV-CT images with metal artifacts for helical tomotherapy
    Jeon, Hosang
    Park, Dahl
    Youn, Hanbean
    Nam, Jiho
    Lee, Jayoung
    Kim, Wontaek
    Ki, Yongkan
    Kim, Yong Ho
    Lee, Ju Hye
    Kim, Dongwon
    Kim, Ho Kyung
    MEDICAL PHYSICS, 2015, 42 (08) : 4654 - 4667
  • [3] A kV-MV approach to CBCT metal artifact reduction using multi-layer MV-CBCT
    Jacobson, Matthew W.
    Harris, Tom
    Myronakis, Marios
    Lehmann, Mathias
    Huber, Pascal
    Ozoemelam, Ikechi
    Hu, Yue-Houng
    Ferguson, Dianne
    Fueglistaller, Rony
    Morf, Daniel
    Berbeco, Ross
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (07):
  • [4] Automatic Segmentation of KV-CT and MV-CT Images of Head and Neck Patients
    Gering, D. T.
    Lu, W.
    Ruchala, K.
    Olivera, G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S490 - S491
  • [5] Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
    Held, Mareike
    Cremers, Florian
    Sneed, Penny K.
    Braunstein, Steve
    Fogh, Shannon E.
    Nakamura, Jean
    Barani, Igor
    Perez-Andujar, Angelica
    Pouliot, Jean
    Morin, Olivier
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 279 - 290
  • [6] Reduction of imaging artifacts during registration of CT and CBCT images
    Nord, J.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S229 - S229
  • [7] Hybrid modality dual-energy imaging aggregating complementary advantages of kV-CT and MV-CBCT: concept proposal and clinical validation
    Qi, Junfeng
    Yu, Shutong
    Dong, Zhengkun
    Liu, Jiang
    Deng, Juan
    Mei, Guojian
    Yin, Chuou
    Li, Qiao
    Li, Tian
    Wang, Shi
    Zhang, Yibao
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (21):
  • [8] A method based on interpolation for metal artifacts reduction in CT images
    Department of Engineering Physics, Tsinghua University, Beijing, 100084, China
    J X Ray Sci Technol, 2006, 1 (11-19):
  • [9] A method based on interpolation for metal artifacts reduction in CT images
    Gu, Jian-wei
    Zhang, Li
    Chen, Zhi-qiang
    Xing, Yu-xiang
    Huang, Zhi-feng
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2006, 14 (01) : 11 - 19
  • [10] Metals in PET/CT: Causes And Reduction of Artifacts in PET Images
    Lemmens, Catherine
    Nuyts, Johan
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3459 - 3467