Rapid Determination of Trace Elements in Water by Total Reflection X-Ray Fluorescence Spectrometry Using Direct Sampling

被引:1
|
作者
Wu Lei [1 ]
Li Ling-yun [2 ]
Peng Yong-zhen [1 ]
机构
[1] Beijing Univ Technol, Natl Engn Lab Adv Municipal Wastewater Treatment, Beijing 100124, Peoples R China
[2] Beijing Enterprises Water Grp Ltd, Beijing 100102, Peoples R China
关键词
TXRF; Rapid; Multi-element Analysis; Low mineral water; Metal element;
D O I
10.3964/j.issn.1000-0593(2023)03-0990-07
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Compared with the complicated enrichment procedure, a rapid, simple and reliable method is urgently needed to determine trace metal elements in drinking water. Total reflection X-ray Fluorescence spectrometry (TXRF) is a convenient and quantitative method for simultaneous analysis of trace multi-metal elements requiring less samples and short measured time and can be analysed directly without samples pretreatment. In this paper, Ga was used as the internal standard. The feasibility of rapid determination of multi-mass concentration gradient and the multi-element metal solution was explored by direct injectionTXRF method and then applied to the low mineral drinking water for the trace metal elements analysis. The experimental results showed that Al, K, Ca, Mn, Fe, Co, Ni, Cu, Zn and Sr can be analyzed immediately. However, the experiment results found that Al, K and Ca, as light elements, are difficult to achieve accurate quantification due to the recovery rate deviating from the standard value, the reason for the high matrix effect and low elements sensitivity. In contrast, other elements achieve the quantitative requirements. It was found that Mn, Fe, Co, Ni, Cu, Zn and Sr showed good accuracy and precision when the concentrations of metal elements were 40 mg center dot L (1), 4 mg center dot L (1), 0. 4 mg center dot L (1) and 40 mu g center dot L (1), respectively. The Recovery Rate (RR) was 80%similar to 112%. The relative Standard Deviation (RSD) was 3. 6 % similar to 10. 5 %, and the Detection Limit (DL) was 0. 001 similar to 0. 07 mg center dot L-1. With the decrease in the concentration gradient, the accuracy and precision appeared to have different degrees of decline. When the mass concentration was at the lowest level of 4 mu g center dot L (1) in this experiment, the RR and RSD of most elements (except Mn) significantly deviated from the standard value. This paper used the direct injection-TXRF method to test the RR of drinking mineral water at low, medium and high levels. The results showed that Mn, Fe, Co, Ni, Cu, Zn and Sr in the samples were basically at the concentration of mu g center dot L-1 levels, the average RR ranged from 90% to 110% and the average RSD was less than 12%, which met the qualification of micro estimation. In summary, Multi-element test results showed that TXRF is more suitable for heavy elements (Z>20) in selecting elements. Water samples with more than ten components of p_g " L (1) level can directly achieve rapid and accurate quantitative analysis without complex pretreatment for enrichment. Preconcentration techniques are needed to improve the accuracy of ultra-trace level samples in the environment.
引用
收藏
页码:990 / 996
页数:7
相关论文
共 14 条
  • [1] TXRF Study on Mobile Forms of Elements in Soils
    Chu Bin-bin
    Wang Ji-yan
    Zhan Xiu-chun
    Yao Wen-sheng
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40 (07) : 2278 - 2282
  • [2] Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater
    Dhara, Sangita
    Misra, N. L.
    [J]. PRAMANA-JOURNAL OF PHYSICS, 2011, 76 (02): : 361 - 366
  • [3] An assessment of landfill leachate influences on water quality in the Boi Pigua river basin (Parana, Brazil) using the TXRF technique
    Dullius, Thaisa
    Espinoza-Quinones, Fernando Rodolfo
    Obregon, Phallcha Luizar
    Modenes, Aparecido Nivaldo
    Ribeiro, Iuri
    de Pauli, Aline Roberta
    [J]. ENVIRONMENTAL EARTH SCIENCES, 2020, 79 (05)
  • [4] Total reflection X-ray fluorescence analysis of highly mineralized water samples using relative intensities and scattered radiation
    Garmay, Andrey, V
    Oskolok, Kirill, V
    Monogarova, Oksana, V
    Alov, Nikolai, V
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2019, 152 : 74 - 83
  • [5] Georgieva RH, 2014, BULG CHEM COMMUN, V46, P840
  • [6] Total reflection X-ray fluorescence analysis of trace elements in Bulgarian bottled mineral waters of low and high mineral content
    Georgieva, Ralitsa
    Detcheva, Albena
    Karadjov, Methody
    Jordanov, Juri
    Ivanova, Elisaveta
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2013, 93 (10) : 1043 - 1051
  • [7] Analysis of trace elements in soft drink, environmental water, and leaching solution of toy by using portable total reflection x-ray fluorescence spectrometer
    Kunimura, Shinsuke
    Watanabe, Daisuke
    Kawai, Jun
    [J]. BUNSEKI KAGAKU, 2008, 57 (02) : 135 - 139
  • [8] Trace elemental analysis of commercial bottled drinking water by a portable total reflection X-ray fluorescence spectrometer
    Kunimura, Shinsuke
    Kawai, Jun
    [J]. ANALYTICAL SCIENCES, 2007, 23 (10) : 1185 - 1188
  • [9] Rainwater analysis by synchrotron radiation-total reflection X-ray fluorescence
    Lopez, Maria L.
    Ceppi, Sergio A.
    Asar, Maria L.
    Buergesser, Rodrigo E.
    Avila, Eldo E.
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2015, 113 : 100 - 105
  • [10] Comparison of total reflection X-ray fluorescence spectroscopy and inductively coupled plasma atomic emission spectroscopy for thequantification of gold nanoparticle uptake
    Mankovskii, Gabriella
    Pejovic-Milic, Ana
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2020, 164