T-wave oversensing after implantation of pacemakers with automatic sensitivity adjustment in patients with atrioventricular block

被引:0
|
作者
Takahashi, Masao [1 ,2 ]
Kujiraoka, Hirofumi [1 ]
Arai, Tomoyuki [1 ]
Hojo, Rintaro [1 ]
Fukamizu, Seiji [1 ]
机构
[1] Tokyo Metropolitan Hiroo Gen Hosp, Dept Cardiol, Cardiol, Tokyo, Japan
[2] Tokyo Metropolitan Hiroo Gen Hosp, Dept Cardiol, 2-34-10 EbisuShibuya ku, Tokyo 1500013, Japan
来源
关键词
atrioventricular block; automatic sensitivity adjustment; pacemaker; T-wave oversensing; ventricular pacing suppression;
D O I
10.1111/pace.14773
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
T-wave oversensing in pacemakers is rare because the set sensitivity is generally fixed. However, several models of pacemaker employ automatic sensitivity adjustment. Here, we present two cases of atrioventricular block treated by implantation of the pacemaker with automatic sensitivity adjustment. After implanting the pacemaker with automatic sensitivity adjustment, ventricular pacing suppression due to T-wave oversensing occurred. In both cases, T-wave oversensing disappeared after adjusting the setting sensitivity from 0.9 to 2.0 mV.
引用
收藏
页码:561 / 563
页数:3
相关论文
共 50 条
  • [1] T-wave oversensing during exercise one year after cardioverter defibrillator implantation for Brugada syndrome
    Otmani, A.
    Rey, J. -L
    Leborgne, L.
    ARCHIVES OF CARDIOVASCULAR DISEASES, 2008, 101 (04) : 292 - 294
  • [2] Role of exercise electrocardiogram to screen for T-wave oversensing after implantation of subcutaneous implantable cardioverter-defibrillator
    Afzal, Muhammad R.
    Evenson, Christopher
    Badin, Auroa
    Patel, Dilesh
    Godara, Hemant
    Essandoh, Michael
    Okabe, Toshimasa
    Tyler, Jaret
    Houmsse, Mahmoud
    Augostini, Ralph
    Hummel, John
    Kalbfleisch, Steven
    Daoud, Emile G.
    Weiss, Raul
    HEART RHYTHM, 2017, 14 (10) : 1436 - 1439
  • [3] Prediction and prevention of T-WAVE oversensing in patients with implantable cardioverter defibrillators
    Yao, T.
    Jo, H.
    Nakazawa, Y.
    Ashihara, T.
    Itoh, H.
    Sugimoto, Y.
    Ito, M.
    Horie, M.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2007, 18 : S93 - S94
  • [4] T-wave window adjustment for the determination of microvolt T-wave alternans for all patients?
    Madias, John E.
    JOURNAL OF ELECTROCARDIOLOGY, 2013, 46 (01) : 72 - 72
  • [5] FUNCTIONAL DISTURBANCES IN PATIENTS WITH PACEMAKERS DUE TO T-WAVE SUPPRESSION
    DIEWITZ, M
    MUNCHENER MEDIZINISCHE WOCHENSCHRIFT, 1980, 122 (08): : 275 - 276
  • [6] Deep T-Wave Inversions After Pacemaker Adjustment
    DeFilippis, Ersilia M.
    Berg, David D.
    Polk, Donna M.
    JAMA INTERNAL MEDICINE, 2016, 176 (06) : 839 - 840
  • [7] T-wave oversensing is more dependent on ICD sensitivity algorithm than lead sensing configuration
    Punjabi, Anil H.
    Dermovsesian, Arlene
    Karnik, Ankur A.
    JOURNAL OF ELECTROCARDIOLOGY, 2018, 51 (04) : 734 - 737
  • [8] T-wave oversensing by an implantable cardioverter defibrillator after successful ablation of idiopathic ventricular fibrillation
    Strohmer, B
    Schernthaner, C
    Pichler, M
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (04): : 431 - 435
  • [9] Device updates successfully reduce T-wave oversensing and inappropriate shocks in subcutaneous ICD patients
    Larbig, R.
    Motloch, L. J.
    Bettin, M.
    Fischer, A.
    Bode, N.
    Frommeyer, G.
    Loeher, A.
    Koebe, J.
    Reinke, F.
    Eckardt, L.
    NETHERLANDS HEART JOURNAL, 2018, 26 (12) : 606 - 611
  • [10] Does manual T-wave window adjustment affect microvolt T-wave alternans results in patients with structural heart disease?
    Ramanan, Tharmathai
    Balakumaran, Kathir
    Ravichandran, Sairekha
    Ganocy, Stephen J.
    Oshodi, Ganiyu
    Costantini, Otto
    Dettmer, Mary
    Leo, Peter J.
    Kaufman, Elizabeth S.
    JOURNAL OF ELECTROCARDIOLOGY, 2016, 49 (06) : 967 - 972