Optimization of Electrolytes with Redox Reagents to Improve the Impedimetric Signal for Use with a Low-Cost Analyzer

被引:0
|
作者
Cheng, Yu-Hsuan [1 ]
Chande, Charmi [1 ]
Li, Zhenglong [1 ]
Haridas Menon, Niranjan [1 ]
Kaaliveetil, Sreerag [1 ]
Basuray, Sagnik [1 ]
机构
[1] New Jersey Inst Technol, Otto H York Dept Chem & Mat Engn, 323 Dr Martin Luther King Jr Blvd, Newark, NJ 07102 USA
来源
BIOSENSORS-BASEL | 2023年 / 13卷 / 12期
关键词
impededance spectroscopy; electrochemical detection; redox probes; low-cost analyzer; IMPEDANCE SPECTROSCOPY; BIOSENSOR; PROTEIN; SENSITIVITY; PLATFORM; SENSORS; LABEL;
D O I
10.3390/bios13120999
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The most well-known criterion for POC devices is ASSURED, and affordability, i.e., using low-cost instrumentation, is the most challenging one. This manuscript provides a pathway for transitioning ESSENCE, an impedance-based biosensor platform, from using an expensive benchtop analyzer-KeySight 4294A (similar to$50k)-to using a significantly portable and cheaper USB oscilloscope-Analog Discovery 2 (similar to$200)-with similar sensitivity (around 100 times price difference). To achieve this, we carried out a fundamental study of the interplay between an electrolyte like potassium chloride (KCl), and an electrolyte buffer like phosphate buffered saline (PBS) in the presence and absence of a redox buffer like ferro/ferricyanide system and ([Ru(bpy)3](2+)). Redox molecules in the electrolyte caused a significant change in the Nyquist curve of the impedance depending on the redox molecule type. The redox species and the background electrolyte have their own RC semicircles in the Nyquist curve, whose overlap depends on the redox concentration and electrolyte ionic strength. We found that by increasing the electrolyte ionic strength or the redox concentration, the RC semicircle moves to higher frequencies and vice versa. Importantly, the use of the buffer electrolyte, instead of KCl, led to a lower standard deviation and overall signal (lesser sensitivity). However, to achieve the best results from the biorecognition signal, we chose a buffered electrolyte like PBS with high ionic strength and lowered the redox probe concentrations to minimize the standard deviation and reduce any noise from migrating to the low-cost analyzer. Comparing the two analyzers shows similar results, with a lowered detection limit from the low-cost analyzer.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Low-cost vector signal analyzer for communication experiments
    University of Wyoming, United States
    Computers in Education Journal, 2008, 18 (02): : 92 - 97
  • [2] LOW-COST WAVE ANALYZER
    OTTONELLO, P
    AMERICAN JOURNAL OF PHYSICS, 1977, 45 (01) : 103 - 104
  • [3] LOW-COST LOGIC ANALYZER
    ADAMS, BC
    WIRELESS WORLD, 1979, 85 (1519): : 44 - 48
  • [4] Low-cost genetic analyzer
    不详
    OUTLOOK ON AGRICULTURE, 1996, 25 (03) : 206 - 206
  • [5] Low-cost and versatile trifluoromethylation reagents
    Qing, Feng-Ling
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2018, 256
  • [6] Low-cost dynamic signal analyzer offers 50-kHz bandwidth, low noise
    Personal Engineering and Instrumentation News, 1997, 14 (08):
  • [7] A LOW-COST TOTALLY INTEGRATING METABOLIC ANALYZER
    BRODIE, DA
    ESTON, RG
    HAMMOND, P
    WAREING, P
    JOURNAL OF PHYSIOLOGY-LONDON, 1994, 477P : P2 - P2
  • [8] Design of a Portable Low-Cost Impedance Analyzer
    Al-Ali, Abdulwadood
    Elwakil, Ahmad
    Ahmad, Abdelaziz
    Maundy, Brent
    PROCEEDINGS OF THE 10TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES, VOL 1: BIODEVICES, 2017, : 104 - 109
  • [9] Building of A Low-Cost Wireless Battery Analyzer
    Tan, Boon-Huat Wilson
    Tiew, Chee-Chiang Derrick
    PROCEEDINGS FIFTH INTERNATIONAL CONFERENCE ON INTELLIGENT SYSTEMS, MODELLING AND SIMULATION, 2014, : 765 - 771
  • [10] SIMPLE LOW-COST DIGITAL EVENTS ANALYZER
    YANGOS, JP
    CHEN, SH
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1980, 51 (03): : 344 - 350