Bidirectional Venturi Flowmeter Based on Capacitive Sensors for Spirometry

被引:2
|
作者
Becerra, Laura L. [1 ]
Rafeedi, Tarek [2 ]
Ramanarayanan, Sankaran [3 ]
Frankel, Ian [3 ]
Miller, Juliana [1 ]
Chen, Alexander X. [2 ]
Qie, Yi [2 ]
Lipomi, Darren J. [2 ]
Garudadri, Harinath [4 ]
Ng, Tse Nga [1 ]
机构
[1] Univ Calif San Diego, Dept Elect & Comp Engn, San Diego, CA 92093 USA
[2] Univ Calif San Diego, Dept Nano & Chem Engn, San Diego, CA 92093 USA
[3] Univ Calif San Diego, Dept Mech & Aerosp Engn, San Diego, CA 92093 USA
[4] Univ Calif San Diego, Qualcomm Inst, 9500 Gilman Dr, San Diego, CA 92093 USA
基金
美国国家科学基金会;
关键词
capacitive sensors; flowmeter; pressure sensing; respiration; spirometry; ASTHMA; GAS;
D O I
10.1002/admt.202300627
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In this work, a portable venturi tube capable of measuring bidirectional respiratory flow is developed and correlated the measurements to pulmonary function. Pressure signals are transduced using flexible and compressible capacitive foam sensors embedded into the wall of the device. In this configuration, the sensors are able to provide differential pressure readings, from which the airflow rate passing through the tube could be extrapolated. Utilizing the venturi effect, the geometry of the spirometer tube is designed through finite element analysis to measure respiratory airflow during inhalation and exhalation. The device tube is 3D-printed and used to measure tidal breathing and deep breathing, along with peak expiratory flow rates, on a healthy individual. This spirometer design allows for easy-to-use point-of-care diagnoses and has the potential to improve the care of respiratory illnesses.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Downwhole venturi flowmeter
    Denney, Dennis
    JPT, Journal of Petroleum Technology, 2005, 57 (11):
  • [2] Optimal Design of Venturi Flowmeter Based on CFD Analysis
    Zhao E.
    Li L.
    Wang G.
    Chen D.
    Xu J.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2022, 56 (06): : 1196 - 1203
  • [3] New Venturi Flowmeter Based on Differential Pressure Method
    Liu Zhizhuang
    Hong Tiansheng
    Zhang Wenzhao
    Li Zhen
    Zhang Kun
    Wen Tao
    PROCEEDINGS OF THE SECOND INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOLS 1-2, 2008, : 189 - 193
  • [4] BIDIRECTIONAL FLOWMETER
    不详
    INSTRUMENTATION TECHNOLOGY, 1972, 19 (02): : 49 - &
  • [5] A highly sensitive frontend IC for very robust capacitive vortex flowmeter sensors
    Schmid, Hanspeter
    Huber, Alex
    Suetterlin, Dirk
    Tanner, Werner
    PROCEEDINGS OF THE 40TH EUROPEAN SOLID-STATE CIRCUIT CONFERENCE (ESSCIRC 2014), 2014, : 299 - +
  • [6] Mass flowmeter for screw conveyors based on capacitive sensing
    Fuchs, Anton
    Zangl, Hubert
    Brasseur, Georg
    2007 IEEE INSTRUMENTATION & MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-5, 2007, : 163 - +
  • [7] A DESIGN FOR A RECORDING VENTURI AND LIFTING VANE FLOWMETER
    TALMAN, AJ
    JOURNAL OF AGRICULTURAL ENGINEERING RESEARCH, 1983, 28 (05): : 463 - 467
  • [8] VENTURI FLOWMETER MEASURES 40 MILLION GPD
    不详
    INSTRUMENTS & CONTROL SYSTEMS, 1981, 54 (03): : 90 - 90
  • [9] DISCHARGE COEFFICIENT FOR VENTURI FLOWMETER WITH SHORT LAYING LENGTH
    INOUE, M
    JOURNAL OF FLUIDS ENGINEERING-TRANSACTIONS OF THE ASME, 1982, 104 (04): : 463 - 468
  • [10] Smart floating element flowmeter based on a capacitive angular displacement transducer
    Xu, Y
    Zhang, T
    Wang, HX
    Chen, DY
    FLOW MEASUREMENT AND INSTRUMENTATION, 2005, 16 (01) : 1 - 6