Comparing Biomechanical Properties of Bioabsorbable Suture Anchors: A Comprehensive Review

被引:0
|
作者
Schonebaum, Dorien I. [1 ,2 ]
Garbaccio, Noelle [1 ]
Escobar-Domingo, Maria J. [1 ]
Wood, Sasha [1 ]
Smith, Jade. E. [1 ]
Foster, Lacey [1 ]
Mehdizadeh, Morvarid [1 ]
Cordero, Justin J. [1 ]
Foppiani, Jose A. [1 ,3 ]
Choudry, Umar [3 ]
Kaplan, David L. [4 ]
Lin, Samuel J. [1 ]
机构
[1] Harvard Med Sch, Beth Israel Deaconess Med Ctr, Dept Plast & Reconstruct Surg, Boston, MA 02215 USA
[2] Amsterdam UMC, Dept Plast & Reconstruct Surg, NL-1105 AZ Amsterdam, Netherlands
[3] Univ Minnesota, Dept Plast & Reconstruct Surg, Minneapolis, MN 55455 USA
[4] Tufts Univ, Dept Biomed Engn, Medford, MA 02155 USA
关键词
suture anchor; absorbable suture; non-absorbable suture; all-soft; all-suture; silk; orthopedic surgery; biocompatibility; strength; soft tissue; osteolysis; cyst; rotator cuff repair; arthroscopic; OPEN SHOULDER STABILIZATION; PULL-OUT STRENGTH; ROTATOR CUFF; ABSORBABLE IMPLANTS; FIXATION STRENGTH; INTERNAL-FIXATION; ALL-SUTURE; REPAIR; FAILURE; BONE;
D O I
10.3390/biomimetics10030175
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Suture anchors (SAs) are medical devices used to connect soft tissue to bone. Traditionally these were made of metal; however, in the past few decades, bio-absorbable suture anchors have been created to overcome revision surgeries and other complications caused by metallic SAs. This systematic review aims to analyze the biomechanical properties of these SAs to gain a better understanding of their safety and utilization. A comprehensive systematic review that adhered to the PRISMA guidelines was conducted. Primary outcomes were that the pull-out strength of SAs, the rate of degradation secondarily, and the biocompatibility of all SAs were analyzed. After screening 347 articles, 16 were included in this review. These studies revealed that the pull-out strength of bio-absorbable SAs was not inferior to that of their non-absorbable comparatives. The studies also revealed that the rate of degradation varies widely from 7 to 90 months. It also showed that not all absorbable SAs were fully absorbed within the expected timeframe. This systematic review demonstrates that existing suture anchor materials exhibit comparable pull-out strengths, material-specific degradation rates, and variable biocompatibility. All-suture anchors had promising results in biocompatibility, but evidence fails to identify a single most favorable material. Higher-powered studies that incorporate tissue-specific characteristics, such as rotator cuff tear size, are warranted. To meet demonstrated shortcomings in strength and biocompatibility, we propose silk fibroin as a novel material for suture anchor design for its customizable properties and superior strength.
引用
收藏
页数:19
相关论文
共 50 条
  • [1] Complications of Bioabsorbable Suture Anchors in the Shoulder
    Dhawan, Aman
    Ghodadra, Neil
    Karas, Vasili
    Salata, Michael J.
    Cole, Brian J.
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2012, 40 (06): : 1424 - 1430
  • [2] The properties of damaged and undamaged suture used in metal and bioabsorbable anchors: An in vitro study
    Wright, Patrick B.
    Budoff, Jeffrey E.
    Yeh, Ming Long
    Kelm, Zachary S.
    Luo, Zong-Ping
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2007, 23 (06): : 655 - 661
  • [3] Biomechanical analysis of bioabsorbable suture anchors for rotator cuff repair using osteoporotic and normal bone models
    Yamauchi, Shohei
    Tsukada, Harehiko
    Sasaki, Eiji
    Sasaki, Shizuka
    Kimura, Yuka
    Yamamoto, Yuji
    Tsuda, Eiichi
    Ishibashi, Yasuyuki
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2022, 27 (01) : 115 - 121
  • [4] Insertional Achilles Tendon Repair with Bioabsorbable Anchors and Suture
    Conle , R. L.
    Saxena, A.
    MLTJ-MUSCLES LIGAMENTS AND TENDONS JOURNAL, 2022, 12 (04): : 484 - 489
  • [5] Biomechanical analysis of suture anchors and suture materials in the canine femur
    Giles, James T., II
    Coker, Demir
    Rochat, Mark C.
    Payton, Mark E.
    Subramarian, Vijay
    Bartels, Kenneth E.
    VETERINARY SURGERY, 2008, 37 (01) : 12 - 21
  • [6] Comparison of Suture-Based Anchors and Traditional Bioabsorbable Anchors in Foot and Ankle Surgery
    Hembree, W. Chad
    Tsai, Michael A.
    Parks, Brent G.
    Miller, Stuart D.
    JOURNAL OF FOOT & ANKLE SURGERY, 2017, 56 (01): : 3 - 7
  • [7] Suture anchors, properties versus material and design: a biomechanical study in ovine model
    I. Bisbinas
    E. A. Magnissalis
    I. Gigis
    T. Beslikas
    I. Hatzokos
    I. Christoforidis
    European Journal of Orthopaedic Surgery & Traumatology, 2011, 21 : 95 - 100
  • [8] Suture anchors, properties versus material and design: a biomechanical study in ovine model
    Bisbinas, I.
    Magnissalis, E. A.
    Gigis, I.
    Beslikas, T.
    Hatzokos, I.
    Christoforidis, I.
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2011, 21 (02): : 95 - 100
  • [9] Biomechanical comparison of different suture anchors used in rotator cuff repair surgery-all-suture anchors are equivalent to other suture anchors: a systematic review and network meta-analysis
    Yang, Yi-Shiuan
    Shih, Chien-An
    Fang, Ching-Ju
    Huang, Tzu-Teng
    Hsu, Kai-Lan
    Kuan, Fa-Chuan
    Su, Wei-Ren
    Hong, Chih-Kai
    JOURNAL OF EXPERIMENTAL ORTHOPAEDICS, 2023, 10 (01)
  • [10] Biomechanical comparison of different suture anchors used in rotator cuff repair surgery–all-suture anchors are equivalent to other suture anchors: a systematic review and network meta-analysis
    Yi-Shiuan Yang
    Chien-An Shih
    Ching-Ju Fang
    Tzu-Teng Huang
    Kai-Lan Hsu
    Fa-Chuan Kuan
    Wei-Ren Su
    Chih-Kai Hong
    Journal of Experimental Orthopaedics, 10