Combining Lunar and Solar On-Orbit Calibration for Polarization Spectral Imager

被引:0
|
作者
Kong J. [1 ]
机构
[1] Beijing Key Laboratory for Precision Optoelectronic Measurement Instrument and Technology, School of Optics and Photonics, Beijing Institute of Technology, Beijing
关键词
diffuse reflector fabrication; lunar calibration; polarization spectrum imager; solar calibration;
D O I
10.15918/j.tbit1001-0645.2022.104
中图分类号
学科分类号
摘要
Due to the comprehensive influence of gravity, stress release, heat and component aging, it is necessary to calibrate the space remote sensor periodically on-orbit. In order to improve the accuracy of spaceborne radiation calibration, an on-orbit calibration method was proposed, combining lunar calibration and solar calibration. Firstly, using lunar imaging, the on-orbit absolute radiation calibration of the central field was carried out, and the satellite ground difference and on-orbit change of radiation characteristics of the instrument were calibrated. Then, using solar calibration, the non-uniformity of the detector was obtained. Finally, the calibration uncertainty of the polarization spectrum imager was analyzed. The results show that the total uncertainty of absolute radiation calibration is 3.77%, meeting the requirement of 5% radiation calibration accuracy. The method proposed in this paper provides an important technical support for improving the calibration accuracy of remote sensors and the detection ability of aerospace load. © 2023 Beijing Institute of Technology. All rights reserved.
引用
收藏
页码:394 / 402
页数:8
相关论文
共 17 条
  • [1] WANG Hongbo, HUANG Xiaoxian, FANG Chenyan, Et al., On-orbit spectral calibration method of grating dispersive imaging spectrometer, Spectroscopy and Spectral Analysis, 38, 1, pp. 296-301, (2018)
  • [2] DAI Caihong, KHLEVNOY B, WU Zhifeng, Et al., Bilateral comparison of spectral irradiance between NIM and VNIIOFI from 250 to 2500 nm[J], Mapan-Journal of Metrology Society of India, 32, pp. 243-250, (2017)
  • [3] WU Zhifeng, DAI Caihong, ZHAO Weiqiang, Et al., Spectral irradiance responsivity calibration using tunable lasers, Spectroscopy and Spectral Analysis, 41, 3, pp. 853-857, (2021)
  • [4] XU Weiwei, ZHANG Liming, CHEN Hongyao, Et al., In-flight radiometric calibration of high resolution optical satellite sensor using reflected point sources, Acta Optica Sinica, 37, 3, (2017)
  • [5] STOVER J C., Optical scattering: measurement and analysis, (2012)
  • [6] ZHANG Jianyong, JIN Weiqi, GAO Zhiyun, Et al., Measurement and calibration of infrared integrating radiometer, Transactions of Beijing Institute of Technology, 23, 6, pp. 749-753, (2003)
  • [7] ZHANG Lu, ZHANG Peng, HU Xiuqing, Et al., A novel hyperspectral lunar irradiance model based on ROLO and mean equigonal albedo[J], Optik, 142, pp. 657-664, (2017)
  • [8] CHEN Lin, XU Na, HU Xiuqing, Et al., Study on orbit radiometric calibration for FY-2 visible band based on deep convective cloud, Spectroscopy and Spectral Analysis, 36, 8, pp. 2639-2645, (2016)
  • [9] WU Ronghua, ZHANG Peng, DENG Xiaobing, Et al., Data collection and irradiance conversion of lunar obsevation for MERSI, Optics and Precision Engineering, 27, 8, pp. 1819-1827, (2019)
  • [10] XIONG Xiaoxiong, ANGAL A, BUTLER J, Et al., Global space-based inter-calibration system reflective solar calibration reference: from Aqua MODIS to S-NPP VIIRS, Proceeding of SPIE Asia-Pacific Remote Sensing, (2016)