Wear analysis of material measures and stylus probes in repetitive tactile calibration

被引:0
|
作者
Eifler, Matthias [1 ,2 ]
Volk, Raimund [3 ]
Feifel, Stefan [3 ]
Diederichs, Stella [4 ]
Kerscher, Eberhard [4 ]
Seewig, Joerg [2 ]
机构
[1] IU Int Univ Appl Sci, Juri Gagarin Ring 152, D-99084 Erfurt, Germany
[2] RPTU Kaiserslautern Landau, Inst Measurement & Sensor Technol, Gottlieb Daimler Str,Geb 44, D-67663 Kaiserslautern, Germany
[3] JENOPT Ind Metrol Germany GmbH, Drachenloch 5, D-78054 Villingen, Germany
[4] RPTU Kaiserslautern Landau, Working Grp Mat Testing, Gottlieb-Daimler Str, Geb 49, D-67663 Kaiserslautern, Germany
关键词
Calibration; Roughness measurement; Stylus instrument; Measurement uncertainty; Wear; SKID; CONTACT;
D O I
10.1016/j.measurement.2024.115894
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Tactile industrial measuring instruments are frequently calibrated with material measures defined in ISO 5436-1 to ensure their function, determine their measurement uncertainty and assess the capability of measuring processes. Since these process needs to be repeated on a regular basis, the long-term stability of both material measures and the stylus is essential to enable a repeatable and reliable measurement. Since the tactile sampling of a material measure involves a mechanical interaction, there is the potential for wear when repetitive measurements are performed. We present a comprehensive experimental study on the effects of wear for different stylus instruments - including a reference plane scanning system and multiple skidded scanning systems. Multiple measurement methods are used to determine the state of the stylus and the surface of the material measure, both before and after their mechanical interaction. The impact of the wear on the accuracy of the results can be described, as can the connection between the wear and the hardness of the skidded probe. The study found that significant wear effects on surfaces with even >500 HV become qualitatively visible much earlier than they can be quantitatively detected, with extreme load cases of >1000 repetitive measurements at the same position being necessary to produce major wear influences. Thus, regular visual inspections in case of extensive repetitive measurements can be recommended, just as well as a low measuring force and a reduction of the number of measurements for measurement capability analysis to 12 measurements.
引用
收藏
页数:17
相关论文
共 50 条
  • [1] Silicon standards for assessment and calibration of stylus probes
    Frühauf, J
    Trumpold, H
    CIRP ANNALS-MANUFACTURING TECHNOLOGY, 2002, 51 (01) : 475 - 478
  • [2] Calibration System for Tactile Measuring Probes
    Hermann, Gy.
    Rudas, I.
    IEEE 13TH INTERNATIONAL SYMPOSIUM ON INTELLIGENT SYSTEMS AND INFORMATICS (SISY), 2015, : 144 - 148
  • [3] Design of Calibration System for Tactile Displacement Probes
    Hermann, Gyula
    2012 4TH IEEE INTERNATIONAL SYMPOSIUM ON LOGISTICS AND INDUSTRIAL INFORMATICS (LINDI), 2012, : 169 - 172
  • [4] Uncertainty analysis of roughness standard calibration using stylus instruments
    Haitjema, H.
    Precision Engineering, 1998, 22 (03): : 110 - 119
  • [5] Uncertainty analysis of roughness standard calibration using stylus instruments
    Haitjema, H
    PRECISION ENGINEERING-JOURNAL OF THE AMERICAN SOCIETY FOR PRECISION ENGINEERING, 1998, 22 (02): : 110 - 119
  • [6] Calibration and Analysis of Tactile Sensors as Slip Detectors
    Van Wyk, Karl
    Falco, Joe
    2018 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2018, : 2744 - 2751
  • [7] Analysis of polyacrylamide as a calibration material
    Roberts, David
    Goss, Lisa
    Murphy, Tyrel
    Pak, Joshua
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2016, 251
  • [8] Calibration and use of grid-based tactile pressure sensors in granular material
    Paikowsky, SG
    Hajduk, EL
    GEOTECHNICAL TESTING JOURNAL, 1997, 20 (02): : 218 - 241
  • [9] Soil material, temperature, and salinity effects on calibration of multisensor capacitance probes
    Baumhardt, RL
    Lascano, RJ
    Evett, SR
    SOIL SCIENCE SOCIETY OF AMERICA JOURNAL, 2000, 64 (06) : 1940 - 1946
  • [10] Comprehensive profile and areal calibration by additively manufactured material measures
    Eifler, Matthias
    Hering-Stratemeier, Julian
    von Freymann, Georg
    Seewig, Joerg
    SURFACE TOPOGRAPHY-METROLOGY AND PROPERTIES, 2024, 12 (01)