Enhancing low-level tritium detection in environmental waters: assessing the Hidex ULLA liquid scintillation counter

被引:1
|
作者
Copia, Lorenzo [1 ]
Wassenaar, Leonard I. [1 ,3 ]
Juvonen, Risto [2 ]
Oikari, Timo [2 ]
机构
[1] Vienna Int Ctr, Int Atom Energy Agcy, A-1400 Vienna, Austria
[2] Hidex Oy, Lemminkaisenkatu 62, FIN-20520 Turku, Finland
[3] WasserCluster Lunz Biolog Stn GmbH, Dr Carl Kupelwieser Promenade 5, A-3293 Lunz Am See, Austria
关键词
Tritium; Groundwater; Dating; Liquid scintillation counting; LSC; 3-H; GROUNDWATER; TRACERS;
D O I
10.1016/j.jenvrad.2024.107545
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Tritium (3H) is a vital tracer in isotope hydrology and environmental studies, necessitating accurate and precise low-level detection methods. This study evaluated the performance of the new Hidex ULLA Liquid Scintillation Counter for ultra-low-level tritium assays in environmental water samples. The optimization of instrumental parameters enabled accurate and reproducible 3H measurements. Test samples (10-132 TU) were analyzed in triplicate, including 3H-free water and NIST traceable standards. The results demonstrated excellent accuracy and reproducibility, as evidenced by Zeta scores and linearity across these tritium concentrations. The system's stability over extended measurement time periods ensured reliable performance for routine ultra-low level 3H analysis. The optimization of key parameters such as coincidence time and region of interest (ROI) achieved the highest figure of merit (FoM) by using a 25 ns coincidence time and the DigPb open mode for optimal spectral fittings. Detection limits (DL) based on ISO11929 recommendations were significantly improved but highlighted the need for sample enrichment for most low-level environmental applications, with LSC counting times preferably exceeding 1000 min. Overall, the Hidex ULLA Liquid Scintillation Counter proved to be a robust and reliable instrument suitable for ultra-low-level tritium detection in environmental water samples.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Low level tritium analysis using liquid scintillation counter
    Yoon, Yoon Yeol
    Cho, Soo Young
    Lee, Kil Yong
    Kim, Yongje
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2007, 20 (05): : 419 - 423
  • [2] LOW-LEVEL LIQUID SCINTILLATION-COUNTER PERFORMANCE IN A LOW-LEVEL SURFACE LABORATORY
    KAIHOLA, L
    KOJOLA, H
    KANANEN, R
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1986, 17 (5-6): : 509 - 510
  • [3] PERFORMANCE EVALUATION OF COMMERCIAL SCINTILLATION COCKTAILS FOR LOW-LEVEL TRITIUM COUNTING BY HIGH-CAPACITY LIQUID SCINTILLATION COUNTER
    Kuwata, Haruka
    Tazoe, Hirofumi
    Kranrod, Chutima
    Fujiwara, Kenso
    Terashima, Motoki
    Matsueda, Makoto
    Hirao, Shigekazu
    Akata, Naofumi
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (13-15) : 1014 - 1018
  • [4] A Method of Calibration for Measurement of Low Level Tritium in Environmental Water using a Liquid Scintillation Counter
    Fukutani, Satoshi
    Fukui, Masami
    Koyama, Akio
    Nishimaki, Kenzo
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2008, : 89 - 92
  • [5] VIAL EFFECT AND BACKGROUND SUBTRACTION METHOD IN LOW-LEVEL TRITIUM MEASUREMENT BY LIQUID SCINTILLATION-COUNTER
    MOMOSHIMA, N
    NAKAMURA, Y
    TAKASHIMA, Y
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1983, 34 (12): : 1623 - 1626
  • [6] EVALUATION OF A PROTOTYPE LOW-LEVEL LIQUID SCINTILLATION MULTISAMPLE COUNTER
    EINARSSON, SA
    RADIOCARBON, 1992, 34 (03) : 366 - 373
  • [7] MODEL DYS LOW-LEVEL LIQUID SCINTILLATION-COUNTER
    JIANG, HY
    LU, SW
    FU, SM
    ZHANG, WX
    ZHANG, TK
    YE, YZ
    LI, MF
    FU, PY
    WANG, SX
    PENG, CH
    JIANG, PD
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1983, 34 (04): : 672 - 672
  • [8] LOW-LEVEL COUNTING IN A STANDARD LIQUID SCINTILLATION-COUNTER
    KESSLER, MJ
    APPLIED RADIATION AND ISOTOPES, 1988, 39 (06) : 593 - 593
  • [9] Low-level liquid scintillation spectrometry in environmental monitoring.
    Schönhofer, F
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2003, 226 : U78 - U79
  • [10] Applicability of a 100-mL Polyethylene Vial for Low-Level Tritium Measurement Using a Low-Background Liquid Scintillation Counter
    Oshimi, Yoshinari
    Ohki, Mayu
    Nagano, Misato
    Yasumatsu, Takuyo
    Hara, Masanori
    Akamaru, Satoshi
    Nakayama, Masato
    Shoji, Miki