Textile electrode-structure effects and motion artifact minimization in heart activity signal acquisition

被引:0
|
作者
Cho H. [1 ]
Yang J. [1 ]
Lee J. [2 ]
Lee J. [2 ]
机构
[1] Institute of Symbiotic Life-TECH, Yonsei University /, 50 Yonsei-ro, Seodaemun-gu, Seoul
[2] College of Science and Technology, Konkuk University, 120 Neungdong-ro Gwagjin-gu, Seoul
[3] Departmet of Clothing & Textiles, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul
来源
Lee, Joohyeon (ljhyeon@yonsei.ac.kr) | 1600年 / Institute of Electronics Engineers of Korea卷 / 10期
基金
新加坡国家研究基金会;
关键词
Configuration method; Heart activity signal sensing performance; Motion artifacts; Textile electrodes; Wearable platforms;
D O I
10.5573/IEIESPC.2021.10.3.280
中图分类号
学科分类号
摘要
The purpose of this study is to compare the heart activity signal sensing performance by textile electrodes (based on size and configuration method) in order to obtain the structural requirements of motion artifact-minimizing textile electrodes suitable for heart activity signal acquisition. Computer embroidery was used to create six types of contact textile electrodes for measuring heart activity by manipulating the electrode size and method of configuration. These electrodes were attached to separate chest belts worn sequentially by eight males in good physical health. Heart activity signals were detected while the subjects stood in a static posture for 60 seconds and while they walked in place at a rate of 80 beats per minute for 60 seconds. The measurements were repeated four times. To collect the heart activity signals, a BIOPAC Systems, Inc. ECG100 electrocardiogram amplifier was used at a sampling rate of 1 ㎑, and the detected raw signals were fed through a bandpass filter. The signal power ratio of the heart activity signal detected by each type of electrode was calculated, and statistical analysis was applied to determine the heart activity signal-acquisition performance of the six electrode configurations. The results of this study indicate that, with regard to the textile electrodes’ configuration method, convex electrodes detected higher-quality signals than flat electrodes; however, no significant performance difference was found in detecting heart activity signals among the three electrode sizes. Among the structural requirements of textile electrodes for heart activity signal acquisition, the electrode configuration method was found to be an important factor that affects heart activity signal sensing in static and dynamic states. Plans for implementing a wearable platform integrating the convex textile electrodes based on this study will be investigated in future studies, along with performance improvements to develop smart clothing technology that can detect high-quality heart activity signals without time or space limitations. © 2021 Institute of Electronics and Information Engineers. All rights reserved.
引用
收藏
页码:280 / 289
页数:9
相关论文
共 23 条
  • [1] Investigating the possible effect of electrode support structure on motion artifact in wearable bioelectric signal monitoring
    Alper Cömert
    Jari Hyttinen
    BioMedical Engineering OnLine, 14
  • [2] Investigating the possible effect of electrode support structure on motion artifact in wearable bioelectric signal monitoring
    Comert, Alper
    Hyttinen, Jari
    BIOMEDICAL ENGINEERING ONLINE, 2015, 14
  • [3] Signal Quality Assessment of a Novel ECG Electrode for Motion Artifact Reduction
    Halvaei, Hesam
    Sornmo, Leif
    Stridh, Martin
    SENSORS, 2021, 21 (16)
  • [4] Integration Design of Portable ECG Signal Acquisition With Deep-Learning Based Electrode Motion Artifact Removal on an Embedded System
    Jhang, Yu-Syuan
    Wang, Szu-Ting
    Sheu, Ming-Hwa
    Wang, Szu-Hong
    Lai, Shin-Chi
    IEEE ACCESS, 2022, 10 : 57555 - 57564
  • [5] Integration Design of Portable ECG Signal Acquisition With Deep-Learning Based Electrode Motion Artifact Removal on an Embedded System
    Jhang, Yu-Syuan
    Wang, Szu-Ting
    Sheu, Ming-Hwa
    Wang, Szu-Hong
    Lai, Shin-Chi
    IEEE Access, 2022, 10 : 57555 - 57564
  • [6] Motion Artifact Removal in Ambulatory ECG Signal for Heart Rate Variability Analysis
    Berwal, Deepak
    Vandana, C. R.
    Dewan, Sourya
    Jiji, C., V
    Baghini, Maryam Shojaei
    IEEE SENSORS JOURNAL, 2019, 19 (24) : 12432 - 12442
  • [7] Water-retentive, 3D knitted textile electrode for long-term and motion state bioelectrical signal acquisition
    Zhao, Jingjing
    Deng, Junwen
    Liang, Weijin
    Zhao, Liuyan
    Dong, Ying
    Wang, Xiaohao
    Lin, Liwei
    COMPOSITES SCIENCE AND TECHNOLOGY, 2022, 227
  • [8] A 2.4μA Continuous-time Electrode-Skin Impedance Measurement Circuit For Motion Artifact Monitoring in ECG Acquisition Systems
    Kim, Sunyoung
    Yazicioglu, Refet Firat
    Torfs, Tom
    Dilpreet, Buxi
    Julien, Penders
    Van Hoof, Chris
    2010 SYMPOSIUM ON VLSI CIRCUITS, DIGEST OF TECHNICAL PAPERS, 2010, : 219 - +
  • [9] Effects of Cable Sway, Electrode Surface Area, and Electrode Mass on Electroencephalography Signal Quality during Motion
    Symeonidou, Evangelia-Regkina
    Nordin, Andrew D.
    Hairston, W. David
    Ferris, Daniel P.
    SENSORS, 2018, 18 (04)
  • [10] Development of a differential suction electrode for improved intravaginal recordings of pelvic floor muscle activity: Reliability and motion artifact assessment
    Keshwani, Nadia
    McLean, Linda
    NEUROUROLOGY AND URODYNAMICS, 2012, 31 (08) : 1272 - 1278