In this study, a label-free interdigitated array (IDA) microelectrode-based immunosensor was designed and tested for rapid and sensitive detection of E. coli O157:H7. The surface of gold IDA microelectrodes (25 pairs of fingers with 700 μm length and 10 μm width and space) was modified with protein A and further immobilized with polyclonal antibodies against E. coli O157:H7. A 20 μL sample containing E. coli O157:H7 in 0.01 M phosphate buffered saline (PBS) at pH 7.4 was dropped on the surface of the modified microelectrodes and incubated for 2 h at room temperature. After the target bacteria were captured by the antibodies immobilized on the microelectrode, a wash with PBS was applied to rinse off all unbound parts of the sample. Finally, the impedance, both magnitude and phase angle, was measured over a broad frequency range from 1 Hz to 1 MHz. The results showed that the impedance in a frequency range of 1 Hz to 1 kHz increased with the increasing number of E. coli O157:H7 in a range from 80 to 2 × 105 cells per 20 μL. A linear relationship between the magnitude of measured impedance and the log number of cells was determined at the frequency of 1 kHz. This biosensing method was able to detect as few as 80 cells of E. coli O157:H7 within 2 h. The specificity of the immunosensor was validated using five non-target bacteria, and none of them generated detectable signals. The application of the immunosensor for food safety was demonstrated with food samples including chicken carcasses, ground beef, and fresh cut broccoli inoculated with E. coli O157:H7. At the same time, the capture of E. coli O157:H7 cells on the microelectrode surface was confirmed using atomic force microscopy. © 2010 ASABE ISSN 1934-2799.
引用
收藏
页码:49 / 62
页数:13
相关论文
共 50 条
  • [1] Interdigitated array microelectrode-based electrochemical impedance immunosensor for detection of Escherichia coli O157:H7
    Yang, LJ
    Li, YB
    Erf, GF
    ANALYTICAL CHEMISTRY, 2004, 76 (04) : 1107 - 1113
  • [2] Double interdigitated array microelectrode-based impedance biosensor for detection of viable Escherichia coli O157:H7 in growth medium
    Varshney, Madhukar
    Li, Yanbin
    TALANTA, 2008, 74 (04) : 518 - 525
  • [3] Interdigitated microelectrode geometry for simple electrical Escherichia coli O157:H7 DNA detection
    Nadzirah, Sharipah
    Hashim, Uda
    MICROELECTRONICS INTERNATIONAL, 2017, 34 (02) : 99 - 107
  • [4] A label-free, microfluidics and interdigitated array microelectrode-based impedance biosensor in combination with nanoparticles immunoseparation for detection of Escherichia coli O157:H7 in food samples
    Varshney, Madhukar
    Li, Yanbin
    Srinivasan, Balaji
    Tung, Steve
    SENSORS AND ACTUATORS B-CHEMICAL, 2007, 128 (01): : 99 - 107
  • [5] Immunosensor for detection of Escherichia coli O157:H7 using imaging ellipsometry
    Bae, Young Min
    Park, Kwang-Won
    Oh, Byung-Keun
    Choi, Jeong-Woo
    JOURNAL OF MICROBIOLOGY AND BIOTECHNOLOGY, 2006, 16 (08) : 1169 - 1173
  • [6] Conductometric immunosensor for specific Escherichia coli O157:H7 detection on chemically funcationalizaed interdigitated aptasensor
    Uda, Muhammad Nur Afnan
    Dafhalla, Alaa Kamal Yousif
    Dhahi, Thikra S.
    Adam, Tijjani
    Gopinath, Subash Chandra Bose
    Ambek, Asral Bahari
    Uda, Muhammad Nur Aiman
    Mohammed, Mohammed
    Parmin, Nor Azizah
    Ibrahim, Nur Hulwani
    Hashim, Uda
    HELIYON, 2024, 10 (05)
  • [7] Detection of Escherichia coli O157:H7 using immunosensor based on surface plasmon resonance
    Oh, BK
    Kim, YK
    Bae, YM
    Lee, WH
    Choi, JW
    JOURNAL OF MICROBIOLOGY AND BIOTECHNOLOGY, 2002, 12 (05) : 780 - 786
  • [8] Liposome-based microcapillary immunosensor for detection of Escherichia coli O157:H7
    Ho, JAA
    Hsu, HW
    Huang, MR
    ANALYTICAL BIOCHEMISTRY, 2004, 330 (02) : 342 - 349
  • [9] Rapid detection of Escherichia coli O157:H7 and Salmonella Typhimurium in foods using an electrochemical immunosensor based on screen-printed interdigitated microelectrode and immunomagnetic separation
    Xu, Meng
    Wang, Ronghui
    Li, Yanbin
    TALANTA, 2016, 148 : 200 - 208
  • [10] Amperometric immunosensor for the detection of Escherichia coli O157:H7 in food specimens
    Li, Yan
    Cheng, Ping
    Gong, Junhua
    Fang, Lichao
    Deng, Jun
    Liang, Wenbin
    Zheng, Junsong
    ANALYTICAL BIOCHEMISTRY, 2012, 421 (01) : 227 - 233