Detection limit of DNA analyzer

被引:0
|
作者
Wang, Jie [1 ,2 ]
Wang, Li-Qiang [1 ]
Shi, Yan [1 ]
Zheng, Hua [1 ,3 ]
Lu, Zu-Kang [1 ]
机构
[1] National Engineering and Technology Research Center for Optical Instrument, Zhejiang University, Hangzhou 310027, China
[2] School of Electronic and Information, Guangdong Polytechnic Normal University, Guangzhou 510665, China
[3] School of Physics and Optoelectronics Technology, Fujian Normal University, Fuzhou 350007, China
来源
Guangzi Xuebao/Acta Photonica Sinica | 2008年 / 37卷 / 03期
关键词
Capillary electrophoresis - DNA - Fluorescence - Fluorophores - Photomultipliers;
D O I
暂无
中图分类号
学科分类号
摘要
The sensitivity of the DNA analyzer, which was designed by authors group to detect the samples of capillary electrophoresis (CE) with Laser induced fluorescence (LIF), was analyzed in theory and experiment. The detection sensitivity of fluorescence depends on both the fluorescent intensity entering the sensors (PMT or CCD) and the performance of sensors. Given an actual DNA analyzer, the calculated detection limit of TO (thiazole orange) fluorophore was 0.1271-0.1272 fluors per cubic micron. The experimental result of CE about double-stranded (ds) TO probes (the standard sample pGEM-3Zf(+)/Hae III Markers) is 0.1454 fluors per cubic micron. The study shows that the approaches for improving detection limit are to make the good coating on inner walls of capillary, choose the suitable sieving matrix, enlarge laser power, increase numerical aperture of objective as much as possible and raise the detection lower limit of photomultiplier.
引用
收藏
页码:543 / 546
相关论文
共 50 条
  • [1] TRACE DNA DETECTION BY AN AUTOMATED DNA ANALYZER
    KAMBARA, H
    NAGAI, K
    KAWAMOTO, K
    FASEB JOURNAL, 1992, 6 (01): : A170 - A170
  • [2] Pushing DNA detection limit
    Anon
    Chemical and Engineering News, 2002, 80 (35):
  • [3] Bandwidth and detection limit of a crossing-sampling spectrum analyzer
    Review of Scientific Instruments, 1997, 68 (1 pt 1):
  • [4] Bandwidth and detection limit of a crossing-sampling spectrum analyzer
    Daria, VR
    Saloma, C
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (01): : 240 - 242
  • [5] The diffraction limit of an optical spectrum analyzer
    Kolobrodov, V. G.
    Tymchik, G. S.
    Kolobrodov, M. S.
    TWELFTH INTERNATIONAL CONFERENCE ON CORRELATION OPTICS, 2015, 9809
  • [6] A microfluidic streaming potential analyzer for label-free DNA detection
    Li, Yufeng
    Lai, Sze Nga
    Zheng, Bo
    SENSORS AND ACTUATORS B-CHEMICAL, 2018, 259 : 871 - 877
  • [7] Rapid bio-barcode assay for multiplex DNA detection based on capillary DNA Analyzer
    He, MingXing
    Li, Kai
    Mao, JunHua
    Zhou, YuXun
    JOURNAL OF VIROLOGICAL METHODS, 2008, 151 (01) : 126 - 131
  • [8] LIMIT FOR QUALITATIVE DETECTION OF BENZOYLECGONINE IN URINE USING EMIT REAGENTS AND THE HITACHI 705 AUTOMATIC ANALYZER
    HALL, TD
    SANCHEZ, I
    DUNLAP, L
    DEKANEL, J
    JOURNAL OF ANALYTICAL TOXICOLOGY, 1989, 13 (06) : 371 - 373
  • [9] Rapid detection of natriuretic peptides by a microfluidic LabChip analyzer with DNA aptamers: Application of natriuretic peptide detection
    Lin, Ming-Cheng
    Nawarak, Jiraporn
    Chen, Tai-Yuan
    Tsai, Hsien-Yu
    Hsieh, Jung-Feng
    Sinchaikul, Supachok
    Chen, Shui-Tein
    BIOMICROFLUIDICS, 2009, 3 (03)
  • [10] Toward a portable DNA analyzer
    Borman, S