Microwave imaging solutions for medical imaging using re-weighted basic pursuit algorithm

被引:0
|
作者
Anjit T.A. [1 ]
Benny R. [1 ]
Cherian P. [2 ]
Mythili P. [1 ]
机构
[1] Cochin University of Science and Technology, Cochin, Kerala
[2] College of Engineering, Chengannur, Kerala
关键词
Inverse problems - Errors - Computerized tomography - Convex optimization - Imaging systems - Iterative methods;
D O I
10.2528/pierm20063007
中图分类号
学科分类号
摘要
This paper describes an innovative technique for the quantitative reconstruction of the dielectric and conductivity distribution of objects in a microwave tomography framework using sparse data. The proposed method tries to extract information about the size, shape, localisation, and dielectric distribution of various inclusions within the object under study using an iterative reconstruction methodology in the sparse domain. The proposed algorithm combines the Distorted Born Iteration method (DBIM) and a convex optimization technique for solving the inverse ill-posed problem. The Re-weighted Basis Pursuit (RwBP) algorithm is chosen as the convex optimization technique in this work. The performance of the proposed algorithm has been compared with the TV-norm method, and the results obtained are highly encouraging. The proposed method produces a significant reduction in the reconstruction error as compared to the TV norm method with an error value of 0.083 as against 0.32 in the case of TV norm in the presence of 25 dB noise. By accurately preserving the edges of the inclusions, the proposed technique is found to provide an overall improvement in the reconstruction in terms of tissue differentiation (permittivity and conductivity), dimensions of inclusions, resolution, shape, size, and coordinate localisation of inclusions. The proposed algorithm converges within 10–12 iterations as compared to other complex imaging algorithms available in the literature. Further, this proposed technique is validated using experimental data from an actual breast imaging setup. The three inclusions of 10 mm, 6 mm, and 3 mm have been localised with errors of 0.052, 0.04, and 0.09, respectively. The results obtained from the real-time data show the applicability and feasibility of the proposed algorithm in breast tumor imaging application. © 2020, Electromagnetics Academy. All rights reserved.
引用
收藏
页码:13 / 24
页数:11
相关论文
共 50 条
  • [1] Microwave Imaging Solutions for Medical Imaging Using Re-Weighted Basic Pursuit Algorithm
    Anjit, Thathamkulam A.
    Benny, Ria
    Cherian, Philip
    Mythili, Palayyan
    PROGRESS IN ELECTROMAGNETICS RESEARCH M, 2020, 97 : 13 - 24
  • [2] Tomographic Bioluminescence Imaging by an Iteratively Re-weighted Minimization
    Wu, Ping
    Liu, Kai
    Xue, Zhenwen
    Guo, Wei
    Qin, Chenghu
    Tian, Jie
    MEDICAL IMAGING 2012: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2012, 8317
  • [3] ISAR IMAGING BASED ON HOMOTOPY RE-WEIGHTED l1-NORM MINIMIZATION
    Gao, Yuexin
    Zhang, Xinyu
    Xing, Mengdao
    Fu, Jixiang
    Zhang, Zijing
    Wang, Ying
    2019 IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM (IGARSS 2019), 2019, : 1204 - 1207
  • [4] Microwave imaging using a genetic algorithm
    Caorsi, S
    Pastorino, M
    ELECTROMAGNETIC NONDESTRUCTIVE EVALUATION (II), 1998, 14 : 233 - 242
  • [5] Sparse signal reconstruction from limited data using FOCUSS: A re-weighted minimum norm algorithm
    Gorodnitsky, IF
    Rao, BD
    IEEE TRANSACTIONS ON SIGNAL PROCESSING, 1997, 45 (03) : 600 - 616
  • [6] ROBUST REGRESSION USING ITERATIVELY RE-WEIGHTED LEAST-SQUARES
    HOLLAND, PW
    WELSCH, RE
    COMMUNICATIONS IN STATISTICS PART A-THEORY AND METHODS, 1977, 6 (09): : 813 - 827
  • [7] Muscle and Adipose Mimicking Solutions for Applications in Microwave Medical Imaging
    Pollacco, D. A.
    Conti, M. Caligari
    Farrugia, L.
    Sammut, C. V.
    2019 13TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2019,
  • [8] Damage Imaging of an Isotropic Plate Using Matching Pursuit Algorithm
    Kim, H. W.
    Yuan, F. G.
    SENSORS AND SMART STRUCTURES TECHNOLOGIES FOR CIVIL, MECHANICAL, AND AEROSPACE SYSTEMS 2017, 2017, 10168
  • [9] Microwave Imaging Update For medical imaging applications
    Stancliff, Roger
    2017 SBMO/IEEE MTT-S INTERNATIONAL MICROWAVE AND OPTOELECTRONICS CONFERENCE (IMOC), 2017,
  • [10] Iteratively re-weighted and refined least squares algorithm for robust inversion of geophysical data
    Gholami, Ali
    Aghamiry, Hossein S.
    GEOPHYSICAL PROSPECTING, 2017, 65 : 201 - 215